Inappropriate shock incidence in patients with subcutaneous implantable cardioverter-defibrillators with concomitant cardiac implantable electronic devices: A single-center cohort study

被引:0
|
作者
Okazaki, Makiko [1 ,2 ]
Sahashi, Yuki [3 ]
Nagase, Takahiko [4 ]
Inoue, Kanki [4 ]
Sekiguchi, Yukio [4 ]
Nitta, Junichi [4 ]
Shinoda, Satoru [5 ]
Shimizu, Sayuri [2 ]
Kuroki, Makoto [2 ]
Isobe, Mitsuaki [6 ]
Mihara, Takahiro [2 ,7 ]
机构
[1] Sakakibara Heart Inst, Dept Clin Engn, Fuchu, Tokyo, Japan
[2] Yokohama City Univ, Grad Sch Data Sci, Dept Hlth Data Sci, Yokohama, Kanagawa, Japan
[3] Gifu Univ, Grad Sch Med, Dept Cardiol, Gifu, Japan
[4] Sakakibara Heart Inst, Dept Cardiol, Fuchu, Tokyo, Japan
[5] Yokohama City Univ, Sch Med, Dept Biostat, Yokohama, Kanagawa, Japan
[6] Sakakibara Heart Inst, Fuchu, Tokyo, Japan
[7] Yokohama City Univ, Grad Sch Data Sci, Dept Hlth Data Sci, Yokohama 2360027, Japan
来源
关键词
cardiac implantable electronic devices; concomitant; inappropriate shock; pacing; subcutaneous implantable cardioverter-defibrillator; ICD; OUTCOMES;
D O I
10.1111/pace.14887
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundSubcutaneous implantable cardioverter defibrillators (S-ICDs) are occasionally used in combination with other cardiac implantable electronic devices (CIEDs). However, whether the incidence of inappropriate shock increases in patients with S-ICDs and concomitant CIEDs remains unclear. This study aimed to investigate the association between the concomitant use of CIEDs and the incidence of inappropriate shock in patients with current-generation S-ICDs.MethodsA total of 127 consecutive patients received an S-ICD. Patients were assigned to two groups depending on concomitant use of CIEDs at the time of S-ICD implantation: patients without other CIEDs (non-combined group, 106 patients) and patients with other CIEDs (combined group, 21 patients). CIEDs included pacemakers, implantable cardioverter defibrillators, cardiac resynchronization therapy pacemakers, and cardiac resynchronization therapy defibrillators. The primary outcome was inappropriate shock, defined as a shock other than ventricular arrhythmia. Hazard ratios and 95% confidence intervals were calculated using a time-varying Cox proportional hazards model which was adjusted for age because age differed between the groups and could be a confounder.ResultsDuring a median follow-up period of 2.2 years (interquartile range, 1.0-3.4 years), inappropriate shock events occurred in 17 (16%) and five (19%) patients of the non-combined and combined groups, respectively. While the age-adjusted hazard ratio for inappropriate shock was 24% higher in the combined than in the non-combined group (hazard ratio = 1.24, 95% confidence interval, 0.39-3.97), this difference was insignificant (p = .71).ConclusionThe incidence of inappropriate shock did not differ between patients with and without concomitant use of CIEDs, suggesting that S-ICDs could potentially be combined with other CIEDs without increasing the number of inappropriate shocks. Further studies are warranted to confirm the safety and feasibility of concomitant use of S-ICDs and CIEDs.
引用
收藏
页码:131 / 138
页数:8
相关论文
共 50 条
  • [41] Perioperative Considerations for Patients With Subcutaneous Implantable Cardioverter-Defibrillators Undergoing Noncardiac Surgery
    Essandoh, Michael
    Daoud, Emile G.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2016, 30 (03) : 756 - 761
  • [42] Cardiac Implantable Electronic Device Therapy Permanent Pacemakers, Implantable Cardioverter Defibrillators, and Cardiac Resynchronization Devices
    Steffen, Melanie M.
    Osborn, Jeffery S.
    Cutler, Michael J.
    MEDICAL CLINICS OF NORTH AMERICA, 2019, 103 (05) : 931 - +
  • [43] Treatment of patients with cardiac pacemakers and implantable cardioverter-defibrillators during radiotherapy
    Solan, AN
    Solan, MJ
    Bednarz, G
    Goodkin, MB
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (03): : 897 - 904
  • [44] Predictors of pace-terminable rhythms in patients with subcutaneous implantable cardioverter-defibrillators
    Rao, Birju R.
    Jain, Vardhmaan
    Mekary, Wissam
    Ibrahim, Rand
    Bhatia, Neal K.
    El Chami, Mikhael
    Knops, Reinoud E.
    Lloyd, Michael S.
    HEART RHYTHM, 2024, 21 (11) : 2346 - 2347
  • [45] Subcutaneous implantable cardioverter-defibrillators: long-term results of the EFFORTLESS study
    Lambiase, Pier D.
    Theuns, Dominic A.
    Murgatroyd, Francis
    Barr, Craig
    Eckardt, Lars
    Neuzil, Petr
    Scholten, Marcoen
    Hood, Margaret
    Kuschyk, Jurgen
    Brisben, Amy J.
    Carter, Nathan
    Stivland, Timothy M.
    Knops, Reinoud
    Boersma, Lucas V. A.
    EUROPEAN HEART JOURNAL, 2022, 43 (21) : 2037 - 2050
  • [46] Practice Advisory for the Perioperative Management of Patients with Cardiac Implantable Electronic Devices: Pacemakers and Implantable Cardioverter-Defibrillators 2020 An Updated Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Implantable Electronic Devices*
    Apfelbaum, Jeffrey L.
    Schulman, Peter M.
    Mahajan, Aman
    Connis, Richard T.
    Agarkar, Madhulika
    ANESTHESIOLOGY, 2020, 132 (02) : 225 - 252
  • [47] Eligibility of cardiac resynchronization therapy patients for subcutaneous implantable cardioverter defibrillators
    Giammaria, Massimo
    Lucciola, Maria Teresa
    Amellone, Claudia
    Orlando, Fabrizio
    Mazzone, Giuseppe
    Chiarenza, Sara
    Lovecchio, Mariolina
    Valsecchi, Sergio
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2019, 54 (01) : 49 - 54
  • [48] Eligibility of cardiac resynchronization therapy patients for subcutaneous implantable cardioverter defibrillators
    Moubarak, Ghassan
    Maison-Blanche, Pierre
    Thomas, Olivier
    JOURNAL OF ELECTROCARDIOLOGY, 2018, 51 (03) : 531 - 533
  • [49] Is it safe to allow patients with implantable cardioverter-defibrillators to drive? Learnings from a single center experience
    Curnis, Antonio
    Mascioli, Giosue
    Bontempi, Luca
    Cerini, Manuel
    Bignotti, Tommaso
    Bonetti, Gabriele
    Cas, Livio Dei
    JOURNAL OF CARDIOVASCULAR MEDICINE, 2008, 9 (12) : 1241 - 1245
  • [50] Atrial leads and SVT discriminators do not decrease the incidence inappropriate shocks in pediatric patients with Implantable cardioverter-defibrillators
    Cannon, B
    Friedman, RA
    Fenrich, AL
    LeWellen, M
    Kim, JJ
    Minor, M
    Kertesz, NJ
    CIRCULATION, 2005, 112 (17) : U610 - U610