Vaginal Applicators in High-Dose Rate Brachytherapy for Endometrial Carcinoma: A Dosimetric Study

被引:0
|
作者
Khan, Aijaz Ahmad [1 ,2 ]
Vijay, Anuj [1 ]
Wani, Shaqul Qamar [3 ]
Haq, Malik Mohib-ul [2 ]
机构
[1] GLA Univ, Inst Appl Sci & Humanities, Dept Phys, Chaumuhan, Uttar Pradesh, India
[2] Sherikashmir Inst Med Sci, Dept Radiol Phys & Bioengn, Srinagar, India
[3] Sherikashmir Inst Med Sci, Dept Radiat Oncol, Srinagar, India
来源
TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY | 2023年 / 38卷 / 02期
关键词
Brachytherapy; Elasto-Gel; overestimation of dose; treatment planning system; vaginal applicators; EXTERNAL-BEAM RADIOTHERAPY; QUALITY-OF-LIFE; RADIATION-THERAPY; STAGE-I; CANCER; PORTEC-2; TRIAL; AAPM; IR-192;
D O I
10.5505/tjo.2022.3765
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
OBJECTIVE Intracavitary brachytherapy is one of the most important methods of gynecological cancer treatment. The effect of attenuation by the applicators is not considered in the dose calculation method released by the AAPM TG-43. In this study, the attenuation due to various segmented vaginal cylinders made up of polyetheretherketone was measured. METHODS A plan created by the treatment planning system (TPS) for the patient was executed using Gafchromic films that were taped on the surface of the applicator, and the dose recorded on the films was measured. The same plan was also executed using a well-type ionization chamber. First, a water equivalent material, Elasto-Gel was used in a well-type ionization chamber for charge collection, and then a segmented vaginal cylinder was used for charge collection measurement; thus, the attenuation due to the various segmented vaginal cylinders was obtained. RESULTS The doses recorded on the Gafchromic films were measured and showed that the TPS overestimated the dose for the segmented vaginal applicators and that the maximum overestimation of the dose was 4.75% for 35 mm diameter vaginal cylinders. This was further confirmed by calculating the percentage attenuation due to vaginal cylinders using a well-type ionization chamber. CONCLUSION Due to the attenuation by the various applicators used in vaginal vault brachytherapy, it can be concluded that due to overestimation of dose by the TPS, the dose actually received by the target and the organs at risk in vaginal vault brachytherapy are less and needs to be considered for actual dose estimation.
引用
收藏
页码:191 / 200
页数:10
相关论文
共 50 条
  • [21] A unique approach to high-dose-rate vaginal mold brachytherapy of gynecologic malignancies
    Nilsson, Sanna
    Moutrie, Zoe
    Cheuk, Robyn
    Chan, Philip
    Lancaster, Craig
    Markwell, Tim
    Dawes, Jodi
    Back, Phil
    BRACHYTHERAPY, 2015, 14 (02) : 267 - 272
  • [22] Daily schedule for high-dose-rate brachytherapy in postoperative treatment of endometrial carcinoma
    Rovirosa, Angeles
    Valduvieco, Izaskun
    Ascaso, Carlos
    Herreros, Antonio
    Bautista, Carlos
    Romera, Ingrid
    Arenas, Meritxell
    Pahisa, Jaume
    Biete, Albert
    CLINICAL & TRANSLATIONAL ONCOLOGY, 2013, 15 (02) : 111 - 116
  • [23] EROS study: evaluation between high-dose-rate and low-dose-rate vaginal interventional radiotherapy (brachytherapy) in terms of overall survival and rate of stenosis
    Autorino, Rosa
    Tagliaferri, Luca
    Campitelli, Maura
    Smaniotto, Daniela
    Nardangeli, Alessia
    Mattiucci, Gian Carlo
    Macchia, Gabriella
    Gui, Benedetta
    Micco, Maura
    Mascilini, Floriana
    Ferrandina, Gabriella
    Kovacs, Gyorgy
    Valentini, Vincenzo
    Gambacorta, Maria Antonietta
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2018, 10 (04) : 315 - 320
  • [24] A SMALL TOLERANCE FOR CATHETER DISPLACEMENT IN HIGH-DOSE RATE PROSTATE BRACHYTHERAPY IS NECESSARY AND FEASIBLE
    Tiong, Albert
    Bydder, Sean
    Ebert, Martin
    Caswell, Nikki
    Waterhouse, David
    Spry, Nigel
    Camille, Peter
    Joseph, David
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (04): : 1066 - 1072
  • [25] Assessment of air pockets in high-dose-rate vaginal cuff brachytherapy using cylindrical applicators
    Hassouna, Ashraf
    Bahadur, Yasir Abdulaziz
    Constantinescu, Camelia
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2014, 6 (03) : 271 - 275
  • [26] Process Mapping and Time Study to Improve Efficiency of New Procedure Implementation for High-Dose Rate Prostate Brachytherapy
    Doyle, Laura A.
    Yondorf, Menachem
    Peng, Cheng
    Harrison, Amy S.
    Den, Robert B.
    JOURNAL FOR HEALTHCARE QUALITY, 2018, 40 (01) : 19 - 26
  • [27] High-dose-rate brachytherapy for high-grade vaginal intraepithelial neoplasia: a dosimetric analysis
    Barcellini, Amelia
    Lecchi, Mara
    Tenconi, Chiara
    Macciotta, Alessandra
    Pignoli, Emanuele
    Pappalardi, Brigid
    Mozzarella, Ester
    Carrara, Mauro
    Giandini, Tommaso
    Fallai, Carlo
    Verderio, Paolo
    Cerrotta, Annamaria
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2019, 11 (02) : 146 - 151
  • [28] 3D-printed applicators for high dose rate brachytherapy: Dosimetric assessment at different infill percentage
    Ricotti, Rosalinda
    Vavassori, Andrea
    Bazani, Alessia
    Ciardo, Delia
    Pansini, Floriana
    Spoto, Ruggero
    Sammarco, Vittorio
    Cattani, Federica
    Baroni, Guido
    Orecchia, Roberto
    Jereczek-Fossa, Barbara Alicja
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (12): : 1698 - 1706
  • [29] Salvage high-dose-rate brachytherapy for isolated vaginal recurrence of endometrial cancer
    Baek, Sungjae
    Isohashi, Fumiaki
    Yamaguchi, Hiroko
    Mabuchi, Seiji
    Yoshida, Ken
    Kotsuma, Tadayuki
    Yamazaki, Hideya
    Tanaka, Eiichi
    Sumida, Iori
    Tamari, Keisuke
    Otani, Keisuke
    Seo, Yuji
    Suzuki, Osamu
    Yoshioka, Yasuo
    Kimura, Tadashi
    Ogawa, Kazuhiko
    BRACHYTHERAPY, 2016, 15 (06) : 812 - 816
  • [30] Vaginal toxicity after high-dose-rate endovaginal brachytherapy: 20 years of results
    Delishaj, Durim
    Barcellini, Amelia
    D'Amico, Romerai
    Ursino, Stefano
    Pasqualetti, Francesco
    Fumagalli, Ilaria Costanza
    Soatti, Carlo Pietro
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2018, 10 (06) : 559 - 566