The influence of surgical technique guidance and surgeon's experience on the femoral head assembly in total hip arthroplasty

被引:1
作者
Darowski, Martin [1 ]
Ruehrmund, Leo [1 ]
Kluess, Daniel [1 ]
Klinder, Annett [1 ]
Bader, Rainer [1 ]
Mittelmeier, Wolfram [1 ]
机构
[1] Rostock Univ Med Ctr, Dept Orthopaed, Doberaner Str 142, D-18057 Rostock, Germany
关键词
Total hip arthroplasty; Taper connection; Femoral head; Assembly procedure; Surgical technique guide; Morse taper; ASSISTED CREVICE CORROSION; FRETTING CORROSION; NECK JUNCTION; TAPER; STRENGTH; STEM; INTERFACE; IMPACTION; FIXATION; WEAR;
D O I
10.1007/s00402-024-05282-w
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction The importance of the assembly procedure on the taper connection strength is evident. However, existent surgical technique guides frequently lack comprehensive and precise instructions in this regard. The aim of our experimental study was to evaluate the influence of the surgical technique guide on the femoral head assembly procedure in surgeons with differing levels of experience in total hip arthroplasty.Materials and methods Twenty-eight participants, divided into four groups based on their lifetime experience in total hip arthroplasty, conducted a femoral head assembly procedure in a simulated intraoperative environment before and after reviewing the surgical technique guide. Demographic information and the number of hammer blows were documented. Hammer velocity and impaction angle were recorded using an optical motion capturing system, while the impaction force was measured using a dynamic force sensor within the impactor.Results We observed a high variation in the number of hammer blows, maximum force, and impaction angle. Overall, the number of hammer blows decreased significantly from 3 to 2.2 after reviewing the surgical technique guide. The only significant intragroup difference in the number of hammer blows was observed in the group with no prior experience in total hip arthroplasty. No correlation was found between individual factors (age, weight, height) or experience and the measured parameters (velocity, maximum force and angle).Conclusions The present study demonstrated a high variation in the parameters of the femoral head assembly procedure. Consideration of the surgical technique guide was found to be a limited factor among participants with varying levels of experience in total hip arthroplasty. These findings underline the importance of sufficient preoperative training, to standardize the assembly procedure, including impaction force, angle, and use of instruments.
引用
收藏
页码:2391 / 2401
页数:11
相关论文
共 40 条
[1]  
[Anonymous], 2015, STANDARD TEST METHOD
[2]   Gross Trunnion Failure After Primary Total Hip Arthroplasty [J].
Banerjee, Samik ;
Cherian, Jeffrey J. ;
Bono, James V. ;
Kurtz, Steven M. ;
Geesink, Rudolph ;
Meneghini, R. Michael ;
Delanois, Ronald E. ;
Mont, Michael A. .
JOURNAL OF ARTHROPLASTY, 2015, 30 (04) :641-648
[3]  
Ben-Shlomo Y., 2021, National Joint Registry Annual Reports
[4]  
Brial C, 2016, 62TH ANN M ORTH RES
[5]  
Brial C, 2017, 63 ANN MEET ORTH RES
[6]   The effect of different assembly loads on taper junction fretting wear in total hip replacements [J].
English, Russell ;
Ashkanfar, Ariyan ;
Rothwell, Glynn .
TRIBOLOGY INTERNATIONAL, 2016, 95 :199-210
[7]   The Role of the Assembly Force in the Tribocorrosion Behaviour of Hip Implant Head-Neck Junctions: An Adaptive Finite Element Approach [J].
Fallahnezhad, Khosro ;
Feyzi, Mohsen ;
Hashemi, Reza ;
Taylor, Mark .
BIOENGINEERING-BASEL, 2022, 9 (11)
[8]  
Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149
[9]  
Grostefon JD, 2016, 62 ANN MEET ORTH RES
[10]   What Surgeons Need to Know About Adverse Local Tissue Reaction in Total Hip Arthroplasty [J].
Hall, Deborah J. ;
Pourzal, Robin ;
Jacobs, Joshua J. .
JOURNAL OF ARTHROPLASTY, 2020, 35 (06) :S55-S59