An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of phenobarbital in human serum and plasma

被引:1
|
作者
Schierscher, Tobias [1 ]
Salzmann, Linda [1 ]
Singh, Neeraj [2 ]
Bachmann, Martina [2 ]
Kobel, Anja [1 ]
Wild, Janik [1 ]
Bauland, Friederike [3 ]
Geistanger, Andrea [2 ]
Risch, Lorenz [1 ]
Geletneky, Christian [2 ]
Seger, Christoph [1 ]
Taibon, Judith [2 ,4 ]
机构
[1] Dr Risch Ostschweiz AG, Buchs, Switzerland
[2] Roche Diagnost GmbH, Penzberg, Germany
[3] Chrestos Concept GmbH & Co KG, Essen, Germany
[4] Roche Diagnost GmbH, Nonnenwald 2, D-82377 Penzberg, Germany
关键词
phenobarbital; ID-LC-MS/MS; reference measurement procedure; SI units; traceability; ANTIEPILEPTIC DRUGS; SPECIFICATIONS; PERFORMANCE; METABOLITES; IONIZATION; EPILEPSY; EFFICACY; POSITION; BLOOD; MS;
D O I
10.1515/cclm-2023-1104
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Phenobarbital serves as an antiepileptic drug (AED) and finds application in the treatment of epilepsy either as monotherapy or adjunctive therapy. This drug exhibits various pharmacodynamic properties that account for its beneficial effects as well as potential side effects. Accurate measurement of its concentration is critical for optimizing AED therapy through appropriate dose adjustments. Therefore, our objective was to develop and validate a new reference measurement procedure (RMP) for the accurate quantification of phenobarbital levels in human serum and plasma. Methods: A sample preparation protocol based on protein precipitation followed by a high dilution step was established in combination with a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method using a C8 column to separate target analytes from known and unknown interferences. Assay validation and determination of measurement uncertainty were performed based on current guidelines. Selectivity and Specificity were assessed using spiked serum and plasma samples; to investigate possible matrix effects (MEs) a post-column infusion experiment and a comparison of standard line slopes was performed. Precision and accuracy were determined within a multiday precision experiment. Results: The RMP was shown to be highly selective and specific, with no evidence of matrix interferences. It can be used to quantify phenobarbital in the range of 1.92 to 72.0 mu g/mL. Intermediate precision was less than 3.2 %, and repeatability coefficient of variation (CV) ranged from 1.3 to 2.0 % across all concentration levels. The relative mean bias ranged from -3.0 to -0.7 % for native serum levels, and from -2.8 to 0.8 % for Li-heparin plasma levels. The measurement uncertainties (k=1) for single measurements and target value assignment were 1.9 to 3.3 % and 0.9 to 1.6 %, respectively. Conclusions: A novel LC-MS/MS-based candidate RMP for the quantification of phenobarbital in human serum and plasma is presented which can be used for the standardization of routine assays and the evaluation of clinically relevant samples.
引用
收藏
页码:1314 / 1326
页数:13
相关论文
共 50 条
  • [41] A simple and reliable measurement procedure for determination of glycocholic acid in human serum by isotope-dilution liquid chromatography-tandem mass spectrometry
    Wang, Ahui
    Xu, Yuzhu
    Ou, Yangfen
    Han, Liqiao
    Zhang, Qiaoxuan
    Zhan, Min
    Liu, Hongcan
    Ke, Peifeng
    Yan, Jun
    Huang, Xianzhang
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2021, 464
  • [42] Accurate determination of chloramphenicol in meat by isotope dilution liquid chromatography mass spectrometry (ID-LC/MS)
    Song, Hyemin
    Kim, Byungjoo
    Jeong, Hyuk
    Ahn, Seonghee
    ANALYTICAL SCIENCE AND TECHNOLOGY, 2010, 23 (06) : 524 - 530
  • [43] LIQUID CHROMATOGRAPHY-ELECTROSPRAY TANDEM MASS SPECTROMETRY (LC-MS/MS) DETERMINATION OF LANSOPRAZOLE IN HUMAN PLASMA
    Ganesh, M.
    BhagiyaLakshmi, M.
    Hemalatha, P.
    Patil, Rahul
    Sakthimanigandan, K.
    Jang, Hyun Tae
    Rajasekar, K.
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2011, 34 (02) : 129 - 142
  • [44] Comparison of bracketing calibration and classical calibration curve quantification methods in establishing a candidate reference measurement procedure for human serum 17β-estradiol by isotope dilution liquid chromatography tandem mass spectrometry
    Zhang, Qiaoxuan
    Cai, Zhiliang
    Liu, Yingyi
    Lin, Haibiao
    Wang, Qiqin
    Yan, Jun
    Han, Liqiao
    Wang, Jianbing
    Ke, Peifeng
    Zhuang, Junhua
    Huang, Xianzhang
    MICROCHEMICAL JOURNAL, 2020, 152
  • [45] Quantitation of Phenolic Benzotriazole Class Compounds in Plasma by Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS)
    Mutlu, Esra
    South, Natalie
    Pierfelice, Jessica
    Djonabaye, Alison
    Pauff, Mindy
    Burback, Brian
    Waidyanatha, Suramya
    ANALYTICAL LETTERS, 2022, 55 (13) : 2074 - 2088
  • [46] Measurement of serum 17-hydroxyprogesterone using isotope dilution liquid chromatography-tandem mass spectrometry candidate reference method and evaluation of the performance for three routine methods
    Long, Qichen
    Zhang, Tianjiao
    Yan, Ying
    Zhao, Haijian
    Zhou, Weiyan
    Zeng, Jie
    Li, Shuijun
    Zhang, Jiangtao
    Zeng, Qingzhang
    Zhao, Beibei
    Zhang, Chuanbao
    Chen, Wenxiang
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2021, 59 (03) : 523 - 532
  • [47] An isotope dilution LC-MS/MS based candidate reference method for the quantification of cyclosporine A, tacrolimus, sirolimus and everolimus in human whole blood
    Taibon, Judith
    van Rooij, Milou
    Schmid, Rupert
    Singh, Neeraj
    Albrecht, Eva
    Wright, Jo Anne
    Geletneky, Christian
    Schuster, Carina
    Moerlein, Sophie
    Vogeser, Michael
    Seger, Christoph
    Pongratz, Stephan
    Kobold, Uwe
    CLINICAL BIOCHEMISTRY, 2020, 82 : 73 - 84
  • [48] Advances in Liquid Chromatography-Tandem Mass Spectrometry (LC-MS-MS)-Based Quantitation of Biopharmaceuticals in Biological Samples
    van de Merbel, Nico C.
    LC GC EUROPE, 2015, : 38 - 44
  • [49] Measurement of serum progesterone by isotope dilution liquid chromatography tandem mass spectrometry: a candidate reference method and its application to evaluating immunoassays
    Wang, Yufei
    Zhang, Tianjiao
    Zhao, Haijian
    Zhou, Weiyan
    Zeng, Jie
    Zhang, Jiangtao
    Zhang, Chuanbao
    Chen, Wenxiang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2019, 411 (11) : 2363 - 2371
  • [50] Measurement of serum progesterone by isotope dilution liquid chromatography tandem mass spectrometry: a candidate reference method and its application to evaluating immunoassays
    Yufei Wang
    Tianjiao Zhang
    Haijian Zhao
    Weiyan Zhou
    Jie Zeng
    Jiangtao Zhang
    Chuanbao Zhang
    Wenxiang Chen
    Analytical and Bioanalytical Chemistry, 2019, 411 : 2363 - 2371