An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of phenobarbital in human serum and plasma

被引:1
|
作者
Schierscher, Tobias [1 ]
Salzmann, Linda [1 ]
Singh, Neeraj [2 ]
Bachmann, Martina [2 ]
Kobel, Anja [1 ]
Wild, Janik [1 ]
Bauland, Friederike [3 ]
Geistanger, Andrea [2 ]
Risch, Lorenz [1 ]
Geletneky, Christian [2 ]
Seger, Christoph [1 ]
Taibon, Judith [2 ,4 ]
机构
[1] Dr Risch Ostschweiz AG, Buchs, Switzerland
[2] Roche Diagnost GmbH, Penzberg, Germany
[3] Chrestos Concept GmbH & Co KG, Essen, Germany
[4] Roche Diagnost GmbH, Nonnenwald 2, D-82377 Penzberg, Germany
关键词
phenobarbital; ID-LC-MS/MS; reference measurement procedure; SI units; traceability; ANTIEPILEPTIC DRUGS; SPECIFICATIONS; PERFORMANCE; METABOLITES; IONIZATION; EPILEPSY; EFFICACY; POSITION; BLOOD; MS;
D O I
10.1515/cclm-2023-1104
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Phenobarbital serves as an antiepileptic drug (AED) and finds application in the treatment of epilepsy either as monotherapy or adjunctive therapy. This drug exhibits various pharmacodynamic properties that account for its beneficial effects as well as potential side effects. Accurate measurement of its concentration is critical for optimizing AED therapy through appropriate dose adjustments. Therefore, our objective was to develop and validate a new reference measurement procedure (RMP) for the accurate quantification of phenobarbital levels in human serum and plasma. Methods: A sample preparation protocol based on protein precipitation followed by a high dilution step was established in combination with a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method using a C8 column to separate target analytes from known and unknown interferences. Assay validation and determination of measurement uncertainty were performed based on current guidelines. Selectivity and Specificity were assessed using spiked serum and plasma samples; to investigate possible matrix effects (MEs) a post-column infusion experiment and a comparison of standard line slopes was performed. Precision and accuracy were determined within a multiday precision experiment. Results: The RMP was shown to be highly selective and specific, with no evidence of matrix interferences. It can be used to quantify phenobarbital in the range of 1.92 to 72.0 mu g/mL. Intermediate precision was less than 3.2 %, and repeatability coefficient of variation (CV) ranged from 1.3 to 2.0 % across all concentration levels. The relative mean bias ranged from -3.0 to -0.7 % for native serum levels, and from -2.8 to 0.8 % for Li-heparin plasma levels. The measurement uncertainties (k=1) for single measurements and target value assignment were 1.9 to 3.3 % and 0.9 to 1.6 %, respectively. Conclusions: A novel LC-MS/MS-based candidate RMP for the quantification of phenobarbital in human serum and plasma is presented which can be used for the standardization of routine assays and the evaluation of clinically relevant samples.
引用
收藏
页码:1314 / 1326
页数:13
相关论文
共 50 条
  • [31] State-of-the-art of serum testosterone measurement by isotope dilution-liquid chromatography-tandem mass spectrometry
    Thienpont, Linda M.
    Van Uytfanghe, Katleen
    Blincko, Stuart
    Ramsay, Carol S.
    Xie, Hui
    Doss, Robert C.
    Keevil, Brian G.
    Owen, Laura J.
    Rockwood, Alan L.
    Kushnir, Mark M.
    Chun, Kelly Y.
    Chandler, Donald W.
    Field, Helen P.
    Sluss, Patrick M.
    CLINICAL CHEMISTRY, 2008, 54 (08) : 1290 - 1297
  • [32] Development of a simplified sample pretreatment procedure as part of an isotope dilution-liquid chromatography/tandem mass spectrometry candidate reference measurement procedure for serum total thyroxine
    Van Uytfanghe, K
    Stöckl, D
    Thienpont, LM
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (13) : 1539 - 1540
  • [33] Evaluation of a bracketing calibration-based isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma
    Zhang, Qiaoxuan
    Zhang, Lu
    Lin, Haibiao
    Cai, Zhiliang
    Yan, Jun
    Wang, Qiqin
    Han, Liqiao
    Wang, Jianbing
    Ke, Peifeng
    Zhuang, Junhua
    Huang, Xianzhang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2019, 411 (27) : 7095 - 7104
  • [34] An isotope dilution LC-MS/MS-based candidate reference method for the quantification of androstenedione in human serum and plasma
    Gradl, Katrin
    Taibon, Judith
    Singh, Neeraj
    Albrecht, Eva
    Geistanger, Andrea
    Pongratz, Stephan
    Hutzler, Stefan
    Mayer, Magdalena
    Kleinschmidt, Christine
    Geletneky, Christian
    Hofmann, Verena
    Koeppl, Daniel
    Rauh, Manfred
    Kobold, Uwe
    CLINICAL MASS SPECTROMETRY, 2020, 16 : 1 - 10
  • [35] Quantification of mycophenolic acid in plasma by isotope dilution liquid chromatography-tandem mass spectrometry candidate reference methodQuantification of mycophenolic acid in plasma by isotope dilution liquid chromatography-tandem mass spectrometry candidate reference methodJ. Lin et al.
    Jing Lin
    Min Shen
    Ting Yu
    Huimin Wang
    Jingjue Pan
    Gaipeng Huang
    Quanle Li
    Analytical and Bioanalytical Chemistry, 2025, 417 (8) : 1607 - 1617
  • [36] ESTIMATION OF MELOXICAM IN HUMAN PLASMA BY LIQUID CHROMATOGRAPHY-TANDEM MASS SPECTROMETRY (LC-MS/MS)
    Ganesh, Mani
    Hemalatha, Pushparaj
    Peng, Mei Mei
    Sakthimanigandan, Kalaimani
    Jang, Hyun Tae
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2013, 36 (07) : 867 - 880
  • [37] Simultaneous measurement of testosterone, androstenedione and dehydroepiandrosterone (DHEA) in serum and plasma using Isotope-Dilution 2-Dimension Ultra High Performance Liquid-Chromatography Tandem Mass Spectrometry (ID-LC-MS/MS)
    Buttler, Rahel M.
    Martens, Frans
    Kushnir, Mark M.
    Ackermans, Mariette T.
    Blankenstein, Marinus A.
    Heijboer, Annemieke C.
    CLINICA CHIMICA ACTA, 2015, 438 : 157 - 159
  • [38] Development of Triglyceride Certified Reference Materials in Human Frozen Serum Using Isotope Dilution-Liquid Chromatography-Tandem Mass Spectrometry
    Lee, Sun Young
    Kim, Kwonseong
    Kwon, Hyun Soo
    Kim, Young Eun
    Kim, Ki Young
    Jeong, Ji-Seon
    Oh, Han Bin
    Kim, Tae Young
    Hong, Jongki
    Kang, Dukjin
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2019, 40 (05) : 418 - 423
  • [39] Accurate Quantification of Saccharin Using Isotope Dilution Liquid Chromatography Mass Spectrometry (ID-LC/MS)
    Lee, Yunjung
    Kim, Byungjoo
    Kim, Jeongkwon
    Ahn, Seonghee
    MASS SPECTROMETRY LETTERS, 2011, 2 (02) : 37 - 40
  • [40] Isotope-Dilution Liquid Chromatography-Tandem Mass Spectrometry Candidate Reference Method for Total Testosterone in Human Serum
    Botelho, Julianne Cook
    Shacklady, Christopher
    Cooper, Hans C.
    Tai, Susan S. -C.
    Van Uytfanghe, Katleen
    Thienpont, Linda M.
    Vesper, Hubert W.
    CLINICAL CHEMISTRY, 2013, 59 (02) : 372 - 380