Subcutaneous implantable cardioverter-defibrillator implantation position predicts successful defibrillation in obese and non-obese patients

被引:7
|
作者
Hoster, Clay [1 ]
Rahman, Annas [1 ]
Goyal, Ansh [2 ]
Peigh, Graham [2 ]
Trohman, Richard [1 ]
Knight, Bradley P. P. [2 ]
Huang, Henry [1 ]
Krishnan, Kousik [3 ]
Larsen, Timothy [1 ]
Mazur, Alexander [1 ]
Sharma, Parikshit [1 ]
Engelstein, Erica [1 ]
Verma, Nishant [2 ]
Wasserlauf, Jeremiah [1 ]
机构
[1] Rush Univ, Div Cardiol, Med Ctr, 1717 W Congress Pkwy, 345 Kellogg, Chicago, IL 60612 USA
[2] Northwestern Univ, Feinberg Sch Med, Div Cardiol, Chicago, IL USA
[3] Midwest Cardiovasc Inst, Naperville, IL USA
关键词
ICD; S-ICD; Defibrillation; Obesity; PRAETORIAN;
D O I
10.1007/s10840-022-01462-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Subcutaneous implantable cardioverter-defibrillators (S-ICD) are an alternative to transvenous ICDs for patients without a need for cardiac pacing. Obese patients have been proposed to be at higher risk for conversion failure with S-ICDs due to subcutaneous fat underneath the device. Optimal device positioning may promote equivalent outcomes between obese and non-obese patients by minimizing the effects of excess adipose tissue. Methods A retrospective analysis of patients undergoing defibrillation testing at the time of S-ICD implantation was performed. The primary endpoint was the rate of successful conversion of ventricular fibrillation (VF) at the time of implant. The secondary endpoint was shock impedance. Results A total of 184 patients were included in the study. The rate of successful conversion of VF was 90.3% for obese patients (n = 72) and 96.4% for non-obese patients (n = 112) (p = 0.086). Compared to non-obese patients, obese patients had a higher mean PRAETORIAN score (78.5 +/- 58.1 vs. 48.8 +/- 35.5, p < 0.001) and higher measured mean impedance (82.0 ohms +/- 26.5 vs. 69.8 ohms +/- 19.3, p < 0.001). Patients with a PRAETORIAN score < 90 all had successful defibrillation testing regardless of BMI. Conclusions In this study, a PRAETORIAN score < 90 was associated with a 100% success rate of defibrillation testing following S-ICD implantation regardless of patient body mass index (BMI). Thus, the impact of obesity on impedance and the risk of failed shocks may be minimized with close attention to implantation technique to achieve a low PRAETORIAN score.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Implantation of the Subcutaneous Implantable Cardioverter-Defibrillator An Evaluation of 4 Implantation Techniques
    Brouwer, Tom F.
    Miller, Marc A.
    Quast, Anne-Floor B. E.
    Palaniswamy, Chandrasekar
    Dukkipati, Srinivas R.
    Reddy, Vivek
    Wilde, Arthur A.
    Willner, Jonathan M.
    Knops, Reinoud E.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2017, 10 (01):
  • [22] New surgical approach to implantation of the subcutaneous implantable cardioverter-defibrillator
    Kennergren, Charles E. H.
    EUROPACE, 2017, 19 (12): : 1907 - 1908
  • [23] CONSISTENT SUBCUTANEOUS PREPECTORAL IMPLANTATION OF A NEW IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    STANTON, MS
    HAYES, DL
    MUNGER, TM
    TRUSTY, JM
    ESPINOSA, RE
    SHEN, WK
    OSBORN, MJ
    PACKER, DL
    HAMMILL, SC
    MAYO CLINIC PROCEEDINGS, 1994, 69 (04) : 309 - 314
  • [24] Effect of defibrillation testing on management during implantable cardioverter-defibrillator implantation
    Leong-Sit, Peter
    Gula, Lorne J.
    Diamantouros, Pantelis
    Krahn, Andrew D.
    Skanes, Allan C.
    Yee, Raymond
    Klein, George J.
    AMERICAN HEART JOURNAL, 2006, 152 (06) : 1104 - 1108
  • [25] Implantation of the subcutaneous implantable cardioverter-defibrillator with truncal plane blocks
    Miller, Marc A.
    Bhatt, Himani V.
    Weiner, Menachem
    Brouwer, Tom F.
    Mittnacht, Alexander J.
    Shariat, Ali
    Jeng, Christina L.
    Eden, Caroline
    Lin, Hung-Mo
    Salter, Benjamin
    Dukkipati, Srinivas R.
    Reddy, Vivek Y.
    HEART RHYTHM, 2018, 15 (07) : 1108 - 1111
  • [26] Defibrillation threshold testing: Is it necessary during implantable cardioverter-defibrillator implantation?
    Liu, Qi-Ming
    Bai, Zhong-Le
    Liu, Zhen-Jiang
    Li, Xu-Ping
    Zhou, Sheng-Hua
    MEDICAL HYPOTHESES, 2009, 72 (02) : 147 - 149
  • [27] Totally submuscular implantation of subcutaneous implantable cardioverter defibrillator: a safe and effective solution for obese or oversized patients
    Droghetti, Andrea
    Locatelli, Alessandro
    Casiraghi, Bruno
    Malacrida, Maurizio
    Arupi, Michele
    Ragusa, Mark
    CLINICAL CASE REPORTS, 2016, 4 (10): : 1009 - 1011
  • [28] CHRONIC DEFIBRILLATION THRESHOLD TESTING IN PATIENTS UNDERGOING SUBCUTANEOUS IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR GENERATOR CHANGE
    Patel, Raj
    Shahab, Ahmed
    Broman, Michael
    Aziz, Zaid
    Beaser, Andrew D.
    Upadhyay, Gaurav A.
    Kannan, Arun
    Jameria, Zenith
    Chen, Andrew
    Ozcan, Cevher
    Tung, Roderick
    Nayak, Hemal M.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 75 (11) : 538 - 538
  • [29] Defibrillation threshold decrease with the supradiaphragmatic extracardiac implantable cardioverter-defibrillator implantation technique
    Berruezo, Antonio
    Zeljko, Hrvojka M.
    Bartrons, Joaquim
    Mayol, Javier
    Prada, Fredy
    Brugada, Josep
    EUROPACE, 2010, 12 (11): : 1649 - 1651
  • [30] AWARENESS DURING ANESTHESIA FOR IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR IMPLANTATION - RECALL OF DEFIBRILLATION SHOCKS
    MOERMAN, A
    HERREGODS, L
    FOUBERT, L
    POELAERT, J
    JORDAENS, L
    DHONT, L
    ROLLY, G
    ANAESTHESIA, 1995, 50 (08) : 733 - 735