Intentional deep overfit learning for patient-specific dose predictions in adaptive radiotherapy

被引:11
作者
Maniscalco, Austen [1 ]
Liang, Xiao [1 ]
Lin, Mu-Han [1 ]
Jiang, Steve [1 ]
Nguyen, Dan [1 ]
机构
[1] Univ Texas Southwestern Med Ctr, Dept Radiat Oncol, Med Artificial Intelligence & Automat Lab, Dallas, TX 75390 USA
关键词
adaptive; artificial intelligence; deep learning; dose prediction; fine-tuning; head and neck cancer; overfit; radiation therapy; NECK-CANCER; RADIATION-THERAPY; HEAD;
D O I
10.1002/mp.16616
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundThe framework of adaptive radiation therapy (ART) was crafted to address the underlying sources of intra-patient variation that were observed throughout numerous patients' radiation sessions. ART seeks to minimize the consequential dosimetric uncertainty resulting from this daily variation, commonly through treatment planning re-optimization. Re-optimization typically consists of manual evaluation and modification of previously utilized optimization criteria. Ideally, frequent treatment plan adaptation through re-optimization on each day's computed tomography (CT) scan may improve dosimetric accuracy and minimize dose delivered to organs at risk (OARs) as the planning target volume (PTV) changes throughout the course of treatment. PurposeRe-optimization in its current form is time-consuming and inefficient. In response to this ART bottleneck, we propose a deep learning based adaptive dose prediction model that utilizes a head and neck (H & N) patient's initial planning data to fine-tune a previously trained population model towards a patient-specific model. Our fine-tuned, patient-specific (FT-PS) model, which is trained using the intentional deep overfit learning (IDOL) method, may enable clinicians and treatment planners to rapidly evaluate relevant dosimetric changes daily and re-optimize accordingly. MethodsAn adaptive population (AP) model was trained using adaptive data from 33 patients. Separately, 10 patients were selected for training FT-PS models. The previously trained AP model was utilized as the base model weights prior to re-initializing model training for each FT-PS model. Ten FT-PS models were separately trained by fine-tuning the previous model weights based on each respective patient's initial treatment plan. From these 10 patients, 26 ART treatment plans were withheld from training as the test dataset for retrospective evaluation of dose prediction performance between the AP and FT-PS models. Each AP and FT-PS dose prediction was compared against the ground truth dose distribution as originally generated during the patient's course of treatment. Mean absolute percent error (MAPE) evaluated the dose differences between a model's prediction and the ground truth. ResultsMAPE was calculated within the 10% isodose volume region of interest for each of the AP and FT-PS models dose predictions and averaged across all test adaptive sessions, yielding 5.759% and 3.747% respectively. MAPE differences were compared between AP and FT-PS models across each test session in a test of statistical significance. The differences were statistically significant in a paired t-test with two-tailed p-value equal to 3.851x10-9$3.851 \times {10}<<^>>{ - 9}$ and 95% confidence interval (CI) equal to [-2.483, -1.542]. Furthermore, MAPE was calculated using each individually segmented structure as an ROI. Nineteen of 24 structures demonstrated statistically significant differences between the AP and FT-PS models. ConclusionWe utilized the IDOL method to fine-tune a population-based dose prediction model into an adaptive, patient-specific model. The averaged MAPE across the test dataset was 5.759% for the population-based model versus 3.747% for the fine-tuned, patient-specific model, and the difference in MAPE between models was found to be statistically significant. Our work demonstrates the feasibility of patient-specific models in adaptive radiotherapy, and offers unique clinical benefit by utilizing initial planning data that contains the physician's treatment intent.
引用
收藏
页码:5354 / 5363
页数:10
相关论文
共 13 条
  • [1] Quantification of volumetric and geometric changes occurring during fractionated radiotherapy for head-and-neck cancer using an integrated CT/linear accelerator system
    Barker, JL
    Garden, AS
    Ang, KK
    O'Daniel, JC
    Wang, H
    Court, LE
    Morrison, WH
    Rosenthal, DI
    Chao, KSC
    Tucker, SL
    Mohan, R
    Dong, L
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (04): : 960 - 970
  • [2] Varian ethos online adaptive radiotherapy for prostate cancer: Early results of contouring accuracy, treatment plan quality, and treatment time
    Byrne, Mikel
    Archibald-Heeren, Ben
    Hu, Yunfei
    Teh, Amy
    Beserminji, Rhea
    Cai, Emma
    Liu, Guilin
    Yates, Angela
    Rijken, James
    Collett, Nick
    Aland, Trent
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (01):
  • [3] Intentional deep overfit learning (IDOL): A novel deep learning strategy for adaptive radiation therapy
    Chun, Jaehee
    Park, Justin C.
    Olberg, Sven
    Zhang, You
    Nguyen, Dan
    Wang, Jing
    Kim, Jin Sung
    Jiang, Steve
    [J]. MEDICAL PHYSICS, 2022, 49 (01) : 488 - 496
  • [4] 3D radiotherapy dose prediction on head and neck cancer patients with a hierarchically densely connected U-net deep learning architecture
    Dan Nguyen
    Jia, Xun
    Sher, David
    Lin, Mu-Han
    Iqbal, Zohaib
    Liu, Hui
    Jiang, Steve
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (06)
  • [5] Automatic treatment planning based on three-dimensional dose distribution predicted from deep learning technique
    Fan, Jiawei
    Wang, Jiazhou
    Chen, Zhi
    Hu, Chaosu
    Zhang, Zhen
    Hu, Weigang
    [J]. MEDICAL PHYSICS, 2019, 46 (01) : 370 - 381
  • [6] Offline adaptive radiotherapy for bladder cancer using cone beam computed tomography
    Foroudi, F.
    Wong, J.
    Haworth, A.
    Baille, A.
    McAlpine, J.
    Rolfo, A.
    Kron, T.
    Roxby, P.
    Paneghel, A.
    Williams, S.
    Duchesne, G.
    Tai, K. H.
    [J]. JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2009, 53 (02) : 226 - 233
  • [7] Deep Learning-Based Dose Prediction for Automated, Individualized Quality Assurance of Head and Neck Radiation Therapy Plans
    Gronberg, Mary P.
    Beadle, Beth M.
    Garden, Adam S.
    Skinner, Heath
    Gay, Skylar
    Netherton, Tucker
    Cao, Wenhua
    Cardenas, Carlos E.
    Chung, Christine
    Fuentes, David T.
    Fuller, Clifton D.
    Howell, Rebecca M.
    Jhingran, Anuja
    Lim, Tze Yee
    Marquez, Barbara
    Mumme, Raymond
    Olanrewaju, Adenike M.
    Peterson, Christine B.
    Vazquez, Ivan
    Whitaker, Thomas J.
    Wooten, Zachary
    Yang, Ming
    Court, Laurence E.
    [J]. PRACTICAL RADIATION ONCOLOGY, 2023, 13 (03) : E282 - E291
  • [8] Image-guided radiation therapy: Physicians perspectives
    Gupta, T.
    Narayan, C. Anand
    [J]. JOURNAL OF MEDICAL PHYSICS, 2012, 37 (04) : 174 - 182
  • [9] Clinical implementation of artificial intelligence-driven cone-beam computed tomography-guided online adaptive radiotherapy in the pelvic region
    Sibolt, Patrik
    Andersson, Lina M.
    Calmels, Lucie
    Sjostrom, David
    Bjelkengren, Ulf
    Geertsen, Poul
    Behrens, Claus F.
    [J]. PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 17 : 1 - 7
  • [10] Sonke J-J., SEM RAD ONC 2019