Estimation of Effective Dose in Monte Carlo Simulation Method for CT Coronary Angiography Patients

被引:0
|
作者
Sarma, Arnabjyoti Deva [1 ]
Singha, Mrinal Kanti [1 ,2 ,3 ]
Sharma, Jibon [2 ,3 ]
Kashyap, Manash Pratim [4 ]
机构
[1] Assam downtown Univ, Dept Radiol, Gauhati, Assam, India
[2] Down Town Hosp, Gauhati, Assam, India
[3] Gauhati Med Coll, State Canc Inst, Gauhati, Assam, India
[4] Assam Down Town Univ, Dept Stat, Gauhati, Assam, India
来源
INTERNATIONAL JOURNAL OF LIFE SCIENCE AND PHARMA RESEARCH | 2023年 / 13卷 / 02期
关键词
Radiation; Computed Tomography; Effective Dose; Angiography; Monte Carlo Simulation; COMPUTED-TOMOGRAPHY EXAMINATIONS; DIAGNOSTIC REFERENCE LEVELS; MULTISLICE;
D O I
10.22376/ijlpr.2023.13.2.L194-L201
中图分类号
R914 [药物化学];
学科分类号
100701 ;
摘要
Computed tomography (CT) scans of Coronary Angiography have significantly aided in the diagnosis of disorders of the coronary artery. In contrast to other radiological examinations, the patient's radiation exposure is notably higher. This study aimed to optimize the radiation dose and estimate the adequate amount in computed tomography (CT) for coronary angiography. A total of 380 patients were referred to the Primus Diagnostic Centre and Heath City Hospital, Guwahati Assam, during the study with coronary artery disturbances. Data on the technical parameters used in CT procedures were taken in 2022. The aim and study's objective was Organ and surface dose to specific radiosensitive organs (Chest) estimation using software imPACT 1.0.4 from National Radiological Protection Board (NRPB) SR250 Monte Carlo dataset. The study population (n = 380) comprised 190 men and 190 women with an average age of 29 to 75 years. The Mean +/- SD of BMI and ED are 22.42 +/- 1.06 and 21.57 +/- 4.27 respectively. The mean DLP is 854.67, and the mean ED is 21.57. The effective doses for males (13-27) mSv were in females (13-29) mSv. This study was a pioneer in presenting actual amounts of CT examinations in Assam because other countries have already started with more advanced CT procedures, such as dosages for paediatrics, coronary angiography and CT fluoroscopy. With this study, there may be more opportunities to create complex new studies or enhance the data from related studies that may be done in future work. The high precision with minimum risk, the current study can be considered as the need of the hour.
引用
收藏
页码:L194 / L201
页数:8
相关论文
共 50 条
  • [21] Monte Carlo simulation of bowtie filter scatter on a wide-cone low-dose CT system
    Liu, Xin
    Srivastava, Anjali
    Lee, Hyoung-Koo
    Hsieh, Jiang
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [22] ESTIMATION OF SYSTEM RELIABILITY BY MONTE CARLO SIMULATION
    Naess, A.
    Leira, B. J.
    Batsevych, O.
    OMAE 2009, VOL 2: STRUCTURES, SAFETY AND RELIABILITY, 2009, : 421 - 426
  • [23] Organ Doses, Effective Dose, and Cancer Risk From Coronary CT Angiography Examinations
    Mahmoodi, Maede
    Chaparian, Ali
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 214 (05) : 1131 - 1136
  • [24] Comparing organ and effective dose of various CT localizer acquisition strategies: A Monte Carlo study
    Oostveen, Luuk J.
    Tunissen, Sjoerd
    Sechopoulos, Ioannis
    MEDICAL PHYSICS, 2024, : 576 - 584
  • [25] Radiation dose reduction in CT coronary angiography
    Shapiro B.P.
    Young P.M.
    Kantor B.
    Choe Y.H.
    McCollough C.H.
    Gerber T.C.
    Current Cardiology Reports, 2010, 12 (1) : 59 - 67
  • [26] Comparison of organ and effective dose estimations from different Monte Carlo simulation-based software methods in infant CT and comparison with direct phantom measurements
    Lawson, Michael
    Berk, Kemal
    Badawy, Mohamed
    Qi, Yujin
    Kuganesan, Ahilan
    Metcalfe, Peter
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (06):
  • [27] Photon dose estimation from ultraintense laser-solid interactions and shielding calculation with Monte Carlo simulation
    Yang, Bo
    Qiu, Rui
    Li, Junli
    Lu, Wei
    Wu, Zhen
    Li, Chunyan
    RADIATION PHYSICS AND CHEMISTRY, 2017, 131 : 13 - 21
  • [28] Dose reduction during CT coronary angiography
    Willoteaux, S.
    Sibileau, E.
    Caroff, J.
    Nedelcu, C.
    Kalil, W. Abi
    Delepine, S.
    Thouveny, F.
    JOURNAL DE RADIOLOGIE, 2010, 91 (11): : 1220 - 1224
  • [29] The effect of overranging on patient effective dose and organs' doses during parathyroid CT examinations utilising Monte Carlo software
    Dimitroukas, Christos P.
    Metaxas, Vasileios I.
    Efthymiou, Fotios O.
    Kalogeropoulou, Christina P.
    Zampakis, Petros E.
    Panayiotakis, George S.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 212
  • [30] A new Monte Carlo tool for organ dose estimation in computed tomography
    Adrien, C.
    Le Loirec, C.
    Dreuil, S.
    Bordy, J. -M.
    RADIOPROTECTION, 2020, 55 (02) : 123 - 134