A cryogenic 14-channel 13C receiver array for 3T human head imaging

被引:6
|
作者
Wang, Wenjun [1 ]
Sanchez-Heredia, Juan Diego [2 ]
Olin, Rie Beck [2 ]
Hansen, Esben Sovso Szocska [3 ]
Laustsen, Christoffer [3 ]
Zhurbenko, Vitaliy [1 ]
Ardenkjaer-Larsen, Jan Henrik [2 ]
机构
[1] Tech Univ Denmark, Natl Space Inst, Lyngby, Denmark
[2] Tech Univ Denmark, Dept Hlth Technol, Lyngby, Denmark
[3] Aarhus Univ, MR Res Ctr, Dept Clin Med, Aarhus, Denmark
关键词
carbon imaging; cryogenic coil; head coil; MRI array; HIGH-TEMPERATURE SUPERCONDUCTOR; SURFACE COIL; MRI; PERFORMANCE; MOUSE; SENSITIVITY; MICROSCOPY; DESIGN; PROBES; SNR;
D O I
10.1002/mrm.29508
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This article presents a novel 14-channel receive-only array for C-13 human head imaging at 3 T that explores the SNR gain by operating at cryogenic temperature cooled by liquid nitrogen. Methods Cryostats are developed to evaluate single-coil bench SNR performance and cool the 14-channel array with liquid nitrogen while having enough thermal insulation between the coils and the sample. The temperature distribution for the coil array is measured. Circuits are adapted to the -189 degrees C environment and implemented in the 14-channel array. C-13 images are acquired with the array at cryogenic and room temperature in a 3T scanner. Results Compared with room temperature, the array at cryogenic temperature provides 27%-168% SNR improvement over all voxels and 47% SNR improvement near the image center. The measurements show a decrease of the element noise correlation at cryogenic temperature. Conclusion It is demonstrated that higher SNR can be achieved by cryogenically cooling the 14-channel array. A cryogenic array suitable for clinical imaging can be further developed on the array proposed. The cryogenic coil array is most likely suited for scenarios in which high SNR deep in a head and decent SNR on the periphery are required.
引用
收藏
页码:1265 / 1277
页数:13
相关论文
共 24 条
  • [21] Technique development of 3D dynamic CS-EPSI for hyperpolarized 13C pyruvate MR molecular imaging of human prostate cancer
    Chen, Hsin-Yu
    Larson, Peder E. Z.
    Gordon, Jeremy W.
    Bok, Robert A.
    Ferrone, Marcus
    van Criekinge, Mark
    Carvajal, Lucas
    Cao, Peng
    Pauly, John M.
    Kerr, Adam B.
    Park, Ilwoo
    Slater, James B.
    Nelson, Sarah J.
    Munster, Pamela N.
    Aggarwal, Rahul
    Kurhanewicz, John
    Vigneron, Daniel B.
    MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (05) : 2062 - 2072
  • [22] A Flexible 32-channel Receive Array Combined With a Homogeneous Transmit Coil for Human Lung Imaging With Hyperpolarized 3He at 1.5 T
    Deppe, Martin H.
    Parra-Robles, Juan
    Marshall, Helen
    Lanz, Titus
    Wild, Jim M.
    MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (06) : 1788 - 1797
  • [23] A metabolite-specific 3D stack-of-spiral bSSFP sequence for improved lactate imaging in hyperpolarized [1-13C]pyruvate studies on a 3T clinical scanner
    Tang, Shuyu
    Bok, Robert
    Qin, Hecong
    Reed, Galen
    VanCriekinge, Mark
    Delos Santos, Romelyn
    Overall, William
    Santos, Juan
    Gordon, Jeremy
    Wang, Zhen Jane
    Vigneron, Daniel B.
    Larson, Peder E. Z.
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (03) : 1113 - 1125
  • [24] Quantitative 7T sodium magnetic resonance imaging of the human brain using a 32-channel phased-array head coil: Application to patients with secondary progressive multiple sclerosis
    Wilferth, Tobias
    Mennecke, Angelika
    Gast, Lena, V
    Lachner, Sebastian
    Mueller, Max
    Rothhammer, Veit
    Huhn, Konstantin
    Uder, Michael
    Doerfler, Arnd
    Nagel, Armin M.
    Schmidt, Manuel
    NMR IN BIOMEDICINE, 2022, 35 (12)