Development of Simple HPLC-UV Method for the Simultaneous Determination of Repaglinide, Dexamethasone, and Remdesivir, and its Application to Synthetic Mixture and Human Plasma

被引:0
作者
Smerikarova, Miglena [1 ]
Bozhanov, Stanislav [1 ]
Mateeva, Alexandrina [2 ]
Maslarska, Vania [1 ]
机构
[1] Med Univ Sofia, Fac Pharm, Dept Chem, Dunav Str 2, Sofia 1000, Bulgaria
[2] Med Univ Sofia, Fac Pharm, Dept Pharmaceut Chem, Dunav Str 2, Sofia 1000, Bulgaria
关键词
HPLC; bioanalysis; remdesivir; dexamethasone; repaglinide; plasma; BULK DRUG; COVID-19; VALIDATION; MORTALITY;
D O I
10.2174/0115734129263384230928052923
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Background: The onset of the COVID-19 pandemic caused numerous difficulties in the treatment of cardiovascular diseases and diabetes mellitus. A persistent risk of developing severe complications and increased mortality from the COVID-19 infection has been reported. In the clinical studies, patients receiving remdesivir and dexamethasone as COVID-19 combination therapy simultaneously with some type II diabetes therapeutic regimens had been reported to have a considerably better state and recover faster. Unfortunately, there is not enough information on the combination of meglitinides, remdesivir, and dexamethasone, and therefore, careful monitoring of the patients' everyday health condition is needed.Objectives: The present study aimed to describe a high-performance liquid chromatographic method for the determination of repaglinide, dexamethasone, and remdesivir in laboratoryprepared mixtures and human plasma by UV detection.Methods: Isocratic elution of the mobile phase (consisting of 0.1% trifluoroacetic acid in water and acetonitrile in the ratio 70:30 v/v) was set at a flow rate of 1.0 ml/min, and the developed analytical procedure has been found to be fast and simple. Chromatographic determination was performed on a Purospher (R) RP - 18 column at room temperature and a UV detector was set at 235 nm.Results: The developed method was validated for linearity in the range 2-32 mu g/ml. Calibration curves were linear over the selected range with correlation coefficients (R-2) greater than 0.996. The coefficients of variation for intraday and interday assay were < 2% and the recovery percentages from plasma ranged from 93.83 to 106.49%.Conclusion: The developed effective and specific method can be applied in routine quality control and clinical laboratory practice.
引用
收藏
页码:662 / 672
页数:11
相关论文
共 50 条
  • [21] Development and Validation of HPLC-UV Method for the Determination of a Potent Synthetic Cannabinoid THJ-2201 in Mouse Plasma and Application in a Pharmacokinetic Study
    Aboul-Enein, Hassan Y.
    Mostafa, Gamal A. E.
    AlRabiah, Haitham
    Al-Ramadi, Mohammed
    Attia, Sabry M.
    Wagdy, Hebatallah A.
    CURRENT PHARMACEUTICAL ANALYSIS, 2020, 16 (04) : 404 - 411
  • [22] A Simple and Sensitive HPLC-UV Method for the Estimation of Ofloxacin in Plasma
    Hussain, Niaz
    Shah, Nisar Hussain
    Khurram Syed, Shahzada
    Rasool, Muhammad F.
    Nisar, Naveed
    Majeed, Abdul
    LATIN AMERICAN JOURNAL OF PHARMACY, 2021, 40 (11): : 2731 - 2734
  • [23] HPLC-UV Method for Simultaneous Determination of Moxifloxacin and Prednisolone Acetate in Eye Drops
    Sireesha, K. R.
    Prakash, K.
    ASIAN JOURNAL OF CHEMISTRY, 2013, 25 (03) : 1255 - 1258
  • [24] Development of rapid and sensitive method for determination of mycophenolic acid in human plasma by HPLC-UV
    Pawinski, T.
    Szlaska, I
    Wakulicz, A.
    THERAPEUTIC DRUG MONITORING, 2011, 33 (04) : 545 - 545
  • [25] Development and Validation of an HPLC-UV Method for Anilocaine Determination in Blood Plasma
    Ryzhikova, V. A.
    Kursakov, S. V.
    Belov, V. Yu.
    Sevast'yanov, V. I.
    PHARMACEUTICAL CHEMISTRY JOURNAL, 2018, 52 (02) : 166 - 170
  • [26] Development and Validation of an HPLC-UV Method for Anilocaine Determination in Blood Plasma
    V. A. Ryzhikova
    S. V. Kursakov
    V. Yu. Belov
    V. I. Sevast’yanov
    Pharmaceutical Chemistry Journal, 2018, 52 : 166 - 170
  • [27] Development and validation of an HPLC-UV detection assay for the determination of rufinamide in human plasma and saliva
    Mazzucchelli, Iolanda
    Rapetti, Manuela
    Fattore, Cinzia
    Franco, Valentina
    Gatti, Giuliana
    Perucca, Emilio
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 401 (03) : 1013 - 1021
  • [28] DEVELOPMENT AND VALIDATION OF HPLC-UV METHOD FOR THE DETERMINATION OF CLOPIDOGREL IN PHARMACEUTICAL DOSAGE FORM AND HUMAN PLASMA
    Javed, Muhammad. K.
    Iqbal, Zafar
    Khan, Abbas
    Khan, Abad
    Shah, Yasar
    Ahmad, Lateef
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2011, 34 (18) : 2118 - 2129
  • [29] A fast and simple HPLC-UV method for simultaneous determination of three anti-cancer agents in plasma of breast cancer patients and its application to clinical pharmacokinetics
    Ahmad, Mahmood
    Usman, Muhammad
    Madni, Asadullah
    Zubair, M.
    Qamar-uz-Zaman
    Qureshi, M. Saleem
    Munir, Abubakar
    Ahmad, Mahtab
    Mahmood, Arshad
    AFRICAN JOURNAL OF PHARMACY AND PHARMACOLOGY, 2011, 5 (07): : 915 - 922
  • [30] Simplified HPLC-UV method for the determination of α-tocopherol in plasma
    Renzi, M
    Righi, F
    Quarantelli, C
    Quarantelli, A
    Bonomi, A
    ITALIAN JOURNAL OF ANIMAL SCIENCE, 2005, 4 (02) : 191 - 195