Biomechanical analysis of the door-shaped titanium plate in single-level anterior cervical discectomy and fusion

被引:0
|
作者
Li, Senli [1 ]
Yan, Peng [1 ]
Fan, Yanwei [1 ]
Wang, Ruibo [1 ]
Zhang, Changjiang [1 ]
机构
[1] Zhengzhou Univ, Dept Orthoped 2, Affiliated Hosp 5, Zhengzhou 450052, Peoples R China
关键词
Anterior cervical discectomy and fusion; Door-shaped titanium plate; Range of motion; Neutral zone; Biomechanical study; ADJACENT-SEGMENT PATHOLOGY; FOLLOW-UP; INTERBODY FUSION; SPONDYLOTIC MYELOPATHY; ILIAC-GRAFT; SPINE; CAGE; OSSIFICATION; DISC; DYSPHAGIA;
D O I
10.1186/s13018-023-04474-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Analyse and discuss the immediate stability of the cervical spine after anterior cervical discectomy and fusion using a door-shaped titanium plate and compare it with the traditional titanium plate, to provide biomechanical evidence for the rationality and effectiveness of the door-shaped titanium plate in clinical applications.Methods Ten adult goat C4/5 vertebral bodies were obtained, and models were prepared using denture base resin. Biomechanical experiments were performed on the specimens before internal fixation. MTS was used to conduct non-destructive biomechanical loading tests in six directions, including flexion, extension, left-right bending, and left-right torsion, recording the range of motion (ROM) and neutral zone (NZ) of each specimen. The specimens were then randomly divided into two groups: the study group was fixed with a door-shaped titanium plate, and the control group was fixed with a traditional titanium plate. ROM and NZ in each direction were measured again. After measurements, both groups were subjected to 0.5 Hz torsion loading with a torque of 2 N m for a total of 3000 cycles, followed by measuring ROM and NZ in six directions once more.Results Compared to before fixation, ROM and NZ in both groups significantly decreased in all six directions after fixation, with statistical significance (P < 0.05); after fixation, the study group showed slightly lower values for various mechanical reference parameters compared to the control group, with no statistical significance (P > 0.05); after 3000 torsional loads, both internal fixation groups showed increased ROM and NZ compared to after fixation but to a lower extent, and no screw or titanium plate loosening was observed. Compared to before fixation, the differences were still statistically significant (P < 0.05), with the study group having slightly lower ROM and NZ values in all directions compared to the control group, with no statistical significance (P > 0.05).Conclusion The door-shaped titanium plate exhibits mechanical properties similar to the traditional titanium plate in all directions, and its smaller size and simpler surgical operation can be used for anterior cervical endoscopic surgery, reducing surgical trauma. It is clinically feasible and deserves further research and promotion.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Biomechanical analysis of the door-shaped titanium plate in single-level anterior cervical discectomy and fusion
    Senli Li
    Peng Yan
    Yanwei Fan
    Ruibo Wang
    Changjiang Zhang
    Journal of Orthopaedic Surgery and Research, 18
  • [2] Biomechanical performance of the novel assembled uncovertebral joint fusion cage in single-level anterior cervical discectomy and fusion: A finite element analysis
    Zhang, Xiang
    Yang, Yi
    Shen, Yi-Wei
    Zhang, Ke-Rui
    Ma, Li-Tai
    Ding, Chen
    Wang, Bei-Yu
    Meng, Yang
    Liu, Hao
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2023, 11
  • [3] Adjacent-Level Ossification Development in Single-Level Standalone Anterior Cervical Discectomy and Fusion Versus Anterior Cervical Discectomy and Fusion With Plate
    Huang, Christopher
    Mobbs, Ralph
    Selby, Michael
    Phan, Kevin
    Rao, Prashanth
    GLOBAL SPINE JOURNAL, 2021, 11 (03) : 292 - 298
  • [4] The Impact of Menopause on Bone Fusion after the Single-Level Anterior Cervical Discectomy and Fusion
    Park, Sung Bae
    Chung, Chun Kee
    Lee, Sang Hyung
    Yang, Hee-Jin
    Son, Young-Je
    Chung, Young Seob
    JOURNAL OF KOREAN NEUROSURGICAL SOCIETY, 2013, 54 (06) : 496 - 500
  • [5] Predictive Factors of Postoperative Dysphagia in Single-Level Anterior Cervical Discectomy and Fusion
    Vaishnav, Avani S.
    Saville, Philip
    McAnany, Steven
    Patel, Dil
    Haws, Brittany
    Khechen, Benjamin
    Singh, Kern
    Gang, Catherine Himo
    Qureshi, Sheeraz A.
    SPINE, 2019, 44 (07) : E400 - E407
  • [6] The effect of cervical plating on single-level anterior cervical discectomy and fusion
    Wang, JC
    McDonough, PW
    Endow, K
    Kanim, LEA
    Delamarter, RB
    JOURNAL OF SPINAL DISORDERS, 1999, 12 (06): : 467 - 471
  • [7] Risk Factors for Postoperative Subsidence of Single-Level Anterior Cervical Discectomy and Fusion The Significance of the Preoperative Cervical Alignment
    Lee, Young-Seok
    Kim, Young-Baeg
    Park, Seung-Won
    SPINE, 2014, 39 (16) : 1280 - 1287
  • [8] Does smoking have an impact on fusion rate in single-level anterior cervical discectomy and fusion with allograft and rigid plate fixation?
    Luszczyk, Myles
    Smith, Justin S.
    Fischgrund, Jeffrey S.
    Ludwig, Steven C.
    Sasso, Rick C.
    Shaffrey, Christopher I.
    Vaccaro, Alexander R.
    JOURNAL OF NEUROSURGERY-SPINE, 2013, 19 (05) : 527 - 531
  • [9] Clinical and radiological comparison of the zero-profile anchored cage and traditional cage-plate fixation in single-level anterior cervical discectomy and fusion
    Zhang, Jun
    Wang, Shanxi
    Tang, Xiangyu
    Xiong, Wei
    Wu, Hua
    Liu, Chaoxu
    Li, Feng
    EUROPEAN JOURNAL OF MEDICAL RESEARCH, 2022, 27 (01)
  • [10] Comparison of radiological outcomes and complications between single-level and multilevel anterior cervical discectomy and fusion (ACDF) by using a polyetheretherketone (PEEK) cage-plate fusion system
    Lee, Han-Chung
    Chen, Chao-Hsuan
    Wu, Chih-Ying
    Guo, Jeng-Hung
    Chen, Yueh-Sheng
    MEDICINE, 2019, 98 (05)