Optimising the parameters of cochlear implant imaging with cone-beam computed tomography

被引:3
作者
Soederqvist, Samuel [1 ,2 ,3 ]
Sivonen, Ville [1 ,2 ]
Aarnisalo, Antti [1 ,2 ]
Karppi, Harri [4 ]
Sinkkonen, Saku T. [1 ,2 ,3 ]
Koivisto, Juha [5 ]
机构
[1] Helsinki Univ Hosp, Head & Neck Ctr, Dept Otorhinolaryngol Head & Neck Surg, Helsinki, Finland
[2] Univ Helsinki, Helsinki, Finland
[3] Helsinki Univ Hosp, Head & Neck Ctr, Dept Otorhinolaryngol Head & Neck Surg, Tauno Palva Lab, Helsinki, Finland
[4] Planmeca Oy, Helsinki, Finland
[5] Univ Helsinki, Dept Phys, Helsinki, Finland
关键词
TEMPORAL BONE; CT; OPTIMIZATION; DOSIMETRY; SCANNERS; QUALITY;
D O I
10.1093/rpd/ncad019
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
With computed tomography (CT), the delicate structures of the inner ear may be hard to visualise, which a cochlear implant (CI) electrode array can further complicate. The usefulness of a novel cone-beam CT device in CI recipient's inner ear imaging was evaluated and the exposure parameters were optimised to attain adequate clinical image quality at the lowest effective dose (ED). Six temporal bones were implanted with a Cochlear Slim Straight electrode array and imaged with six different imaging protocols. Contrast-to-noise ratio was calculated for each imaging protocol, and three observers evaluated independently the image quality of each imaging protocol and temporal bone. The overall image quality of the inner ear structures did not differ between the imaging protocols and the most relevant inner ear structures of CI recipient's inner ear can be visualised with a low ED. To visualise the most delicate structures in the inner ear, imaging protocols with higher radiation exposure may be required.
引用
收藏
页码:462 / 470
页数:9
相关论文
共 23 条
[1]   Effective dose reduction using collimation function in digital panoramic radiography and possible clinical implications in dentistry [J].
Benehimol, Daniel ;
Koivisto, Quha ;
Kadesjo, Nils ;
Shi, Xie-Qi .
DENTOMAXILLOFACIAL RADIOLOGY, 2018, 47 (07)
[2]   Evaluation of artifacts of cochlear implant electrodes in cone beam computed tomography [J].
Bevis, Nicholas ;
Effertz, Thomas ;
Beutner, Dirk ;
Gueldner, Christian .
EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2021, 278 (05) :1381-1386
[3]   Radiohistologic Comparison Study of Temporal Bone Specimens After Cochlear Implant Electrode Array Insertion: Is Cone-Beam CT Superior to MDCT? [J].
Burck, Iris ;
Schneider, Serena, V ;
Balster, Sven ;
Lehn, Annette ;
Yel, Ibrahim ;
Albrecht, Moritz H. ;
Helbig, Silke ;
Stoever, Timo ;
Kaltenbach, Benjamin ;
Vogl, Thomas J. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2021, 216 (03) :752-758
[4]  
Deutsches Institut fur Normung, 2016, 6868161 DIN
[5]   Intraoperative Cone-Beam Computed Tomography and Multi-Slice Computed Tomography in Temporal Bone Imaging for Surgical Treatment [J].
Erovic, Boban M. ;
Chan, Harley H. L. ;
Daly, Michael J. ;
Pothier, David D. ;
Yu, Eugene ;
Coulson, Chris ;
Lai, Philip ;
Irish, Jonathan C. .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2014, 150 (01) :107-114
[6]   Cone-beam CT versus Multidetector CT in Postoperative Cochlear Implant Imaging: Evaluation of Image Quality and Radiation Dose [J].
Helal, R. A. ;
Jacob, R. ;
Elshinnawy, M. A. ;
Othman, A., I ;
Al-Dhamari, I. M. ;
Paulus, D. W. ;
Abdelaziz, T. T. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2021, 42 (02) :362-367
[7]   Cone-beam imaging: Applications in ENT [J].
Hodez, C. ;
Griffaton-Taillandier, C. ;
Bensimon, I. .
EUROPEAN ANNALS OF OTORHINOLARYNGOLOGY-HEAD AND NECK DISEASES, 2011, 128 (02) :65-78
[8]  
International commission on Radiological Protection (ICRP), 2008, ICRP PUBL, V103, P23
[9]   Cone-beam computed tomography: Time to move from ALARA to ALADA [J].
Jaju, Prashant P. ;
Jaju, Sushma P. .
IMAGING SCIENCE IN DENTISTRY, 2015, 45 (04) :263-265
[10]   Cone-Beam CT Compared to Multi-Slice CT for the Diagnostic Analysis of Conductive Hearing Loss: A Feasibility Study [J].
Kemp, Pieter ;
Van Stralen, Jiska ;
De Graaf, Pim ;
Berkhout, Erwin ;
Van Horssen, Pepijn ;
Merkus, Paul .
JOURNAL OF INTERNATIONAL ADVANCED OTOLOGY, 2020, 16 (02) :222-+