Influence of the background in Compton camera images for proton therapy treatment monitoring

被引:3
作者
Borja-Lloret, M. [1 ]
Barrientos, L. [1 ]
Bernabeu, J. [1 ]
Lacasta, C. [1 ]
Munoz, E. [1 ]
Ros, A. [1 ]
Roser, J. [1 ]
Viegas, R. [1 ]
Llosa, G. [1 ]
机构
[1] Inst Fis Corpuscular IFIC, CSIC UV, Valencia, Spain
关键词
Compton imaging; Compton camera; proton therapy; treatment monitoring; Monte Carlo simulation; image reconstruction; background; VIVO RANGE VERIFICATION; PROMPT-GAMMA EMISSION; MONTE-CARLO; 1ST TEST; SIMULATION; PERFORMANCE; TELESCOPE; TOOLKIT; ENERGY;
D O I
10.1088/1361-6560/ace024
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Background events are one of the most relevant contributions to image degradation in Compton camera imaging for hadron therapy treatment monitoring. A study of the background and its contribution to image degradation is important to define future strategies to reduce the background in the system. Approach. In this simulation study, the percentage of different kinds of events and their contribution to the reconstructed image in a two-layer Compton camera have been evaluated. To this end, GATE v8.2 simulations of a proton beam impinging on a PMMA phantom have been carried out, for different proton beam energies and at different beam intensities. Main results. For a simulated Compton camera made of Lanthanum (III) Bromide monolithic crystals, coincidences caused by neutrons arriving from the phantom are the most common type of background produced by secondary radiations in the Compton camera, causing between 13% and 33% of the detected coincidences, depending on the beam energy. Results also show that random coincidences are a significant cause of image degradation at high beam intensities, and their influence in the reconstructed images is studied for values of the time coincidence windows from 500 ps to 100 ns. Significance. Results indicate the timing capabilities required to retrieve the fall-off position with good precision. Still, the noise observed in the image when no randoms are considered make us consider further background rejection methods.
引用
收藏
页数:16
相关论文
共 50 条
  • [31] The Effects of Compton Camera Data Acquisition and Readout Timing on PG Imaging for Proton Range Verification
    Polf, Jerimy C.
    Maggi, Paul
    Panthi, Rajesh
    Peterson, Stephen
    Mackin, Dennis
    Beddar, Sam
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2022, 6 (03) : 366 - 373
  • [32] Uncertainty analysis of a Compton camera imaging system for radiation therapy dose reconstruction
    Mundy, Daniel W.
    Herman, Michael G.
    MEDICAL PHYSICS, 2010, 37 (05) : 2341 - 2350
  • [33] Feasibility of the Compton Suppressed Gamma-ray spectrometry for proton therapy applications
    Tabrizi, Ehsan Hassanzadeh
    Ghasemi, Behjat
    Riazi, Zafar
    Jokar, Alireza
    RADIATION PHYSICS AND CHEMISTRY, 2025, 226
  • [34] Experimental evaluation of depth-encoding absorber designs for prompt-gamma Compton imaging in proton therapy
    Tsai, Hui-Yu
    Chen, Hsin-Yu
    Lee, Ming-Wei
    Wang, Ze
    Tseng, Sheng-Pin
    Hong, Ji-Hong
    Jan, Meei-Ling
    RADIATION MEASUREMENTS, 2019, 127
  • [35] Dosimetric accuracy and radiobiological implications of ion computed tomography for proton therapy treatment planning
    Meyer, Sebastian
    Kamp, Florian
    Tessonnier, Thomas
    Mairani, Andrea
    Belka, Claus
    Carlson, David J.
    Gianoli, Chiara
    Parodi, Katia
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (12)
  • [36] Design and performance evaluation of a slit-slat camera for 2D prompt gamma imaging in proton therapy monitoring: A Monte Carlo simulation study
    Malekzadeh, Etesam
    Rajabi, Hossein
    Tajik-Mansoury, Mohammad Ali
    Sabouri, Pouya
    Fiorina, Elisa
    Kalantari, Faraz
    MEDICAL PHYSICS, 2023, 50 (06) : 3701 - 3718
  • [37] Prompt Gamma Rays Detected with a BGO Block Compton Camera Reveal Range Deviations of Therapeutic Proton Beams
    Hueso-Gonzalez F.
    Pausch G.
    Petzoldt J.
    Romer K.E.
    Enghardt W.
    1600, Institute of Electrical and Electronics Engineers Inc. (01): : 76 - 86
  • [38] Dose reconstruction with Compton camera during proton therapy via subset-driven origin ensemble and double evolutionary algorithm
    Yao, Zhi-Yang
    Xiao, Yong-Shun
    Zhao, Ji-Zhong
    NUCLEAR SCIENCE AND TECHNIQUES, 2023, 34 (04)
  • [39] Dose reconstruction with Compton camera during proton therapy via subset-driven origin ensemble and double evolutionary algorithm
    Zhi-Yang Yao
    Yong-Shun Xiao
    Ji-Zhong Zhao
    Nuclear Science and Techniques, 2023, 34
  • [40] Background contributions in the electron-tracking Compton camera aboard SMILE-2+
    Ikeda, Tomonori
    Takada, Atsushi
    Takemura, Taito
    Yoshikawa, Kei
    Nakamura, Yuta
    Onozaka, Ken
    Abe, Mitsuru
    Tanimori, Toru
    Mizumura, Yoshitaka
    PHYSICAL REVIEW D, 2023, 108 (12)