摘要
Objective. Background events are one of the most relevant contributions to image degradation in Compton camera imaging for hadron therapy treatment monitoring. A study of the background and its contribution to image degradation is important to define future strategies to reduce the background in the system. Approach. In this simulation study, the percentage of different kinds of events and their contribution to the reconstructed image in a two-layer Compton camera have been evaluated. To this end, GATE v8.2 simulations of a proton beam impinging on a PMMA phantom have been carried out, for different proton beam energies and at different beam intensities. Main results. For a simulated Compton camera made of Lanthanum (III) Bromide monolithic crystals, coincidences caused by neutrons arriving from the phantom are the most common type of background produced by secondary radiations in the Compton camera, causing between 13% and 33% of the detected coincidences, depending on the beam energy. Results also show that random coincidences are a significant cause of image degradation at high beam intensities, and their influence in the reconstructed images is studied for values of the time coincidence windows from 500 ps to 100 ns. Significance. Results indicate the timing capabilities required to retrieve the fall-off position with good precision. Still, the noise observed in the image when no randoms are considered make us consider further background rejection methods.
引用
收藏
页数:16
相关论文
共 50 条
  • [11] Regularized origin ensemble with a beam prior for range verification in particle therapy with Compton-camera data
    Kasprzak, Jona
    Roser, Jorge
    Werner, Julius
    Kohlhase, Nadja
    Bolke, Andreas
    Kaufmann, Lisa-Marie
    Rafecas, Magdalena
    PHYSICS IN MEDICINE AND BIOLOGY, 2025, 70 (07)
  • [12] Secondary Particle Interactions in a Compton Camera Designed for in vivo Range Verification of Proton Therapy
    Panthi, Rajesh
    Maggi, Paul
    Peterson, Stephen
    Mackin, Dennis
    Polf, Jerimy
    Beddar, Sam
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2021, 5 (03) : 383 - 391
  • [13] Use of a LYSO-based Compton camera for prompt gamma range verification in proton therapy
    Jan, Meei-Ling
    Hsiao, Ing-Tsung
    Huang, Hsuan-Ming
    MEDICAL PHYSICS, 2017, 44 (12) : 6261 - 6269
  • [14] Compton Camera and Prompt Gamma Ray Timing: Two Methods for In Vivo Range Assessment in Proton Therapy
    Hueso-Gonzalez, Fernando
    Fiedler, Fine
    Golnik, Christian
    Kormoll, Thomas
    Pausch, Guntram
    Petzoldt, Johannes
    Roemer, Katja E.
    Enghardt, Wolfgang
    FRONTIERS IN ONCOLOGY, 2016, 6
  • [15] Prompt Gamma Imaging of a Pencil Beam with a High Efficiency Compton Camera at a Clinical Proton Therapy Facility
    Hueso-Gonzalez, F.
    Petzoldt, J.
    Roemer, K. E.
    Schoene, S.
    Fiedler, F.
    Golnik, C.
    Kormoll, T.
    Pausch, G.
    Enghardt, W.
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [16] Monitoring of positron using high-energy gamma camera for proton therapy
    Yamamoto, Seiichi
    Toshito, Toshiyuki
    Komori, Masataka
    Morishita, Yuki
    Okumura, Satoshi
    Yamaguchi, Mitsutaka
    Saito, Yuichi
    Kawachi, Naoki
    Fujimaki, Shu
    ANNALS OF NUCLEAR MEDICINE, 2015, 29 (03) : 268 - 275
  • [17] Influence of sub-nanosecond time of flight resolution for online range verification in proton therapy using the line-cone reconstruction in Compton imaging
    Livingstone, Jayde
    Dauvergne, Denis
    Etxebeste, Ane
    Fontana, Mattia
    Gallin-Martel, Marie-Laure
    Huisman, Brent
    Letang, Jean Michel
    Marcatili, Sara
    Sarrut, David
    Testa, Etienne
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (12)
  • [18] Comparison of reconstructed prompt gamma emissions using maximum likelihood estimation and origin ensemble algorithms for a Compton camera system tailored to proton range monitoring
    Lozano, Ingrid Valencia
    Dedes, George
    Peterson, Steve
    Mackin, Dennis
    Zoglauer, Andreas
    Beddar, Sam
    Avery, Stephen
    Polf, Jerimy
    Parodi, Katia
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (02): : 124 - 134
  • [19] Enhancement of Compton camera images reconstructed by inversion of a conical Radon transform
    Maxim, Voichita
    INVERSE PROBLEMS, 2019, 35 (01)
  • [20] Reconstruction of Compton-camera images using artificial neural networks
    Karg, T
    Pauli, J
    Anton, G
    Beulertz, W
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 2152 - 2155