Influence of the background in Compton camera images for proton therapy treatment monitoring

被引:3
作者
Borja-Lloret, M. [1 ]
Barrientos, L. [1 ]
Bernabeu, J. [1 ]
Lacasta, C. [1 ]
Munoz, E. [1 ]
Ros, A. [1 ]
Roser, J. [1 ]
Viegas, R. [1 ]
Llosa, G. [1 ]
机构
[1] Inst Fis Corpuscular IFIC, CSIC UV, Valencia, Spain
关键词
Compton imaging; Compton camera; proton therapy; treatment monitoring; Monte Carlo simulation; image reconstruction; background; VIVO RANGE VERIFICATION; PROMPT-GAMMA EMISSION; MONTE-CARLO; 1ST TEST; SIMULATION; PERFORMANCE; TELESCOPE; TOOLKIT; ENERGY;
D O I
10.1088/1361-6560/ace024
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Background events are one of the most relevant contributions to image degradation in Compton camera imaging for hadron therapy treatment monitoring. A study of the background and its contribution to image degradation is important to define future strategies to reduce the background in the system. Approach. In this simulation study, the percentage of different kinds of events and their contribution to the reconstructed image in a two-layer Compton camera have been evaluated. To this end, GATE v8.2 simulations of a proton beam impinging on a PMMA phantom have been carried out, for different proton beam energies and at different beam intensities. Main results. For a simulated Compton camera made of Lanthanum (III) Bromide monolithic crystals, coincidences caused by neutrons arriving from the phantom are the most common type of background produced by secondary radiations in the Compton camera, causing between 13% and 33% of the detected coincidences, depending on the beam energy. Results also show that random coincidences are a significant cause of image degradation at high beam intensities, and their influence in the reconstructed images is studied for values of the time coincidence windows from 500 ps to 100 ns. Significance. Results indicate the timing capabilities required to retrieve the fall-off position with good precision. Still, the noise observed in the image when no randoms are considered make us consider further background rejection methods.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] Noise evaluation of Compton camera imaging for proton therapy
    Ortega, P. G.
    Torres-Espallardo, I.
    Cerutti, F.
    Ferrari, A.
    Gillam, J. E.
    Lacasta, C.
    Llosa, G.
    Oliver, J. F.
    Sala, P. R.
    Solevi, P.
    Rafecas, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (05) : 1845 - 1863
  • [2] Requirements for a Compton camera for in vivo range verification of proton therapy
    Rohling, H.
    Priegnitz, M.
    Schoene, S.
    Schumann, A.
    Enghardt, W.
    Hueso-Gonzalez, F.
    Pausch, G.
    Fiedler, F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (07) : 2795 - 2811
  • [3] Machine learning-based event recognition in SiFi Compton camera imaging for proton therapy monitoring
    Kazemi Kozani, Majid
    Magiera, Andrzej
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (15)
  • [4] Proton therapy monitoring by Compton imaging: influence of the large energy spectrum of the prompt-γ radiation
    Hilaire, Estelle
    Sarrut, David
    Peyrin, Francoise
    Maxim, Voichita
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (08) : 3127 - 3147
  • [5] The SiFi-CC project - Feasibility study of a scintillation-fiber-based Compton camera for proton therapy monitoring
    Kasper, Jonas
    Rusiecka, Katarzyna
    Hetzel, Ronja
    Kozani, Majid Kazemi
    Lalik, Rafal
    Magiera, Andrzej
    Stahl, Achim
    Wronska, Aleksandra
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 76 : 317 - 325
  • [6] First Images of a Three-layer compton Telescope prototype for Treatment Monitoring in hadron Therapy
    Llosa, Gabriela
    Trovato, Marco
    Barrio, John
    Etxebeste, Ane
    Munoz, Enrique
    Lacasta, Carlos
    Oliver, Josep F.
    Rafecas, Magdalena
    Solaz, Carles
    Solevi, Paola
    FRONTIERS IN ONCOLOGY, 2016, 6
  • [7] Monitoring Ion Beam Therapy With a Compton Camera: Simulation Studies of the Clinical Feasibility
    Fontana, M.
    Ley, J. -L.
    Dauvergne, D.
    Freud, N.
    Krimmer, J.
    Letang, J. M.
    Maxim, V.
    Richard, M. -H.
    Rinaldi, I.
    Testa, E.
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2020, 4 (02) : 218 - 232
  • [8] Evaluation of a multistage CdZnTe Compton camera for prompt γ imaging for proton therapy
    McCleskey, M.
    Kaye, W.
    Mackin, D. S.
    Beddar, S.
    He, Z.
    Polf, J. C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 785 : 163 - 169
  • [9] Simulation study of Compton camera imaging for human head phantom proton therapy
    Yao, Zhiyang
    Chen, Zhiqiang
    Xiao, Yongshun
    SECOND SYMPOSIUM ON NOVEL TECHNOLOGY OF X-RAY IMAGING, 2019, 11068
  • [10] Clinical applicability of the Compton camera for Prompt γ-ray Imaging during proton therapy
    Rohling, H.
    Priegnitz, M.
    Schoene, S.
    Schumann, A.
    Enghardt, W.
    Golnik, C.
    Hueso-Gonzalez, F.
    Kormoll, T.
    Pausch, G.
    Petzoldt, J.
    Roemer, K.
    Fiedler, F.
    RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S90 - S91