Count rate correction for pulse pileup in CdZnTe photon counting detectors

被引:0
|
作者
Kang, Yang [1 ,2 ]
Wu, Rui [1 ,2 ]
Li, Peizheng [1 ,2 ]
Li, Qingpei [1 ,2 ]
Yin, Ziang [1 ,2 ]
Wu, Sen [1 ,2 ]
Tan, Tingting [1 ,2 ]
Li, Yingrui [1 ,2 ,3 ]
Zha, Gangqiang [1 ,2 ,3 ]
机构
[1] Northwestern Polytech Univ, State Key Lab Solidificat Proc, Xian 710072, Peoples R China
[2] Northwestern Polytech Univ, MIIT Key Lab Radiat Detect Mat & Devices, Xian 710072, Peoples R China
[3] Northwestern Polytech Univ, MIIT Key Lab Radiat Detect Mat & Devices, Xian 710072, Peoples R China
关键词
Photon counting; Pulse pileup; CdZnTe detector; Count rate correction; X-RAY; CDTE;
D O I
10.1016/j.mssp.2024.108142
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In the practical application of CdZnTe photon counting detectors, severe detector counting distortion can be caused by pulse pileup. A count rate correction method in this paper is proposed to solve this problem. In this method, the case where multi -photons are recorded by higher energy regions due to pulse pileup crossing the energy threshold is considered to calculate the true count rates under all energy bins. A verification method based on Poisson distribution is also proposed to evaluate the accuracy of the correction method. A 16 -pixel linear array CdZnTe photon counting detector was used for verification. Taking the dual -energy bin as an example, this method was introduced, and the true count rates of the high- and low -energy bins are successfully calculated. The result was highly consistent with the verification result, which proves the reliability of the proposed method.
引用
收藏
页数:5
相关论文
共 50 条
  • [31] Image-based spectral distortion correction for photon-counting x-ray detectors
    Ding, Huanjun
    Molloi, Sabee
    MEDICAL PHYSICS, 2012, 39 (04) : 1864 - 1876
  • [32] Single-Photon Counting Detectors
    Cova, Sergio D.
    Ghioni, Massimo
    IEEE PHOTONICS JOURNAL, 2011, 3 (02): : 274 - 277
  • [33] High Event Rate ROICs (HEROICs) for astronomical UV photon counting detectors
    Harwit, Alex
    France, Kevin
    Argabright, Vic
    Franka, Steve
    Freymiller, Ed
    Ebbets, Dennis
    HIGH ENERGY, OPTICAL, AND INFRARED DETECTORS FOR ASTRONOMY VI, 2014, 9154
  • [34] Signal to noise ratio of energy selective x-ray photon counting systems with pileup
    Alvarez, Robert E.
    MEDICAL PHYSICS, 2014, 41 (11)
  • [35] Tantalum STJ for photon counting detectors
    Jorel, C
    Feautrier, P
    Villégier, JC
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2005, 15 (02) : 587 - 590
  • [36] Investigation of the Influence of Infrared Illumination on the Pulse Shapes of Output Signals of CdZnTe Detectors
    Ivanov, Victor
    Fjodorovs, Viktors
    Hinoverovs, Sergejs
    Loutchanski, Anatoli
    Ogorodniks, Vadims
    Vidinejevs, Sergejs
    SENSORS, 2023, 23 (24)
  • [37] Photon-counting x-ray detectors for CT
    Danielsson, Mats
    Persson, Mats
    Sjolin, Martin
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03)
  • [38] Enabling Photon Counting Detectors with Dynamic Attenuators
    Hsieh, Scott S.
    Pelc, Norbert J.
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [39] Photon-counting chip for avalanche detectors
    Zappa, F
    Lotito, A
    Tisa, S
    IEEE PHOTONICS TECHNOLOGY LETTERS, 2005, 17 (01) : 184 - 186
  • [40] Vision 20/20: Single photon counting x-ray detectors in medical imaging
    Taguchi, Katsuyuki
    Iwanczyk, Jan S.