Count rate correction for pulse pileup in CdZnTe photon counting detectors

被引:0
|
作者
Kang, Yang [1 ,2 ]
Wu, Rui [1 ,2 ]
Li, Peizheng [1 ,2 ]
Li, Qingpei [1 ,2 ]
Yin, Ziang [1 ,2 ]
Wu, Sen [1 ,2 ]
Tan, Tingting [1 ,2 ]
Li, Yingrui [1 ,2 ,3 ]
Zha, Gangqiang [1 ,2 ,3 ]
机构
[1] Northwestern Polytech Univ, State Key Lab Solidificat Proc, Xian 710072, Peoples R China
[2] Northwestern Polytech Univ, MIIT Key Lab Radiat Detect Mat & Devices, Xian 710072, Peoples R China
[3] Northwestern Polytech Univ, MIIT Key Lab Radiat Detect Mat & Devices, Xian 710072, Peoples R China
关键词
Photon counting; Pulse pileup; CdZnTe detector; Count rate correction; X-RAY; CDTE;
D O I
10.1016/j.mssp.2024.108142
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In the practical application of CdZnTe photon counting detectors, severe detector counting distortion can be caused by pulse pileup. A count rate correction method in this paper is proposed to solve this problem. In this method, the case where multi -photons are recorded by higher energy regions due to pulse pileup crossing the energy threshold is considered to calculate the true count rates under all energy bins. A verification method based on Poisson distribution is also proposed to evaluate the accuracy of the correction method. A 16 -pixel linear array CdZnTe photon counting detector was used for verification. Taking the dual -energy bin as an example, this method was introduced, and the true count rates of the high- and low -energy bins are successfully calculated. The result was highly consistent with the verification result, which proves the reliability of the proposed method.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Count statistics and pileup correction for nonparalyzable photon counting detectors with finite pulse length
    Gronberg, Fredrik
    Sjolin, Martin
    Danielsson, Mats
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [2] Analytical model for pulse pileup spectra and count statistics in photon counting detectors with seminonparalyzable behavior
    Yang, Yirong
    Pelc, Norbert J.
    Wang, Adam S.
    MEDICAL PHYSICS, 2025,
  • [3] Analytical model for pulse pileup in photon counting detectors with seminonparalyzable behavior
    Yang, Yirong
    Pelc, Norbert J.
    Wang, Adam S.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [4] Sensitivity Analysis of Pulse Pileup Model Parameter in Photon Counting Detectors
    Shunhavanich, Picha
    Pelc, Norbert J.
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [5] Pulse pileup statistics for energy discriminating photon counting x-ray detectors
    Wang, Adam S.
    Harrison, Daniel
    Lobastov, Vladimir
    Tkaczyk, J. Eric
    MEDICAL PHYSICS, 2011, 38 (07) : 4265 - 4275
  • [6] Count statistics of nonparalyzable photon-counting detectors with nonzero pulse length
    Gronberg, Fredrik
    Danielsson, Mats
    Sjolin, Martin
    MEDICAL PHYSICS, 2018, 45 (08) : 3800 - 3811
  • [7] Model-based pulse pileup and charge sharing compensation for photon counting detectors: A simulation study
    Taguchi, Katsuyuki
    Polster, Christoph
    Segars, W. Paul
    Aygun, Nafi
    Stierstorfer, Karl
    MEDICAL PHYSICS, 2022, 49 (08) : 5038 - 5051
  • [8] An analytical model of the effects of pulse pileup on the energy spectrum recorded by energy resolved photon counting x-ray detectors
    Taguchi, Katsuyuki
    Frey, Eric C.
    Wang, Xiaolan
    Iwanczyk, Jan S.
    Barber, William C.
    MEDICAL PHYSICS, 2010, 37 (08) : 3957 - 3969
  • [9] Investigation on X-ray Photocurrent Response of CdZnTe Photon Counting Detectors
    Li, Yingrui
    Zha, Gangqiang
    Guo, Yu
    Xi, Shouzhi
    Xu, Lingyan
    Jie, Wanqi
    SENSORS, 2020, 20 (02)
  • [10] Effect and mechanism of temperature on photon counting performance of CdZnTe detectors
    Li Y.
    Wu S.
    Guo Y.
    Xi S.
    Fu X.
    Zha G.
    Jie W.
    Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering, 2019, 48 (10):