摘要
Background: Rotating-hinge knee (RHK) implants are an option for knee arthroplasty in cases of instability or severe bone loss. However, the revision rate can be increased compared to unconstrained implants which mandate improvements in implant design. This study investigates a novel RHK design for total knee arthroplasty (TKA). Methods: Retrospective analysis of a single design RHK TKA (GenuX MK, Implantcast) implanted between 2015 and 2019 including 133 patients (21 primary TKA, 112 revisions). The main indication for revision TKA was second-stage reimplantation following infection. The median follow-up amounted to 30 months (IQR 22-47). Results: In total, 13% (18/133) of patients underwent revision surgery after a mean time of 1 month (IQR 0-6). The main reason for the revision was (re-) infection in 8% of patients. All revisions were performed following revision TKA and none following primary TKA. There were no revision surgeries for loosening or instability. The revision-free implant survival of 83% was (95% CI 75-91%) after five years. A higher number of previous surgeries was associated with a greater revision risk (p = 0.05). Conclusion: Revision and complex primary TKA using a single-design RHK implant results in good implant survival at mid-term follow-up although infection remains a major concern.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Mid-term results of complex primary total knee arthroplasty using a rotating-hinge implant
    Spranz, David M.
    Koch, Kevin-Arno
    Reiner, Tobias
    Hetto, Pit
    Gotterbarm, Tobias
    Merle, Christian
    KNEE, 2022, 34 : 34 - 41
  • [2] Primary Rotating-Hinge Total Knee Arthroplasty: Good Outcomes at Mid-Term Follow-Up
    Kowalczewski, Jacek
    Marczak, Dariusz
    Synder, Marek
    Sibinski, Marcin
    JOURNAL OF ARTHROPLASTY, 2014, 29 (06): : 1202 - 1206
  • [3] Mid-term results and survival rates following a single-design rotating hinge knee arthroplasty in non-tumor conditions in a Pakistani population
    Nouman Memon
    Faizan Iqbal
    Syed Shahid Noor
    Kazim Rahim Najjad
    Muhammad Farhan Sozera
    Arsalan Abro
    Noman Khan
    Knee Surgery & Related Research, 33
  • [4] Mid-term results and survival rates following a single-design rotating hinge knee arthroplasty in non-tumor conditions in a Pakistani population
    Memon, Nouman
    Iqbal, Faizan
    Noor, Syed Shahid
    Najjad, Kazim Rahim
    Sozera, Muhammad Farhan
    Abro, Arsalan
    Khan, Noman
    KNEE SURGERY & RELATED RESEARCH, 2021, 33 (01)
  • [5] Rotating-hinge Total Knee for Revision Total Knee Arthroplasty
    Bistolfi, Alessandro
    Massazza, Giuseppe
    Rosso, Federica
    Crova, Maurizio
    ORTHOPEDICS, 2012, 35 (03) : E325 - E330
  • [6] Dislocation of rotating-hinge total knee arthroplasty
    Encinas-Ullan, Carlos A.
    Gomez-Cardero, Primitivo
    Ruiz-Perez, Juan S.
    Carlos Rodriguez-Merchan, E.
    EFORT OPEN REVIEWS, 2021, 6 (02) : 107 - 112
  • [7] KINEMATIC ROTATING-HINGE TOTAL KNEE ARTHROPLASTY
    RAND, JA
    CHAO, EYS
    STAUFFER, RN
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (04): : 489 - 497
  • [8] A case of an atraumatic implant failure after primary rotating-hinge total knee arthroplasty
    Petitqueux, Lucas
    Verhulst, Karen
    Dauwe, Jan
    Dauwe, Dirk
    ACTA ORTHOPAEDICA BELGICA, 2021, 87 (03): : 449 - 451
  • [9] Endo-Modell Rotating-hinge Total Knee for Revision Total Knee Arthroplasty
    Bistolfi, Alessandro
    Rosso, Federica
    Crova, Maurizio
    Massazza, Giuseppe
    ORTHOPEDICS, 2013, 36 (10) : E1299 - E1306
  • [10] Mid-term results after implantation of rotating-hinge knee prostheses: primary versus revision
    Efe, Turgay
    Roessler, Philip P.
    Heyse, Thomas J.
    Hauk, Carsten
    Pahrmann, Caroline
    Getgood, Alan
    Schmitt, Jan
    ORTHOPEDIC REVIEWS, 2012, 4 (04) : 155 - 159