Method development, validation, and application for simultaneous determination of 56 new psychoactive substances in surface water by LC-MS/MS

被引:3
|
作者
Atasoy Aydin, Asli [1 ]
Goren, Ismail Ethem [1 ]
Yavuz Guzel, Evsen [2 ]
Daglioglu, Nebile [1 ]
机构
[1] Ankara Univ, Inst Forens Sci, Dept Forens Toxicol, TR-06620 Ankara, Turkiye
[2] Cukurova Univ, Fac Fisheries, Dept Basic Sci, TR-01330 Adana, Turkiye
关键词
Synthetic cannabinoids; Synthetic cathinones; Surface water; Solid phase extraction; SOLID-PHASE EXTRACTION; INFLUENT WASTE-WATER; LIQUID-CHROMATOGRAPHY; SYNTHETIC CATHINONES; MASS-SPECTROMETRY; ILLICIT DRUGS; METABOLITES; PHARMACEUTICALS; ABUSE;
D O I
10.1007/s11356-023-28495-9
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Despite preventive legislation, the popularity and consumption of new psychoactive substances (NPS) have been steadily increasing in recent years. This study provides a rapid and sensitive method for the quantitation and the detection of 56 NPS from surface water. Sample clean-up and pre-concentration were performed by solid-phase extraction (SPE) with Oasis HLB (6 cc/500 mg) cartridge. Following the chromatographic separation with Shim-pack FC-ODS column, the all substances were quantified by liquid chromatography-tandem mass spectrometry. The method was optimized and validated for all NPS. Despite the wide variety of physicochemical properties of the analytes, the recoveries for all compounds studied were in the range of 69-117%. The limit of quantitation (LOQ) ranging from 2.5 to 15 ng/L was reached for reliable and accurate quantification of analytes. The analytical method developed was successfully applied to the surface water samples. While synthetic cannabinoids were not detected, mephedrone from the synthetic cathinone group was detected under the LOQ. This novel method was expected to be a part of future environmental routine analyses as a satisfactory method.
引用
收藏
页码:85920 / 85929
页数:10
相关论文
共 50 条
  • [41] Development And Validation Of Sensitive LC-MS/MS Method For Determination Of Doravirine In Human Plasma Samples
    Bhadru, Banothu
    Rao, V. Venkata
    Vidyadhara, Suryadevara
    INTERNATIONAL JOURNAL OF LIFE SCIENCE AND PHARMA RESEARCH, 2020, 10 (03): : P29 - P35
  • [42] Validation and application of an LC-MS/MS method for the simultaneous quantification of 13 pharmaceuticals in seawater
    Klaas Wille
    Herlinde Noppe
    Karolien Verheyden
    Julie Vanden Bussche
    Eric De Wulf
    Peter Van Caeter
    Colin R. Janssen
    Hubert F. De Brabander
    Lynn Vanhaecke
    Analytical and Bioanalytical Chemistry, 2010, 397 : 1797 - 1808
  • [43] Development and clinical application of a LC-MS/MS method for simultaneous determination of various tyrosine kinase inhibitors in human plasma
    Goetze, Lutz
    Hegele, Axel
    Metzelder, Stephan Klaus
    Renz, Harald
    Nockher, Wolfgang Andreas
    CLINICA CHIMICA ACTA, 2012, 413 (1-2) : 143 - 149
  • [44] Development and validation of a solid-phase extraction method coupled with LC-MS/MS for the simultaneous determination of 16 antibiotic residues in duck meat
    Zheng, Weijia
    Abd El-Aty, A. M.
    Kim, Seong-Kwan
    Choi, Jeong-Min
    Park, Da-Hee
    Yoo, Kyung-Hee
    Kang, Young-Sun
    Jeon, Jong-Sup
    Hacimuftuoglu, Ahmet
    Shim, Jae-Han
    Shin, Ho-Chul
    BIOMEDICAL CHROMATOGRAPHY, 2019, 33 (05)
  • [45] Development and validation of a reliable LC-MS/MS method for simultaneous determination of deoxynivalenol and T-2 toxin in maize and oats
    Tahoun, Ibrahim F.
    Gab-Allah, Mohamed A.
    Yamani, Randa N.
    Shehata, Adel B.
    MICROCHEMICAL JOURNAL, 2021, 169
  • [46] Development and Validation of a Method using SPE and LC-ESI-MS-MS for the Determination of Multiple Classes of Pesticides and Metabolites in Water Samples
    Demoliner, Adriana
    Caldas, Sergiane S.
    Costa, Fabiane P.
    Goncalves, Fabio F.
    Clementin, Rosilene M.
    Milani, Marcio R.
    Primel, Ednei G.
    JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2010, 21 (08) : 1424 - 1433
  • [47] Development of LC-MS/MS Method for the Simultaneous Detection of Emerging Contaminants in Aquatic Matrices
    Ramadan, Lama
    Ozturk-Ufuk, Irem
    Yuksel, Ebubekir
    Topuz, Emel
    WATER AIR AND SOIL POLLUTION, 2024, 235 (08)
  • [48] Development and validation of a specific method to quantify intact glucosinolates in honey by LC-MS/MS
    Ares, Ana M.
    Valverde, Silvia
    Nozal, Maria J.
    Bernal, Jose L.
    Bernal, Jose
    JOURNAL OF FOOD COMPOSITION AND ANALYSIS, 2016, 46 : 114 - 122
  • [49] Development and validation of a selective and sensitive LC-MS/MS method for determination of cycloserine in human plasma: Application to bioequivalence study
    Patel, Dinesh S.
    Sharma, Naveen
    Patel, Mukesh C.
    Patel, Bhavin N.
    Shrivastav, Pranav S.
    Sanyal, Mallika
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2011, 879 (23): : 2265 - 2273
  • [50] Development and Validation of an LC-MS/MS Method for Simultaneous Determination of Short Peptide-Based Drugs in Human Blood Plasma
    Fisher, Elizaveta N.
    Melnikov, Evgeny S.
    Gegeckori, Vladimir
    Potoldykova, Natalya, V
    Enikeev, Dmitry, V
    Pavlenko, Kirill A.
    Agatonovic-Kustrin, Snezana
    Morton, David W.
    Ramenskaya, Galina, V
    MOLECULES, 2022, 27 (22):