Reversed-Phase Liquid Chromatography Methods Based on C18 Polymer Monoliths for the Determination of Urinary Myoglobin

被引:0
|
作者
Vitek, Renan [1 ]
do Nascimento, Fernando H. [2 ]
Masini, Jorge C. [2 ]
机构
[1] Inst Fed Educ Ciencia & Tecnol Mato Grosso, BR-78043409 Cuiaba, MT, Brazil
[2] Univ Sao Paulo, Dept Quim Fundamental, Inst Quim, Ave Prof Lineu Pestes,748, BR-05508000 Sao Paulo, SP, Brazil
关键词
protein separation; narrow-bore chromatography; monolithic columns; sample preparation; flow analysis; solid phase extraction; SEQUENTIAL INJECTION CHROMATOGRAPHY; SEPARATION; EXTRACTION; PROTEINS; RHABDOMYOLYSIS; FABRICATION; COLUMNS; ALBUMIN; MUSCLE; ONLINE;
D O I
10.21577/0103-5053.20230084
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
This study describes reversed-phase liquid chromatography (RPLC) methods to quantify urinary myoglobin using polymer monolithic columns produced by copolymerization of stearyl methacrylate (SMA) and ethylene glycol dimethacrylate (EDMA). The columns were prepared in the coffins of 1.5 mm internal diameter (i.d.) ethylene tetrafluoroethylene (ETFE) tubing for use in sequential injection chromatography (SIC) and solid phase extraction (SPE), and inside 1.0 mm i.d. Silcosteel & REG; tube for use in narrow-bore liquid chromatography. The monoliths inside the ETFE were produced via UV, whereas thermal polymerization formed the monoliths inside the Silcosteel & REG; tube. The separation of carbonic anhydrase, lysozyme, and myoglobin was demonstrated because they may occur simultaneously in urine samples. Quantification was undertaken by external calibration, and the accuracy was evaluated by the spiking/recovery strategy. The methods exhibited linearity from 5.0 to 60 & mu;g mL-1 (SIC), 2.5 to 50 & mu;g mL-1 (high-performance liquid chromatography (HPLC)), and 1.0 to 7.5 & mu;g mL-1 for an SPE-HPLC method. The lowest limits of detection and quantifications were 0.13 and 0.43 jig L-1, respectively, obtained after concentrating myoglobin by SPE. Recoveries ranged from 98 to 105%. The low cost, simplicity, reusability, and analytical features provided by these polymeric stationary phases make them affordable alternatives to routine analyses of urinary myoglobin.
引用
收藏
页码:1601 / 1610
页数:10
相关论文
共 50 条
  • [1] A new imidazolium-embedded C18 stationary phase with enhanced performance in reversed-phase liquid chromatography
    Qiu, Hongdeng
    Mallik, Abul K.
    Takafuji, Makoto
    Liu, Xia
    Jiang, Shengxiang
    Ihara, Hirotaka
    ANALYTICA CHIMICA ACTA, 2012, 738 : 95 - 101
  • [2] C18 columns for the simultaneous determination of oxytetracycline and its related substances by reversed-phase high performance liquid chromatography and UV detection
    Smyrniotakis, C. G.
    Archontaki, Helen A.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 43 (02) : 506 - 514
  • [3] Practical observations on the performance of bare silica in hydrophilic interaction compared with C18 reversed-phase liquid chromatography
    Heaton, James C.
    Wang, Xiaoli
    Barber, William E.
    Buckenmaier, Stephan M. C.
    McCalley, David V.
    JOURNAL OF CHROMATOGRAPHY A, 2014, 1328 : 7 - 15
  • [4] Highly crosslinked polymeric monoliths for reversed-phase capillary liquid chromatography of small molecules
    Liu, Kun
    Tolley, H. Dennis
    Lee, Milton L.
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1227 : 96 - 104
  • [5] Peptide mapping by reversed-phase high-performance liquid chromatography employing silica rod monoliths
    Hennessy, TP
    Boysen, RI
    Huber, MI
    Unger, KK
    Hearn, MTW
    JOURNAL OF CHROMATOGRAPHY A, 2003, 1009 (1-2) : 15 - 28
  • [6] Polymer Microchips Integrating Solid-Phase Extraction and High-Performance Liquid Chromatography Using Reversed-Phase Polymethacrylate Monoliths
    Liu, Jikun
    Chen, Chien-Fu
    Tsao, Chia-Wen
    Chang, Chien-Cheng
    Chu, Chin-Chou
    DeVoe, Don L.
    ANALYTICAL CHEMISTRY, 2009, 81 (07) : 2545 - 2554
  • [7] Performance of different C18 columns in reversed-phase liquid chromatography with hydro-organic and micellar-organic mobile phases
    Ruiz-Angel, M. J.
    Pous-Torres, S.
    Carda-Broch, S.
    Garcia-Alvarez-Coque, M. C.
    JOURNAL OF CHROMATOGRAPHY A, 2014, 1344 : 76 - 82
  • [8] High efficiency polyethylene glycol diacrylate monoliths for reversed-phase capillary liquid chromatography of small molecules
    Aggarwal, Pankaj
    Lawson, John S.
    Tolley, H. Dennis
    Lee, Milton L.
    JOURNAL OF CHROMATOGRAPHY A, 2014, 1364 : 96 - 106
  • [9] Reversed phase capillary HPLC using polymer-entrapped C18 particles
    Gibson, Graham T. T.
    Marecak, Dale M.
    Oleschuk, Richard D.
    JOURNAL OF SEPARATION SCIENCE, 2009, 32 (23-24) : 4025 - 4032
  • [10] Evaluation of hydrophilic interaction chromatography (HILIC) versus C18 reversed-phase chromatography for targeted quantification of peptides by mass spectrometry
    Simon, Romain
    Enjalbert, Quentin
    Biarc, Jordane
    Lemoine, Jerome
    Salvador, Arnaud
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1264 : 31 - 39