Intra-operative Neurophysiological Monitoring in Patients Undergoing Posterior Spinal Correction Surgery with Pre-operative Neurological Deficit: Its Feasibility and High-risk Factors for Failed Monitoring

被引:5
|
作者
Liu, Wanyou [1 ]
Li, Yinkun [1 ]
Qiu, Junyin [1 ]
Shi, Benlong [1 ]
Liu, Zhen [1 ]
Sun, Xu [1 ]
Qiu, Yong [1 ]
Zhu, Zezhang [1 ]
机构
[1] Nanjing Univ, Nanjing Drum Tower Hosp, Affiliated Hosp, Med Sch,Dept Orthoped Surg,Div Spine Surg, Nanjing 210008, Peoples R China
关键词
High-risk factors; Motor evoked potentials; Neurological deficit; Somatosensory evoked potentials; Spine deformity; SOMATOSENSORY-EVOKED POTENTIALS; DECOMPRESSION; EFFICACY; STENOSIS;
D O I
10.1111/os.13914
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective Considering spinal deformity patients with pre-operative neurological deficit were associated with more intra-operative iatrogenic neurological complications than those without, intra-operative neurophysiological monitoring (IONM) has been used for detecting possible iatrogenic injury timely. However, the IONM waveforms are often unreliable. To analyze the performance of intra-operative neurophysiological monitoring (IONM) including somatosensory evoked potentials (SEP) and motor evoked potentials (MEP) in patients with pre-operative neurological deficit undergoing posterior spinal correction surgery, and to identify the high-risk factors for failed IONM.Methods Patients with pre-operative neurological deficit undergoing posterior spinal correction surgery between October 2017 and January 2022 were retrospectively reviewed. The presence or absence of SEP and MEP of target muscles were separately recorded. The P37/N50 latency and amplitude of SEP, and the MEP amplitude were measured. Any IONM alerts were also recorded. The IONM performance was compared among patients with different etiologies, levels responsible for neurological deficit, and strength of IONM-target muscles. Patients' demographics were analyzed using the descriptive statistics and were presented with mean +/- standard deviation. Comparison analysis was performed using chi(2)-test and statistically significant difference was defined as p < 0.05.Results A total of 270 patients (147 males, 123 females) with an average age of 48.4 +/- 36.7 years were involved. The SEP records were available in 371 (68.7%) lower extremities while MEP records were available in 418 (77.4%). SEP alerts were reported in 31 lower extremities and MEP alerts in 22, and new neurological deficit at post-operation was observed in 11. The etiologies of neuromuscular and syndromic indicated relatively lower success rates of IONM, which were 44.1% and 40.5% for SEP, and 58.8% and 59.5% for MEP (p < 0.001). In addition, patients with pre-operative neurological deficit caused by cervical spine and muscle strength lower than grade 4 suffered from higher risk of failed IONM waveforms (p < 0.001).Conclusion Patients with pre-operative neurological deficit suffered from a higher incidence of failed IONM results. The high-risk for failed IONM waveforms included the neuromuscular and syndromic etiologies, neurological deficit caused by cervical spine, muscle strength lower than grade 4 in patients with pre-operative neurological deficit undergoing posterior spinal correction surgery.
引用
收藏
页码:3146 / 3152
页数:7
相关论文
共 4 条
  • [1] Intra-Operative Neurophysiological Monitoring in Patients with Thoracic Spinal Stenosis: Its Feasibility and High-Risk Factors for New Neurological Deficit
    Liu, Wanyou
    Li, Yinkun
    Qiu, Junyin
    Shi, Benlong
    Liu, Zhen
    Sun, Xu
    Zhu, Zezhang
    Qiu, Yong
    ORTHOPAEDIC SURGERY, 2023, 15 (05) : 1298 - 1303
  • [2] Intra-Operative Neurophysiological Monitoring in Patients with Intraspinal Abnormalities Undergoing Posterior Spinal Fusion
    Qiu, Junyin
    Liu, Wanyou
    Shi, Benlong
    Li, Yang
    Yan, Huang
    Zhu, Zezhang
    Liu, Zhen
    Sun, Xu
    Qiu, Yong
    ORTHOPAEDIC SURGERY, 2022, 14 (08) : 1615 - 1621
  • [3] Feasibility and diagnostic accuracy of intra-operative monitoring of motor evoked potentials in children <2 years of age undergoing tethered cord surgery: results in 100 children
    Nair, Bijesh R.
    Ramamani, Mariappan
    Singh, Georgene
    Babu, Krothapalli Srinivasa
    Rajshekhar, Vedantam
    CHILDS NERVOUS SYSTEM, 2021, 37 (07) : 2289 - 2298
  • [4] Feasibility and diagnostic accuracy of intra-operative monitoring of motor evoked potentials in children <2 years of age undergoing tethered cord surgery: results in 100 children
    Bijesh R. Nair
    Mariappan Ramamani
    Georgene Singh
    Krothapalli Srinivasa Babu
    Vedantam Rajshekhar
    Child's Nervous System, 2021, 37 : 2289 - 2298