Optimal local impedance parameters for successful pulmonary vein isolation in patients with atrial fibrillation

被引:7
作者
Fukaya, Hidehira [1 ]
Mori, Hitoshi [2 ]
Oikawa, Jun [1 ]
Kawano, Daisuke [2 ]
Nakamura, Hironori [1 ]
Ishizue, Naruya [1 ]
Kishihara, Jun [1 ]
Hojo, Rintaro [3 ]
Tsutsui, Kenta [2 ]
Ikeda, Yoshifumi [2 ]
Kato, Ritsushi [2 ]
Fukamizu, Seiji [3 ]
机构
[1] Kitasato Univ, Dept Cardiovasc Med, Sch Med, 1-15-1 Kitasato,Minami Ku, Sagamihara, Kanagawa 2520374, Japan
[2] Saitama Med Univ Int Med Ctr, Dept Cardiovasc Med, Hidaka, Japan
[3] Tokyo Metropolitan Hiroo Gen Hosp, Dept Cardiol, Tokyo, Japan
关键词
atrial fibrillation; IntellaNav STABLEPOINT (TM); local impedance; pulmonary vein isolation; CATHETER ABLATION; PREDICTING STROKE;
D O I
10.1111/jce.15748
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Local impedance (LI) parameters of IntellaNav STABLEPOINT for successful pulmonary vein isolation (PVI) of atrial fibrillation (AF) remain unclear. The purpose of this study was to seek LI data achieving successful PVI. Methods: Consecutive AF patients who underwent catheter ablation with STABLEPOINT were prospectively enrolled in two centers. PVI was performed under a constant 35-or 40-watt power, 20-s duration, and > 5-g contact force. The operators were blinded to the LI data. The characteristics of all ablation points with/without conduction gaps (Unsuccess or Success tags) after the first-attempt PVI were evaluated for the right/left PVs and anterior/posterior wall (RPV/LPV and AW/PW, respectively), and cutoff values of LI data were calculated for successful lesion formation. Results: A total of 5257 ablation points in 102 patients (65 [58-72] years old, 65.7% male) were evaluated. The LI drop values were higher in the Success tags than Unsuccess tags on the LPV-AW and RPV-AW/PW (p < .001), except for the LPV-PW (p = .105). The %LI drop values (LI drop/initial LI) were higher for the Success tags in all areas (15.8 [12.2%-19.6%] vs. 11.6 [9.7%-15.6%] in LPV-AW: p < .001, 15.0 [11.5%-19.3%] vs. 11.4 [8.7%-17.3%] in LPV-PW: p = .035, 15.3 [11.5%-19.4%] vs. 9.9 [8.1%-13.7%] in RPV-AW: p < .001, and 13.3 [10.1%-17.4%] vs. 8.1 [6.3%-9.5%] in RPV-PW, p < .001). The LI drop and %LI drop cutoff values were 20.0 ohms and 11.6%, respectively. Conclusions: An insufficient LI drop with STABLEPOINT was associated with a gap formation during PVI, and the best cutoff values for the LI drop and %LI drop were 20.0 ohms and 11.6%, respectively.
引用
收藏
页码:71 / 81
页数:11
相关论文
共 24 条
[1]   Comparison of two catheters measuring local impedance: local impedance variation vs lesion characteristics and steam pops [J].
Amemiya, Miki ;
Takigawa, Masateru ;
Goya, Masahiko ;
Martin, Claire A. ;
Anzai, Tatsuhiko ;
Takahashi, Kunihiko ;
Shimizu, Yuki ;
Ikenouchi, Takashi ;
Kamata, Tatsuaki ;
Yamamoto, Tasuku ;
Nishimura, Takuro ;
Shirai, Yasuhiro ;
Tao, Susumu ;
Miyazaki, Shinsuke ;
Sasano, Tetsuo .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2022, 65 (02) :419-428
[2]   Effect of Baseline Impedance on Ablation Lesion Dimensions A Multimodality Concept Validation From Physics to Clinical Experience [J].
Barkagan, Michael ;
Rottmann, Markus ;
Leshem, Eran ;
Shen, Changyu ;
Buxton, Alfred E. ;
Anter, Elad .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2018, 11 (10)
[3]  
Calzolari Vittorio, 2017, JACC Clin Electrophysiol, V3, P1126, DOI 10.1016/j.jacep.2017.08.016
[4]   Local catheter impedance drop during pulmonary vein isolation predicts acute conduction block in patients with paroxysmal atrial fibrillation: initial results of the LOCALIZE clinical trial [J].
Das, Moloy ;
Luik, Armin ;
Shepherd, Ewen ;
Sulkin, Matthew ;
Laughner, Jacob ;
Oesterlein, Tobias ;
Duffy, Elizabeth ;
Meyer, Christian ;
Jais, Pierre ;
Duchateau, Josselin ;
Yue, Arthur ;
Ullah, Waqas ;
Ramos, Pablo ;
Garcia-Bolao, Ignacio .
EUROPACE, 2021, 23 (07) :1042-1051
[5]   Ablation index, a novel marker of ablation lesion quality: prediction of pulmonary vein reconnection at repeat electrophysiology study and regional differences in target values [J].
Das, Moloy ;
Loveday, Jonathan J. ;
Wynn, Gareth J. ;
Gomes, Sean ;
Saeed, Yawer ;
Bonnett, Laura J. ;
Waktare, Johan E. P. ;
Todd, Derick M. ;
Hall, Mark C. S. ;
Snowdon, Richard L. ;
Modi, Simon ;
Gupta, Dhiraj .
EUROPACE, 2017, 19 (05) :775-783
[6]   Validation of clinical classification schemes for predicting stroke - Results from the national registry of Atrial Fibrillation [J].
Gage, BF ;
Waterman, AD ;
Shannon, W ;
Boechler, M ;
Rich, MW ;
Radford, MJ .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2001, 285 (22) :2864-2870
[7]   Combined local impedance and contact force for radiofrequency ablation assessment [J].
Garrott, Kara ;
Laughner, Jacob ;
Gutbrod, Sarah ;
Sugrue, Alan ;
Shuros, Allan ;
Sulkin, Matt ;
Yasin, Omar ;
Bush, Jamie ;
Pottinger, Nathan ;
Meyers, Jason ;
Kapa, Suraj .
HEART RHYTHM, 2020, 17 (08) :1371-1380
[8]   A novel assessment of local impedance during catheter ablation: initial experience in humans comparing local and generator measurements [J].
Gunawardene, Melanie ;
Muenkler, Paula ;
Eickholt, Christian ;
Akbulak, Ruken Oe ;
Jularic, Mario ;
Klatt, Niklas ;
Hartmann, Jens ;
Dinshaw, Leon ;
Jungen, Christiane ;
Moser, Julia M. ;
Merbold, Lydia ;
Willems, Stephan ;
Meyer, Christian .
EUROPACE, 2019, 21 :34-42
[9]   Electrophysiological end point for catheter ablation of atrial fibrillation initiated from multiple pulmonary venous foci [J].
Haïssaguerre, M ;
Jaïs, P ;
Shah, DC ;
Garrigue, S ;
Takahashi, A ;
Lavergne, T ;
Hocini, M ;
Peng, JT ;
Roudaut, R ;
Clementy, J .
CIRCULATION, 2000, 101 (12) :1409-1417
[10]   Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins [J].
Haïssaguerre, M ;
Jaïs, P ;
Shah, DC ;
Takahashi, A ;
Hocini, M ;
Quiniou, G ;
Garrigue, S ;
Le Mouroux, A ;
Le Métayer, P ;
Clémenty, J .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (10) :659-666