Assessment of the Dosimetric Performance of the Mobius3D against Portal Dose Measurements in Patient-specific Quality Assurance

被引:1
|
作者
Labagnoy, Yasmine Joy M. [1 ]
Oonsiri, Sornjarod [2 ]
Vimolnoch, Mananchaya [2 ]
Kingkaew, Sakda [2 ]
机构
[1] Chulalongkorn Univ, Fac Med, Dept Radiol, Bangkok, Thailand
[2] King Chulalongkorn Mem Hosp, Dept Radiol, Div Radiat Oncol, Bangkok, Thailand
关键词
Electronic portal imaging device; intensity-modulated radiotherapy; Mobius3D; patient-specific quality assurance; volumetric-modulated arc therapy; IMAGING DEVICE;
D O I
10.4103/jmp.jmp_19_23
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aim: The Mobius3D software addresses limitations lacking in measurement-based methods in patient-specific quality assurance (QA). The objective of this study was to validate its dosimetric performance against conventionally used portal dose measurements using gamma analysis and confidence limits. Materials and Methods: A total of 240 patient-specific QA plans for the Varian Halcyon linear accelerator were collected. The Mobius3D software was commissioned through beam data and plan verification. All plans underwent QA through the electronic portal imaging device, coupled with the Portal Dosimetry software, and the Mobius3D. Data were assessed using >95% gamma pass. Portal measurements were evaluated using 3%/2 mm and 3%/3 mm criteria, whereas Mobius3D was analyzed at 3%/3 mm and 5%/3 mm, at the 10% threshold. Results: Mobius 5%/3 mm mean gamma passes were 99.89% for volumetric-modulated arc therapy (VMAT) and 99.31% for intensity-modulated radiotherapy (IMRT), and correspondingly, the data for portal 3%/2 mm were 99.99% and 99.96%. The Mobius3D at 5%/3 mm can perform like Portal 3%/2 mm for VMAT plans at 0.1% difference, especially for head/neck and pelvic/abdominal cases. In IMRT-based treatments, at 0.7% difference in Mobius3D 5%/3 mm and Portal 3%/2 mm, the performance and error identification in IMRT plans should be applied more carefully due to the amount of failed plans, particularly the chest region. The confidence limits for VMAT plans for Portal 3%/2 mm and Mobius 5%/3 mm are 99.93% and 99.42%, respectively, while for IMRT plans are 99.69% and 97.43%, respectively. Conclusions: At a 5%/3 mm criterion, the Mobius3D may yield percentage gamma pass rates like measurements obtained by Portal Dosimetry 3%/3 mm and Portal Dosimetry 3%/2 mm. As the software is largely dependent on commissioned data, rigorous commissioning and a comprehensive QA program should be implemented.
引用
收藏
页码:350 / 357
页数:8
相关论文
共 50 条
  • [1] Commissioning of Mobius3D/FX for patient-specific quality assurance: The CUIMC-NewYork Presbyterian Hospital experience
    Li, Fiona
    Price, Michael
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (02):
  • [2] Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation
    Basavatia, Amar K.
    Fiedler, Derek A.
    Ritter, Johannes
    Brodin, Patrik
    Goddard, Lee C.
    Jeong, Kyoungkeun
    Hsu, Shu-Hui
    Mynampati, Dinesh K.
    Yaparpalvi, Ravindra
    Tome, Wolfgang A.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (05)
  • [3] Validation and Implementation of Model-Based Patient Specific Quality Assurance Using Mobius3D and MobiusFX in a Clinical Setting
    Galavis, P.
    Osterman, K.
    Jozsef, G.
    Becker, S.
    Dewyngaert, K.
    MEDICAL PHYSICS, 2014, 41 (06) : 350 - 350
  • [4] Sensitivity of DoseCHECK and Mobius3D for patient-specific QA of Ethos Therapy on-line adaptive RT
    Almond, E.
    Fullarton, R.
    Kidane, G.
    Withers, D.
    Ward, A.
    Graham, S.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1326 - S1326
  • [5] Development of Dosimetric Verification System for Patient-Specific Quality Assurance of High-Dose-Rate Brachytherapy
    Kang, Sang-Won
    Chung, Jin-Beom
    Kim, Kyeong-Hyeon
    Choi, Chang Heon
    Kang, Seonghee
    Shin, Dong-Seok
    Cho, Woong
    Eom, Kuen-Yong
    Park, Hae-Jin
    Kim, Jin-Young
    Song, Changhoon
    Kim, In Ah
    Kim, Jae-Sung
    Suh, Tae Suk
    Park, Justin C.
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [6] Assessment of dosimetric leaf gap correction factor in Mobius3D commissioning affected by couch top
    Min Cheol Han
    Jihun Kim
    Chae-Seon Hong
    Kyung Hwan Chang
    Su Chul Han
    Kwangwoo Park
    Dong Wook Kim
    Myoung Kyu Park
    Yu Yun Noh
    Jin Sung Kim
    Physical and Engineering Sciences in Medicine, 2020, 43 : 1069 - 1075
  • [7] Validation of the Mobius system for patient-specific quality assurance using introduced intentional errors
    Ivy Win Long Au
    Laura Ciurlionis
    Neil Campbell
    Daniel Goodwin
    Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 181 - 189
  • [8] Validation of the Mobius system for patient-specific quality assurance using introduced intentional errors
    Au, Ivy Win Long
    Ciurlionis, Laura
    Campbell, Neil
    Goodwin, Daniel
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (01) : 181 - 189
  • [9] Assessment of dosimetric leaf gap correction factor in Mobius3D commissioning affected by couch top
    Han, Min Cheol
    Kim, Jihun
    Hong, Chae-Seon
    Chang, Kyung Hwan
    Han, Su Chul
    Park, Kwangwoo
    Kim, Dong Wook
    Park, Myoung Kyu
    Noh, Yu Yun
    Kim, Jin Sung
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (03) : 1069 - 1075
  • [10] Performance of the Mobius 3D software for cranial SRS patient-specific QA
    Calvo-Ortega, J.
    Hermida-Lopez, M.
    Laosa-Bello, C.
    Moragues-Femenia, S.
    Caslas-Farran, J.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1315 - S1316