Impact of computed tomography metal artifact reduction protocol on periprosthetic tissue characterization after total hip arthroplasty: A cadaveric study

被引:2
作者
Dwyer, Andrew [1 ]
Korlaet, Mishelle [1 ]
Callary, Stuart A. [2 ,3 ]
Robertson, Thomas [2 ,3 ]
Smitham, Peter [2 ,3 ]
Solomon, Lucian B. [2 ,3 ]
机构
[1] South Australian Hlth & Med Res Inst, Clin & Res Imaging Ctr, Adelaide, SA 5000, Australia
[2] Royal Adelaide Hosp, Dept Orthoped & Trauma, Adelaide, SA, Australia
[3] Univ Adelaide, Ctr Orthoped & Trauma Res, Adelaide, SA, Australia
关键词
computed tomography; dual energy; metal artifact reduction; osteolysis; tissue density; DUAL-ENERGY CT;
D O I
10.1002/jor.25391
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Metal artifact reduction (MAR) has improved computed tomography (CT) imaging of total hip arthroplasty (THA) but the assessment of osteolysis and implant to bone contact relies on the accurate depiction of bone defects, cancellous bone, and cement. This study evaluates the impact of available single and dual-energy protocols on periprosthetic tissue characterization in a cadaveric phantom. Bilateral THA was performed on a fresh frozen cadaveric pelvis with simulated osteolytic cavities. CT acquisitions with projection-based MAR and noise equivalence were performed using single energy 140 kVp, single energy 150 kVp with 0.6 mm tin filtration, and dual-energy at 100/150 kVp with 0.6 mm tin filtration, from which simulated energies were extracted. Image subtraction, segmentation, region of interest histograms, and line profiles were used to characterize tissue density and separation. Tissue densities were heavily dependent on the energy profile of the protocol. Cancellous bone ranged from 182 to 45 HU and cement from 1012 to 131 HU using 140 kVp compared to dual-energy with weighted high energy tube, respectively. Spectral separation between cancellous bone, osteolytic defect, and cement was reduced for all protocols compared with 140 kVp. Spectral overlap was most severe using dual-energy with heavily weighted high-energy tubes. Dual-energy algorithms reduced trabecular contrast within the cancellous bone and cortical edge response. Although the dual-energy acquisition has been proposed as an additive to projection-based MAR techniques in THA, reduced density and contrast in clinically relevant periprosthetic tissue compared to 140 kVp single energy may limit its use in characterizing periprosthetic tissues.
引用
收藏
页码:657 / 662
页数:6
相关论文
共 22 条
[1]   Acetabular Component Migration Measured Using Radiostereometric Analysis Following Revision Total Hip Arthroplasty A Scoping Review [J].
Abrahams, John M. ;
Callary, Stuart A. ;
Munn, Zachary ;
Jang, Sung Won ;
Huang, Qiang ;
Howie, Donald W. ;
Solomon, Lucian Bogdan .
JBJS REVIEWS, 2020, 8 (04)
[2]   Metal artefact reduction in CT imaging of hip prostheses-an evaluation of commercial techniques provided by four vendors [J].
Andersson, K. M. ;
Nowik, P. ;
Persliden, J. ;
Thunberg, P. ;
Norrman, E. .
BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1052)
[3]  
Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR), 2020, Hip, Knee & Shoulder Arthroplasty: 2020 Annual Report, P1
[4]   Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation [J].
Bamberg, Fabian ;
Dierks, Alexander ;
Nikolaou, Konstantin ;
Reiser, Maximilian F. ;
Becker, Christoph R. ;
Johnson, Thorsten R. C. .
EUROPEAN RADIOLOGY, 2011, 21 (07) :1424-1429
[5]   Comparison and Combination of Dual-Energy- and Iterative-Based Metal Artefact Reduction on Hip Prosthesis and Dental Implants [J].
Bongers, Malte N. ;
Schabel, Christoph ;
Thomas, Christoph ;
Raupach, Rainer ;
Notohamiprodjo, Mike ;
Nikolaou, Konstantin ;
Bamberg, Fabian .
PLOS ONE, 2015, 10 (11)
[6]   Do the Cup Surface Properties Influence the Initial Stability? [J].
Goriainov, Vitali ;
Jones, Andrew ;
Briscoe, Adam ;
New, Andrew ;
Dunlop, Douglas .
JOURNAL OF ARTHROPLASTY, 2014, 29 (04) :757-762
[7]   Quantifying near metal visibility using dual energy computed tomography and iterative metal artifact reduction in a fracture phantom [J].
Hakvoort, E. T. ;
Wellenberg, R. H. H. ;
Streekstra, G. J. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 69 :9-18
[8]   Metal Artifact Reduction Software Used With Abdominopelvic Dual- Energy CT of Patients With Metal Hip Prostheses: Assessment of Image Quality and Clinical Feasibility [J].
Han, Seung Chol ;
Chung, Yong Eun ;
Lee, Young Han ;
Park, Kwan Kyu ;
Kim, Myeong Jin ;
Kim, Ki Whang .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (04) :788-795
[9]   Metal Artifact Reduction in Pelvic Computed Tomography With Hip Prostheses Comparison of Virtual Monoenergetic Extrapolations From Dual-Energy Computed Tomography and an Iterative Metal Artifact Reduction Algorithmin a Phantom Study [J].
Higashigaito, Kai ;
Angst, Florian ;
Runge, Val M. ;
Alkadhi, Hatem ;
Donati, Olivio F. .
INVESTIGATIVE RADIOLOGY, 2015, 50 (12) :828-834
[10]   Increase in Total Joint Arthroplasty Projected from 2014 to 2046 in Australia: A Conservative Local Model With International Implications [J].
Inacio, Maria C. S. ;
Graves, Stephen E. ;
Pratt, Nicole L. ;
Roughead, Elizabeth E. ;
Nemes, Szilard .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2017, 475 (08) :2130-2137