Optimising total knee replacement imaging: a novel 3D printed PET/CT anthropomorphic phantom for metal artefact simulation

被引:1
作者
Assiri, Rajeh [1 ,2 ]
Knapp, Karen [2 ]
Fulford, Jon [2 ]
Chen, Junning [3 ]
机构
[1] King Khalid Univ, Dept Radiol Sci, Coll Appl Med Sci, Abha 61421, Saudi Arabia
[2] Univ Exeter, Fac Hlth & Life Sci, Dept Med Imaging, St Lukes Campus,Heavitree Rd, Exeter EX1 2LU, England
[3] Univ Exeter, Coll Engn Math & Phys Sci, Exeter, England
关键词
PET/CT; Total knee replacement; TKR; 3D printing; Phantom; Bone; RADIOLOGICAL PROPERTIES; IMAGES; BONE;
D O I
10.1186/s40658-024-00634-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Arthroplasty phantoms, including total knee replacement (TKR) phantoms, have been frequently used to test metal artefact reduction methods applied to positron emission tomography/computed tomography (PET/CT) images. These phantoms generally simulate either simple anatomical features or simple activity distribution around the metal inserts in the PET/CT scans. 3D printing has been used recently to fabricate fillable anthropomorphic phantoms that accurately simulate volume and geometry. This study aims to describe the process of image segmentation, phantom modelling, 3D printing and validation of a population-based fillable TKR phantom that simulates human TKR PET/CT metal artefacts.Methods 10 participants (5 male and 5 female) were scanned using 3T MRI and the images were segmented to create average male and average female 3D knee models, inversely with void cortical and porous trabecular compartments for 3D printing and contrast media. Virtual total knee replacement (TKR) surgery was implemented on these models to prepare the insertion locations for knee prosthetic implants. Subsequently, TKR models were printed using a 3D photopolymer resin printer and then injected with normal saline to test the phantoms for any leaks. Subsequently, diluted iodinated contrast media was injected into the cortical compartment and saline with 18F-FDG was injected into the trabecular compartment and the phantom was scanned with PET/CT. The images were then evaluated and compared to the human knee radiographic features reported in the literature.Results Phantoms were shown to be fluid-tight with distinct compartments. They showed comparable volume and geometry to the segmented human MRI knees. The phantoms demonstrated similar values for x-ray attenuation and Hounsfield units (HU) to the literature for both cortical and trabecular compartments. The phantoms displayed a uniform distribution for the radioactive tracer, resembling that seen in human trabecular bone PET. TKR phantom PET/CT images with metal inserts replicated the clinical metal artefacts seen clinically in the periprosthetic area.Conclusion This novel, 3D-printed, and customisable phantom effectively mimics the geometric, radiographic and radiotracer distribution features of real TKRs. Importantly, it simulates TKR image metal artefacts, making it suitable for repeatable and comprehensive evaluation of various metal artefact reduction methods in future research.
引用
收藏
页数:18
相关论文
共 34 条
[1]  
3D P., 2022, which exposure finder should I use?
[2]  
Abdoli M, Comparative methods for metal artifact reduction in x-ray CT
[3]   Review of the role of dynamic 18F-NaF PET in diagnosing and distinguishing between septic and aseptic loosening in hip prosthesis [J].
Adesanya, Olu ;
Sprowson, Andrew ;
Masters, James ;
Hutchinson, Charles .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2015, 10
[4]   Multi-Resin Masked Stereolithography (MSLA) 3D Printing for Rapid and Inexpensive Prototyping of Microfluidic Chips with Integrated Functional Components [J].
Ahmed, Isteaque ;
Sullivan, Katherine ;
Priye, Aashish .
BIOSENSORS-BASEL, 2022, 12 (08)
[5]   Construction and validation of a low cost paediatric pelvis phantom [J].
Ali, Ali Mohammed ;
Hogg, Peter ;
Johansen, Safora ;
England, Andrew .
EUROPEAN JOURNAL OF RADIOLOGY, 2018, 108 :84-91
[6]   Metal artefact reduction in CT imaging of hip prostheses-an evaluation of commercial techniques provided by four vendors [J].
Andersson, K. M. ;
Nowik, P. ;
Persliden, J. ;
Thunberg, P. ;
Norrman, E. .
BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1052)
[7]   Deep learning-based metal artefact reduction in PET/CT imaging [J].
Arabi, Hossein ;
Zaidi, Habib .
EUROPEAN RADIOLOGY, 2021, 31 (08) :6384-6396
[8]   Correlation of the quantitative methods for the measurement of bone uptake and plasma clearance of 18F-NaF using positron emission tomography. Systematic review and meta-analysis [J].
Assiri, Rajeh ;
Knapp, Karen ;
Fulford, Jon ;
Chen, Junning .
EUROPEAN JOURNAL OF RADIOLOGY, 2022, 146
[9]   F-18-Sodium Fluoride PET/CT and PET/MR Imaging of Bone and Joint Disorders [J].
Beheshti, Mohsen .
PET CLINICS, 2018, 13 (04) :477-490
[10]   Bismuth Infusion of ABS Enables Additive Manufacturing of Complex Radiological Phantoms and Shielding Equipment [J].
Ceh, Justin ;
Youd, Tom ;
Mastrovich, Zach ;
Peterson, Cody ;
Khan, Sarah ;
Sasser, Todd A. ;
Sander, Ian M. ;
Doney, Justin ;
Turner, Clark ;
Leevy, W. Matthew .
SENSORS, 2017, 17 (03)