Influence of reconstruction techniques on PET/CT image quality and quantitative accuracy: a phantom study

被引:1
作者
Su, Xuesong [1 ]
Geng, Jianhua [1 ]
Liu, Jianing [1 ]
Liu, Fengshuo [1 ]
Wu, Yichen [1 ]
Zheng, Rong [1 ]
Wang, Xuejuan [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Canc Hosp, Dept Nucl Med, Natl Canc Ctr,Natl Clin Res Ctr Canc, Beijing 100021, Peoples R China
关键词
Positron emission tomography (PET); Image quality; Quantification accuracy; Reconstruction techniques; Phantom; TIME-OF-FLIGHT; POINT-SPREAD-FUNCTION; PHYSICAL FIGURES; PERFORMANCE; RESOLUTION; IMPACT; ALGORITHMS; NOISE; MERIT; OSEM;
D O I
10.1007/s41605-023-00441-w
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
PurposeTo investigate the influence of different reconstruction techniques on the quantitative accuracy and image quality of PET/CT.MethodsThe NEMA NU2-2018 image quality phantom was scanned on a GE Discovery Elite PET/CT scanner and the spatial resolution was measured based on NEMA NU2 standard. The same raw data were reconstructed using five different algorithms: ordered subset expectation maximization (OSEM), OSEM with point spread function (PSF) modeling, OSEM with time-of-flight (TOF), OSEM with TOF and PSF, and filtered back-projection (FBP). The recovery coefficient (RC), contrast recovery coefficient (CRC), and contrast-to-noise ratio (CNR) were calculated for the six hot spheres, the percent background variability (PBV) and coefficient of variation (COV) were calculated for the background, and the residual error (RE) was calculated for lung insert in different image slices.ResultsThe incorporation of PSF modeling showed the smallest transverse FWHM and FWTM at both 1 and 10 cm radical offsets. The combination of PSF modeling and TOF improved RCmean and CRC for all spheres and resulted in the highest ratings for the detectability of 10 mm spheres in human observer assessment. PSF modeling played a role in reducing the COV within the background region of interest and increasing the CNR of the spheres, and decreased background noise ratings in human observer assessment. Besides, TOF significantly reduced the RE in lung insert. Neither PSF modeling nor TOF had a significant effect on PBV. Compared to FBP, the OSEM algorithm showed significant advantages in PBV, COV, CNR and RE and human observer ratings of image quality, but worse results for RCmax, RCmean and CRC.ConclusionsThe integration of TOF and PSF modeling into the OSEM algorithm achieves improvements in both quantitative accuracy and image quality, providing distinct advantages. PSF modeling improves the spatial resolution and decreases the visual appearance of background noise. The OSEM algorithm shows significantly better image quality than the FBP algorithm but no distinct advantages concerning quantitative accuracy.
引用
收藏
页码:1171 / 1186
页数:16
相关论文
共 50 条
  • [41] Quantitative Image Quality and Histogram-Based Evaluations of an Iterative Reconstruction Algorithm at Low-to-Ultralow Radiation Dose Levels: A Phantom Study in Chest CT
    Lee, Ki Baek
    Goo, Hyun Woo
    KOREAN JOURNAL OF RADIOLOGY, 2018, 19 (01) : 119 - 129
  • [42] A systematic task-based image quality assessment of photon-counting and energy integrating CT as a function of reconstruction kernel and phantom size
    Bhattarai, Mridul
    Bache, Steve
    Abadi, Ehsan
    Samei, Ehsan
    MEDICAL PHYSICS, 2024, 51 (02) : 1047 - 1060
  • [43] iQMetrix-CT: New software for task-based image quality assessment of phantom CT images
    Greffier, Joel
    Barbotteau, Yves
    Gardavaud, Francois
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2022, 103 (11) : 555 - 562
  • [44] Influence of Iterative Reconstruction Algorithms on PET Image Resolution
    Karpetas, G. E.
    Michail, C. M.
    Fountos, G. P.
    Valais, I. G.
    Nikolopoulos, D.
    Kandarakis, I. S.
    Panayiotakis, G. S.
    INTERNATIONAL CONFERENCE ON BIO-MEDICAL INSTRUMENTATION AND RELATED ENGINEERING AND PHYSICAL SCIENCES (BIOMEP 2015), 2015, 637
  • [45] Improvement of image quality of dentomaxillofacial region in ultra-high-resolution CT: a phantom study
    Sakai, Yuki
    Okamura, Kazutoshi
    Kitamoto, Erina
    Shirasaka, Takashi
    Kato, Toyoyuki
    Chikui, Toru
    Ishigami, Kousei
    DENTOMAXILLOFACIAL RADIOLOGY, 2024, : 203 - 209
  • [46] Benefits of Point-Spread Function and Time of Flight for PET/CT Image Quality in Relation to the Body Mass Index and Injected Dose
    Akamatsu, Go
    Mitsumoto, Katsuhiko
    Ishikawa, Kaori
    Taniguchi, Takafumi
    Ohya, Nobuyoshi
    Baba, Shingo
    Abe, Koichiro
    Sasaki, Masayuki
    CLINICAL NUCLEAR MEDICINE, 2013, 38 (06) : 407 - 412
  • [47] Image quality improvement with deep learning-based reconstruction on abdominal ultrahigh-resolution CT: A phantom study
    Shirasaka, Takashi
    Kojima, Tsukasa
    Funama, Yoshinori
    Sakai, Yuki
    Kondo, Masatoshi
    Mikayama, Ryoji
    Hamasaki, Hiroshi
    Kato, Toyoyuki
    Ushijima, Yasuhiro
    Asayama, Yoshiki
    Nishie, Akihiro
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (07): : 286 - 296
  • [48] Quantitative Techniques in PET-CT Imaging
    Basu, Sandip
    Zaidi, Habib
    Holm, Soren
    Alavi, Abass
    CURRENT MEDICAL IMAGING REVIEWS, 2011, 7 (03) : 216 - 233
  • [49] Brain image quality according to beam collimation width and image reconstruction algorithm: A phantom study
    Greffier, Joel
    Viry, Anais
    Durand, Quentin
    Hajdu, Steven David
    Frandon, Julien
    Beregi, Jean Paul
    Dabli, Djamel
    Racine, Damien
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 108
  • [50] Task-based quantification of image quality using a model observer in abdominal CT: a multicentre study
    Racine, Damien
    Ryckx, Nick
    Ba, Alexandre
    Becce, Fabio
    Viry, Anais
    Verdun, Francis R.
    Schmidt, Sabine
    EUROPEAN RADIOLOGY, 2018, 28 (12) : 5203 - 5210