Influence of reconstruction techniques on PET/CT image quality and quantitative accuracy: a phantom study

被引:1
作者
Su, Xuesong [1 ]
Geng, Jianhua [1 ]
Liu, Jianing [1 ]
Liu, Fengshuo [1 ]
Wu, Yichen [1 ]
Zheng, Rong [1 ]
Wang, Xuejuan [1 ]
机构
[1] Chinese Acad Med Sci & Peking Union Med Coll, Canc Hosp, Dept Nucl Med, Natl Canc Ctr,Natl Clin Res Ctr Canc, Beijing 100021, Peoples R China
关键词
Positron emission tomography (PET); Image quality; Quantification accuracy; Reconstruction techniques; Phantom; TIME-OF-FLIGHT; POINT-SPREAD-FUNCTION; PHYSICAL FIGURES; PERFORMANCE; RESOLUTION; IMPACT; ALGORITHMS; NOISE; MERIT; OSEM;
D O I
10.1007/s41605-023-00441-w
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
PurposeTo investigate the influence of different reconstruction techniques on the quantitative accuracy and image quality of PET/CT.MethodsThe NEMA NU2-2018 image quality phantom was scanned on a GE Discovery Elite PET/CT scanner and the spatial resolution was measured based on NEMA NU2 standard. The same raw data were reconstructed using five different algorithms: ordered subset expectation maximization (OSEM), OSEM with point spread function (PSF) modeling, OSEM with time-of-flight (TOF), OSEM with TOF and PSF, and filtered back-projection (FBP). The recovery coefficient (RC), contrast recovery coefficient (CRC), and contrast-to-noise ratio (CNR) were calculated for the six hot spheres, the percent background variability (PBV) and coefficient of variation (COV) were calculated for the background, and the residual error (RE) was calculated for lung insert in different image slices.ResultsThe incorporation of PSF modeling showed the smallest transverse FWHM and FWTM at both 1 and 10 cm radical offsets. The combination of PSF modeling and TOF improved RCmean and CRC for all spheres and resulted in the highest ratings for the detectability of 10 mm spheres in human observer assessment. PSF modeling played a role in reducing the COV within the background region of interest and increasing the CNR of the spheres, and decreased background noise ratings in human observer assessment. Besides, TOF significantly reduced the RE in lung insert. Neither PSF modeling nor TOF had a significant effect on PBV. Compared to FBP, the OSEM algorithm showed significant advantages in PBV, COV, CNR and RE and human observer ratings of image quality, but worse results for RCmax, RCmean and CRC.ConclusionsThe integration of TOF and PSF modeling into the OSEM algorithm achieves improvements in both quantitative accuracy and image quality, providing distinct advantages. PSF modeling improves the spatial resolution and decreases the visual appearance of background noise. The OSEM algorithm shows significantly better image quality than the FBP algorithm but no distinct advantages concerning quantitative accuracy.
引用
收藏
页码:1171 / 1186
页数:16
相关论文
共 50 条
  • [31] Evaluating and comparing the image quality and quantification accuracy of SiPM-PET/CT and PMT-PET/CT
    Yuji Tsutsui
    Shinichi Awamoto
    Kazuhiko Himuro
    Toyoyuki Kato
    Shingo Baba
    Masayuki Sasaki
    Annals of Nuclear Medicine, 2020, 34 : 725 - 735
  • [32] Assessment of organ dose and image quality in head and chest CT examinations: a phantom study
    Gharbi, S.
    Labidi, S.
    Mars, M.
    Chelli, M.
    Meftah, S.
    Ladeb, M. F.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (02) : 807 - 818
  • [33] The influence of SPECT reconstruction algorithms on image quality and diagnostic accuracy in phantom measurements and 99mTc-sestamibi parathyroid scintigraphy
    van Hoorn, Ralph A.
    Vriens, Dennis
    Postema, Jan-Willem A.
    Arens, Anne I. J.
    Pfestroff, Andreas
    Oyen, Wim J. G.
    Gotthardt, Martin
    NUCLEAR MEDICINE COMMUNICATIONS, 2014, 35 (01) : 64 - 72
  • [34] Image quality, diagnostic accuracy, and potential for radiation dose reduction in thoracoabdominal CT, using Sinogram Affirmed Iterative Reconstruction (SAFIRE) technique in a longitudinal study
    Scharf, Michael
    Brendel, Stephanie
    Melzer, Katja
    Hentschke, Christian
    May, Matthias
    Uder, Michael
    Lell, Michael M.
    PLOS ONE, 2017, 12 (07):
  • [35] Influence of PET reconstruction parameters on the TrueX algorithm A combined phantom and patient study
    Knaeusl, B.
    Rausch, I. F.
    Bergmann, H.
    Dudczak, R.
    Hirtl, A.
    Georg, D.
    NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2013, 52 (01): : 28 - 35
  • [36] A biological phantom for evaluation of CT image reconstruction algorithms
    Cammin, J.
    Fung, G. S. K.
    Fishman, E. K.
    Siewerdsen, J. H.
    Stayman, J. W.
    Taguchi, K.
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [37] Standardization and optimization of Siemens Biograph TruePoint PET/CT acquisition and reconstruction parameters: Simultaneous qualitative and quantitative assessments
    Vosoughi, Habibeh
    Emami, Farshad
    Momennezhad, Mehdi
    Geramifar, Parham
    Rahmim, Arman
    IRANIAN JOURNAL OF NUCLEAR MEDICINE, 2024, 32 (02): : 183 - 193
  • [38] Improvement in PET/CT image quality in overweight patients with PSF and TOF
    Takafumi Taniguchi
    Go Akamatsu
    Yukiko Kasahara
    Katsuhiko Mitsumoto
    Shingo Baba
    Yuji Tsutsui
    Kazuhiko Himuro
    Shohei Mikasa
    Daisuke Kidera
    Masayuki Sasaki
    Annals of Nuclear Medicine, 2015, 29 : 71 - 77
  • [39] New standards for phantom image quality and SUV harmonization range for multicenter oncology PET studies
    Akamatsu, Go
    Shimada, Naoki
    Matsumoto, Keiichi
    Daisaki, Hiromitsu
    Suzuki, Kazufumi
    Watabe, Hiroshi
    Oda, Keiichi
    Senda, Michio
    Terauchi, Takashi
    Tateishi, Ukihide
    ANNALS OF NUCLEAR MEDICINE, 2022, 36 (02) : 144 - 161
  • [40] Quality control for quantitative multicenter whole-body PET/MR studies: A NEMA image quality phantom study with three current PET/MR systems
    Boellaard, Ronald
    Rausch, Ivo
    Beyer, Thomas
    Delso, Gaspar
    Yaqub, Maqsood
    Quick, Harald H.
    Sattler, Bernhard
    MEDICAL PHYSICS, 2015, 42 (10) : 5961 - 5969