Commissioning and dosimetric verification of volumetric modulated arc therapy for multiple modalities using electronic portal imaging device-based 3D dosimetry system: a novel approach

被引:0
|
作者
Hajare, Raghavendra [1 ,2 ]
Sreelakshmi, K. K. [1 ]
Kumar, Anil [1 ]
Kalita, Rituraj [3 ]
Kaginelli, Shanmukhappa [2 ]
Mahantshetty, Umesh [1 ]
机构
[1] Homi Bhabha Canc Hosp & Res Ctr, Dept Radiat Oncol, Visakhapatnam, Andhra Pradesh, India
[2] JSS Acad Higher Educ & Res, Div Med Phys, Mysuru, India
[3] Tezpur Canc Ctr, Dept Radiat Oncol, Bihuguri, India
关键词
3D dosimetry system; EPID; VMAT commissioning; In-Vivo; PSQA; AAPM TG 119; IMRT;
D O I
10.1007/s12194-024-00792-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to validate an electronic portal imaging device (EPID) based 3-dimensional (3D) dosimetry system for the commissioning of volumetric modulated arc therapy (VMAT) delivery for flattening filter (FF) and flattening filter free (FFF) modalities based on test suites developed according to American Association of Physicists in Medicine Task Group 119 (AAPM TG 119) and pre-treatment patient specific quality assurance (PSQA).With ionisation chamber, multiple-point measurement in various planes becomes extremely difficult and time-consuming, necessitating repeated exposure of the plan. The average agreement between measured and planned doses for TG plans is recommended to be within 3%, and both the ionisation chamber and PerFRACTION (TM) measurement were well within this prescribed limit. Both point dose differences with the planned dose and gamma passing rates are comparable with TG reported multi-institution results. From our study, we found that no significant differences were found between FF and FFF beams for measurements using PerFRACTION (TM) and ion chamber. Overall, PerFRACTION (TM) produces acceptable results to be used for commissioning and validating VMAT and for performing PSQA. The findings support the feasibility of integrating PerFRACTION (TM) into routine quality assurance procedures for VMAT delivery. Further multi-institutional studies are recommended to establish global baseline values and enhance the understanding of PerFRACTION (TM)'s capabilities in diverse clinical settings.
引用
收藏
页码:412 / 424
页数:13
相关论文
共 27 条
  • [1] 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry
    Mans, Anton
    Remeijer, Peter
    Olaciregui-Ruiz, Igor
    Wendling, Markus
    Sonke, Jan-Jakob
    Mijnheer, Ben
    van Herk, Marcel
    Stroom, Joep C.
    RADIOTHERAPY AND ONCOLOGY, 2010, 94 (02) : 181 - 187
  • [2] On-Line Dosimetric Verification of Respiratory Gated Volumetric Modulated Arc Therapy Using the Electronic Portal Imaging Device
    Schaly, B.
    Xhaferllari, I.
    Gaede, S.
    MEDICAL PHYSICS, 2015, 42 (06) : 3469 - 3470
  • [3] Leaf Position Verification during Dynamic Beam Delivery of Volumetric Modulated Arc Therapy Using Electronic Portal Imaging Device
    Sarkar, B.
    Manikandan, A.
    Holla, R.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [4] Dosimetric precision of 3D gel dosimetry compared with radiochromic films in volumetric modulated arc therapy
    Skyt, P. S.
    Hoye, E. M.
    Yates, E. S.
    Schmidt, M. L.
    Petersen, J. B. B.
    Balling, P.
    Muren, L. P.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S409 - S410
  • [5] A study on dosimetric properties of electronic portal imaging device and its use as a quality assurance tool in Volumetric Modulated Arc Therapy
    Sukumar, Prabakar
    Padmanaban, Sriram
    Jeevanandam, Prakash
    Kumar, S. A. Syam
    Nagarajan, Vivekanandan
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2011, 16 (06) : 248 - 255
  • [6] Electronic Portal Imaging Device-Based Three-Dimensional Volumetric Dosimetry for Intensity-modulated Radiotherapy Pretreatment Quality Assurance
    Arjunan, Manikandan
    Sekaran, Sureka Chandra
    Sarkar, Biplab
    Manavalan, Saran Kumar
    JOURNAL OF MEDICAL PHYSICS, 2019, 44 (03) : 176 - 184
  • [7] Using an Electronic Portal Imaging Device for Patient-Specific Volumetric Modulated Arc Therapy Quality Assurance
    Bakhtiari, M.
    Kumaraswamy, L.
    de Boer, S.
    Malhotra, H.
    Podgorsak, M.
    MEDICAL PHYSICS, 2010, 37 (06) : 3087 - +
  • [8] Clinical Implementation of a Model-Based In Vivo Dose Verification System for Stereotactic Body Radiation Therapy-Volumetric Modulated Arc Therapy Treatments Using the Electronic Portal Imaging Device
    McCowan, Peter M.
    Asuni, Ganiyu
    Van Uytven, Eric
    VanBeek, Timothy
    McCurdy, Boyd M. C.
    Loewen, Shaun K.
    Ahmed, Naseer
    Bashir, Bashir
    Butler, James B.
    Chowdhury, Amitava
    Dubey, Arbind
    Leylek, Ahmet
    Nashed, Maged
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 97 (05): : 1077 - 1084
  • [9] Patient Specific Quality Assurance for Volumetric Modulated Arc Therapy (RapidArc) Using COMPASS 3D Dosimetry System
    Arun, G.
    Subramanian, V. Shanmuga
    Swamy, S. Thirumalai
    Kathirvel, M.
    Ramalingam, K.
    Kala, S.
    Kumar, S. Ashok
    Karthikeyan, K.
    Anantharaman, A.
    Kumar, N. Jagdish
    Arivarasan, I.
    Babaiah, M.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [10] Fast 3D dosimetric verifications based on an electronic portal imaging device using a GPU calculation engine
    Zhu, Jinhan
    Chen, Lixin
    Chen, Along
    Luo, Guangwen
    Deng, Xiaowu
    Liu, Xiaowei
    RADIATION ONCOLOGY, 2015, 10