Laser Target System in Combination with an Aiming Device for Percutaneous CT-Guided Interventions - An Accuracy Study

被引:4
作者
Scharll, Yannick [1 ]
Boehler, Daniel [1 ]
Laimer, Gregor [1 ]
Schullian, Peter [1 ]
Bale, Reto [1 ,2 ]
机构
[1] Med Univ Innsbruck, Dept Radiol, Intervent Oncol Microinvas Therapy SIP, Anichstr 35, A-6020 Innsbruck, Austria
[2] Med Univ Innsbruck, Dept Radiol, Intervent Oncol Microinvas Therapy, Anichstr 35, A-6020 Innsbruck, Austria
关键词
Accuracy; CT-guidance; Laser guidance; Aiming device; Percutaneous; RADIOFREQUENCY ABLATION; NAVIGATION SYSTEM; RADIATION-EXPOSURE; GUIDANCE-SYSTEM; FREEHAND; FEASIBILITY; PUNCTURES; BIOPSIES; LESIONS; TUMORS;
D O I
10.1016/j.acra.2023.03.028
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: To evaluate the targeting accuracy of laser-guided punctures in combination with an aiming device for computed tomography (CT) interventions during in vitro experiments.Materials and Methods: A total of 600 CT-guided punctures were performed using a laser target system, half of them with the additional help of an aiming device. Conically shaped targets in a plexiglass phantom were punctured. The planning CT data sets were acquired with 1.25, 2.5 and 5 mm slice thickness. Needle placement accuracy, as well as procedural time, was assessed. The Euclidean (ED) and normal distances (ND) were calculated at the target point.Results: Using the aiming device, the accomplished mean ND at the target for the 1.25, 2.5 and 5 mm slice thickness was 1.76 mm (SD +/- 0.92), 2.09 mm (SD +/- 1.06) and 1.93 mm (SD +/- 1.38), respectively. Without aiming device, the corresponding results were 2.55 mm (SD +/- 1.42), 2.7 mm (SD +/- 1.43) and 2.31 mm (SD +/- 1.64). At a slice thickness of 1.25 mm and 2.5 mm, punctures with the aiming device were significantly more accurate for both the ED and ND as compared to the punctures without aiming device (p < 0.001). The mean time required to complete the procedure, including image acquisition, trajectory planning, the placement of 10 needles, and the control-CT scan was 24.8 min without and 29.8 min with the aiming device.Conclusion: The additional use of the aiming device in combination with the commercially available laser guidance system significantly increased the level of accuracy during this in vitro experiment compared to freehand passes.
引用
收藏
页码:3047 / 3055
页数:9
相关论文
共 34 条
  • [1] CT-guided Navigation of Percutaneous Hepatic and Renal Radiofrequency Ablation under High-frequency Jet Ventilation: Feasibility Study
    Abderhalden, Susanne
    Biro, Peter
    Hechelhammer, Lukas
    Pfiffner, Roger
    Pfammatter, Thomas
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 22 (09) : 1275 - 1278
  • [2] Robot-assisted radiofrequency ablation of primary and secondary liver tumours: early experience
    Abdullah, Basri Johan Jeet
    Yeong, Chai Hong
    Goh, Khean Lee
    Yoong, Boon Koon
    Ho, Gwo Fuang
    Yim, Carolyn Chue Wai
    Kulkarni, Anjali
    [J]. EUROPEAN RADIOLOGY, 2014, 24 (01) : 79 - 85
  • [3] Preliminary clinical experience with a dedicated interventional robotic system for CT-guided biopsies of lung lesions: a comparison with the conventional manual technique
    Anzidei, Michele
    Argiro, Renato
    Porfiri, Andrea
    Boni, Fabrizio
    Anile, Marco
    Zaccagna, Fulvio
    Vitolo, Domenico
    Saba, Luca
    Napoli, Alessandro
    Leonardi, Andrea
    Longo, Flavia
    Venuta, Federico
    Bezzi, Mario
    Catalano, Carlo
    [J]. EUROPEAN RADIOLOGY, 2015, 25 (05) : 1310 - 1316
  • [4] Navigated CT-guided interventions
    Bale, Reto
    Widmann, Gerlig
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2007, 16 (04) : 196 - 204
  • [5] A novel vacuum device for extremity immobilisation during digital angiography: preliminary clinical experiences
    Bale, RJ
    Lottersberger, C
    Vogele, M
    Prassl, A
    Czermak, B
    Dessl, A
    Sweeney, RA
    Waldenberger, P
    Jaschke, W
    [J]. EUROPEAN RADIOLOGY, 2002, 12 (12) : 2890 - 2894
  • [6] Osteochondral lesions of the talus: Computer-assisted retrograde drilling - Feasibility and accurracy in initial experiences
    Bale, RJ
    Hoser, C
    Rosenberger, R
    Rieger, M
    Benedetto, KP
    Fink, C
    [J]. RADIOLOGY, 2001, 218 (01) : 278 - 282
  • [7] Head and neck tumors: Fractionated frameless stereotactic interstitial brachytherapy - Initial experience
    Bale, RJ
    Freysinger, W
    Gunkel, AR
    Vogele, M
    Sztankay, A
    Auer, T
    Eichberger, P
    Martin, A
    Auberger, T
    Scholtz, AW
    Jaschke, W
    Thumfart, WF
    Lukas, P
    [J]. RADIOLOGY, 2000, 214 (02) : 591 - 595
  • [8] Bartsch H-J., 2001, Taschenbuch mathematischer Formeln
  • [9] Multicenter evaluation of a new laser guidance system for computed tomography intervention
    Brabrand, K
    Aalokken, TM
    Krombach, GA
    Günther, RW
    Tariq, R
    Magnusson, A
    Lindgren, PG
    [J]. ACTA RADIOLOGICA, 2004, 45 (03) : 308 - 312
  • [10] Robotic laser position versus freehand in CT-guided percutaneous microwave ablation for single hepatocellular carcinoma (diameter < 3 cm): a preliminary study
    Dong, Yuru
    Wang, Guisheng
    Zhang, Jingjun
    Zhang, Shu
    Chen, Xiaoxia
    Guo, Qing
    Qu, Feihuan
    Shou, Feng
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2022, 39 (01) : 725 - 732