Noninvasive blood pressure measurement in patients with continuous-flow left ventricular assist devices

被引:1
|
作者
Tanaka, Aiko [1 ]
Kiriyama, Yuki [2 ]
Kubo, Naoko [1 ]
Sakaguchi, Ryota [1 ]
Uchiyama, Akinori [1 ]
Fujino, Yuji [1 ]
机构
[1] Osaka Univ, Grad Sch Med, Dept Anesthesiol & Intens Care Med, 2-15 Yamadaoka, Suita, Osaka 5650871, Japan
[2] Natl Org Osaka Natl Hosp, Div Anesthesiol, Chuo Ku, 2-1-14 Hoenzaka, Osaka 5400006, Japan
关键词
Blood pressure; Noninvasive measurement; Left ventricular assist device; Doppler ultrasound;
D O I
10.1007/s10047-022-01349-w
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Oscillatory blood pressure (OBP) with a slow cuff-deflation system has been proposed as noninvasive measurement of mean arterial pressure (MAP) in patients with continuous-flow left ventricular assist devices (LVADs). However, the challenge is that the measurement is not obtainable in certain patients. We hypothesized that the combined use of color Doppler imaging during OBP measurement (CDBP) could derive MAP accurately. We conducted a prospective observational study in critically ill patients (30 patients with continuous-flow LVADs and 30 control patients without LVADs). Triplicate OBP and CDBP measurements were performed and invasive blood pressure (IBP) was recorded. The overall success rate of OBP was 63.3% in the LVAD group and 98.9% in the control group. The CDBP was successfully obtained in 100% of all study patients. The CDBP in the LVAD group was closest to the MAP of measured IBP, while that in the control group was closest to the systolic IBP. The mean absolute differences in OBP and CDBP from the closest IBP were similar in both the control and LVAD groups. In nonpulsatile LVAD patients with a pulse pressure IBP < 10 mmHg, the success rate of OBP measurement was only 10.0%, and CDBP showed significantly reduced error in MAP measurement (mean absolute difference: OBP 23.2 +/- 8.7 vs CDBP 5.2 +/- 3.6 mmHg, p < 0.001). The validity of OBP measurement with a slow cuff-deflation system limited particularly in nonpulsatile LVAD patients. The concurrent use of color Doppler imaging is encouraged for more accurate measurement of MAP in patients with continuous-flow LVADs.
引用
收藏
页码:160 / 164
页数:5
相关论文
共 50 条
  • [31] The Physiological Rationale for Incorporating Pulsatility in Continuous-Flow Left Ventricular Assist Devices
    Grosman-Rimon, Liza
    Billia, Filio
    Kobulnik, Jeremy
    Bar-Ziv, Stacey Pollock
    Cherney, David Z.
    Rao, Vivek
    CARDIOLOGY IN REVIEW, 2018, 26 (06) : 294 - 301
  • [32] A Comprehensive Imaging Approach to Guide the Management of Patients with Continuous-Flow Left Ventricular Assist Devices
    Hannawi, Bashar
    Estep, Jerry D.
    CARDIOLOGY CLINICS, 2018, 36 (04) : 583 - +
  • [33] Prognostic implications of serial outpatient blood pressure measurements in patients with an axial continuous-flow left ventricular assist device
    Pinsino, Alberto
    Castagna, Francesco
    Zuver, Amelia M.
    Royzman, Eugene A.
    Nasiri, Mojdeh
    Stohr, Eric J.
    Cagliostro, Barbara
    McDonnell, Barry
    Cockcroft, John R.
    Garan, A. Reshad
    Topkara, Veli K.
    Schulze, P. Christian
    Takeda, Koji
    Takayama, Hiroo
    Naka, Yoshifumi
    Demmer, Ryan T.
    Willey, Joshua Z.
    Yuzefpolskaya, Melana
    Colombo, Paolo C.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2019, 38 (04) : 396 - 405
  • [34] Physiologic and Pathologic Changes in Patients with Continuous-Flow Ventricular Assist Devices
    John, Ranjit
    Boyle, Andrew
    Pagani, Frank
    Miller, Leslie
    JOURNAL OF CARDIOVASCULAR TRANSLATIONAL RESEARCH, 2009, 2 (02) : 154 - 158
  • [35] Physiologic and Pathologic Changes in Patients with Continuous-Flow Ventricular Assist Devices
    Ranjit John
    Andrew Boyle
    Frank Pagani
    Leslie Miller
    Journal of Cardiovascular Translational Research, 2009, 2 : 154 - 158
  • [36] Management of hypertension in patients supported with continuous flow left ventricular assist devices
    Krishnarao, Krithika
    Krim, Selim R.
    CURRENT OPINION IN CARDIOLOGY, 2023, 38 (04) : 281 - 286
  • [37] Impact of Cardiac Resynchronization Therapy on Clinical Outcomes in Patients With Continuous-Flow Left Ventricular Assist Devices
    Gopinathannair, Rakesh
    Birks, Emma J.
    Trivedi, Jaimin R.
    Mccants, Kelly C.
    Sutton, Brad S.
    Deam, Allen G.
    Slaughter, Mark S.
    Hottigoudar, Raseimi U.
    JOURNAL OF CARDIAC FAILURE, 2015, 21 (03) : 226 - 232
  • [38] Non-invasive assessment of arterial pulsatility in patients with continuous-flow left ventricular assist devices
    Cross, Troy J.
    Sajgalik, Pavol
    Fabian, Vratislav
    Matera, Lukas
    Kushwaha, Sudhir S.
    Maltais, Simon
    Stulak, John M.
    Schirger, John A.
    Johnson, Bruce D.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2020, 43 (02) : 99 - 108
  • [39] Bleeding in patients with continuous-flow left ventricular assist devices: acquired von Willebrand disease or antithrombotics?
    Consolo, Filippo
    Marasi, Alessandra
    Della Valle, Patrizia
    Bonora, Marta
    Pieri, Marina
    Scandroglio, Anna Mara
    Redaelli, Alberto
    Zangrillo, Alberto
    D'Angelo, Armando
    Pappalardo, Federico
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2022, 62 (01)
  • [40] Risk of pump thrombosis and stroke in patients with continuous-flow left ventricular assist devices and gastrointestinal bleeding
    Szymanski, Thomas W.
    Weeks, Phillip A.
    Patel, Chandni J.
    Jezovnik, Mateja K.
    Gulbis, Brian
    Nathan, Sriram S.
    Jumean, Marwan F.
    Radovancevic, Rajko
    Kar, Biswajit
    Gregoric, Igor D.
    ARTIFICIAL ORGANS, 2020, 44 (11) : 1171 - 1175