A feasibility study using an array of LaBr3(Ce) scintillation detectors as a Compton camera for prompt gamma imaging during BNCT

被引:2
作者
Nutter, Kiran [1 ]
Price, Tony [1 ]
Kokalova, Tzany [1 ]
Green, Stuart [1 ,2 ]
Phoenix, Ben [1 ]
机构
[1] Univ Birmingham, Sch Phys & Astron, Birmingham, England
[2] Univ Hosp Birmingham NHS Fdn Trust, Queen Elizabeth Hosp Birmingham, Birmingham, England
基金
英国工程与自然科学研究理事会;
关键词
BNCT; Compton camera; dosimetry; prompt gamma imaging; LaBr3; EMISSION COMPUTED-TOMOGRAPHY; NEUTRON-CAPTURE THERAPY; SPECT SYSTEM; PG-SPECT; RECONSTRUCTION; RAYS;
D O I
10.3389/fphy.2024.1347929
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Boron Neutron Capture Therapy (BNCT) is a binary cancer therapy where a low energy neutron beam is incident upon a patient who has been administered a tumour-seeking 10B loaded compound. The neutron capture reaction on 10B results in the production of two short range particles, Li-7 and He-4, that deposit all of their energies within the targeted cell. However, accurate, online dosimetry during BNCT is challenging as it requires knowledge of both the neutron fluence and B-10 concentration in cells. An additional product in the neutron capture reaction on B-10 is a 478 keV prompt gamma ray, and if the production vertices of these gamma rays could be imaged by an external camera, the density of the vertices could be used to infer the dose delivered to the patient. In this study, the feasibility of using an array of LaBr3 scintillators as a modified Compton camera for prompt gamma imaging during BNCT was investigated using Geant4 simulations. These simulations demonstrated that a phantom containing a 3 cm diameter region of 400 ppm B-10 could be reconstructed using clinically relevant neutron fluences. This result opens up more possibilities for future research to improve dosimetry during BNCT.
引用
收藏
页数:12
相关论文
共 29 条
[21]   GAMMA-RAY IMAGING WITH COMPTON CAMERAS [J].
PHILLIPS, GW .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1995, 99 (1-4) :674-677
[22]   Prompt gamma tomography during BNCT -: A feasibility study [J].
Rosenschold, Per Munck Af ;
Minarik, David ;
Ostlund, Carl ;
Ljungberg, Michael ;
Ceberg, Crister .
JOURNAL OF INSTRUMENTATION, 2006, 1
[23]  
Seltzer S., 1987, NIST Standard Reference Database, V8, DOI [10.18434/T48G6X, DOI 10.18434/T48G6X]
[24]   Nuclear data sheets for A=80 [J].
Singh, B .
NUCLEAR DATA SHEETS, 2005, 105 (02) :223-+
[26]   AN ELECTRONICALLY COLLIMATED GAMMA-CAMERA FOR SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY .2. IMAGE-RECONSTRUCTION AND PRELIMINARY EXPERIMENTAL MEASUREMENTS [J].
SINGH, M ;
DORIA, D .
MEDICAL PHYSICS, 1983, 10 (04) :428-435
[27]   Compton imaging for medical applications [J].
Tashima, Hideaki ;
Yamaya, Taiga .
RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2022, 15 (03) :187-205
[28]   PROPOSED GAMMA CAMERA [J].
TODD, RW ;
NIGHTINGALE, JM ;
EVERETT, DB .
NATURE, 1974, 251 (5471) :132-134
[29]  
Valda A, 2005, BRAZ J PHYS, V35, P785, DOI 10.1590/S0103-97332005000500017