A feasibility study using an array of LaBr3(Ce) scintillation detectors as a Compton camera for prompt gamma imaging during BNCT

被引:2
作者
Nutter, Kiran [1 ]
Price, Tony [1 ]
Kokalova, Tzany [1 ]
Green, Stuart [1 ,2 ]
Phoenix, Ben [1 ]
机构
[1] Univ Birmingham, Sch Phys & Astron, Birmingham, England
[2] Univ Hosp Birmingham NHS Fdn Trust, Queen Elizabeth Hosp Birmingham, Birmingham, England
基金
英国工程与自然科学研究理事会;
关键词
BNCT; Compton camera; dosimetry; prompt gamma imaging; LaBr3; EMISSION COMPUTED-TOMOGRAPHY; NEUTRON-CAPTURE THERAPY; SPECT SYSTEM; PG-SPECT; RECONSTRUCTION; RAYS;
D O I
10.3389/fphy.2024.1347929
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Boron Neutron Capture Therapy (BNCT) is a binary cancer therapy where a low energy neutron beam is incident upon a patient who has been administered a tumour-seeking 10B loaded compound. The neutron capture reaction on 10B results in the production of two short range particles, Li-7 and He-4, that deposit all of their energies within the targeted cell. However, accurate, online dosimetry during BNCT is challenging as it requires knowledge of both the neutron fluence and B-10 concentration in cells. An additional product in the neutron capture reaction on B-10 is a 478 keV prompt gamma ray, and if the production vertices of these gamma rays could be imaged by an external camera, the density of the vertices could be used to infer the dose delivered to the patient. In this study, the feasibility of using an array of LaBr3 scintillators as a modified Compton camera for prompt gamma imaging during BNCT was investigated using Geant4 simulations. These simulations demonstrated that a phantom containing a 3 cm diameter region of 400 ppm B-10 could be reconstructed using clinically relevant neutron fluences. This result opens up more possibilities for future research to improve dosimetry during BNCT.
引用
收藏
页数:12
相关论文
共 29 条
[1]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]   Experimental validation of a spectroscopic gamma-ray detector based on a LaBr3 scintillator towards real-time dose monitoring in BNCT [J].
Caracciolo, Anita ;
Di Vita, Davide ;
Buonanno, Luca ;
Carminati, Marco ;
Protti, Nicoletta ;
Altieri, Saverio ;
Pola, Andrea ;
Bortot, Davide ;
Fiorini, Carlo .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1041
[3]   Boron neutron capture therapy: Cellular targeting of high linear energy transfer radiation [J].
Coderre, JA ;
Turcotte, JC ;
Riley, KJ ;
Binns, PJ ;
Harling, OK ;
Kiger, WS .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2003, 2 (05) :355-375
[4]   A SOURCE RECONSTRUCTION METHOD FOR MULTIPLE SCATTER COMPTON CAMERAS [J].
DOGAN, N ;
WEHE, DK ;
AKCASU, AZ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (05) :1427-1430
[5]   MULTIPLE COMPTON-SCATTERING GAMMA-RAY IMAGING CAMERA [J].
DOGAN, N ;
WEHE, DK ;
KNOLL, GF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 299 (1-3) :501-506
[6]   Optimization of the Compton camera for measuring prompt gamma rays in boron neutron capture therapy [J].
Gong, Chun-hui ;
Tang, Xiao-bin ;
Shu, Di-yun ;
Yu, Hai-yan ;
Geng, Chang-ran .
APPLIED RADIATION AND ISOTOPES, 2017, 124 :62-67
[7]   Feasibility study of SPECT system for online dosimetry imaging in boron neutron capture therapy [J].
Hales, B. ;
Katabuchi, T. ;
Hayashizaki, N. ;
Terada, K. ;
Igashira, M. ;
Kobayashi, T. .
APPLIED RADIATION AND ISOTOPES, 2014, 88 :167-170
[8]   Predicted performance of a PG-SPECT system using CZT primary detectors and secondary Compton-suppression anti-coincidence detectors under near-clinical settings for boron neutron capture therapy [J].
Hales, Brian ;
Katabuchi, Tatsuya ;
Igashira, Masayuki ;
Terada, Kazushi ;
Hayashizaki, Noriyosu ;
Kobayashi, Tooru .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 875 :51-56
[9]   Boron concentration prediction from Compton camera image for boron neutron capture therapy based on generative adversarial network [J].
Hou, Zhenfeng ;
Geng, Changran ;
Tang, Xiaobin ;
Tian, Feng ;
Zhao, Sheng ;
Qi, Jie ;
Shu, Diyun ;
Gong, Chunhui .
APPLIED RADIATION AND ISOTOPES, 2022, 186
[10]  
IAEA, 2023, Advances in Boron Neutron Capture Therapy