Machine learning and reduced order modelling for the simulation of braided stent deployment

被引:10
作者
Bisighini, Beatrice [1 ,2 ,3 ]
Aguirre, Miquel [1 ,4 ,5 ]
Biancolini, Marco Evangelos [3 ]
Trovalusci, Federica [3 ]
Perrin, David [2 ]
Avril, Stephane [1 ]
Pierrat, Baptiste [1 ]
机构
[1] Univ Lyon, Univ Jean Monnet, Mines St Etienne, INSERM, St Etienne, France
[2] Grande Usine Creat 2, Predisurge, St Etienne, France
[3] Univ Tor Vergata, Dept Enterprise Engn, Rome, Italy
[4] Univ Politecn Cataluna, Lab Calcul Numer, Barcelona, Spain
[5] Gran Capita, Int Ctr Numer Methods Engn CIMNE, Barcelona, Spain
关键词
braided stent; beam elements; contact mechanics; reduced order modelling; machine learning; UNRUPTURED INTRACRANIAL ANEURYSMS; FLOW DIVERSION; SIZE RATIO; IN-VITRO; RUPTURE; PREVALENCE; VALIDATION;
D O I
10.3389/fphys.2023.1148540
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
Endoluminal reconstruction using flow diverters represents a novel paradigm for the minimally invasive treatment of intracranial aneurysms. The configuration assumed by these very dense braided stents once deployed within the parent vessel is not easily predictable and medical volumetric images alone may be insufficient to plan the treatment satisfactorily. Therefore, here we propose a fast and accurate machine learning and reduced order modelling framework, based on finite element simulations, to assist practitioners in the planning and interventional stages. It consists of a first classification step to determine a priori whether a simulation will be successful (good conformity between stent and vessel) or not from a clinical perspective, followed by a regression step that provides an approximated solution of the deployed stent configuration. The latter is achieved using a non-intrusive reduced order modelling scheme that combines the proper orthogonal decomposition algorithm and Gaussian process regression. The workflow was validated on an idealized intracranial artery with a saccular aneurysm and the effect of six geometrical and surgical parameters on the outcome of stent deployment was studied. We trained six machine learning models on a dataset of varying size and obtained classifiers with up to 95% accuracy in predicting the deployment outcome. The support vector machine model outperformed the others when considering a small dataset of 50 training cases, with an accuracy of 93% and a specificity of 97%. On the other hand, real-time predictions of the stent deployed configuration were achieved with an average validation error between predicted and high-fidelity results never greater than the spatial resolution of 3D rotational angiography, the imaging technique with the best spatial resolution (0.15 mm). Such accurate predictions can be reached even with a small database of 47 simulations: by increasing the training simulations to 147, the average prediction error is reduced to 0.07 mm. These results are promising as they demonstrate the ability of these techniques to achieve simulations within a few milliseconds while retaining the mechanical realism and predictability of the stent deployed configuration.
引用
收藏
页数:18
相关论文
共 68 条
  • [11] Care A., 2018, Machine Learning Techniques for Space Weather, P71, DOI DOI 10.1016/B978-0-12-811788-0.00004-4
  • [12] A reduced-order model for wall shear stress in abdominal aortic aneurysms by proper orthogonal decomposition
    Chang, Gary Han
    Schirmer, Clemens M.
    Modarres-Sadeghi, Yahya
    [J]. JOURNAL OF BIOMECHANICS, 2017, 54 : 33 - 43
  • [13] Chatterjee A, 2000, CURR SCI INDIA, V78, P808
  • [14] Physics-informed machine learning for reduced-order modeling of nonlinear problems
    Chen, Wenqian
    Wang, Qian
    Hesthaven, Jan S.
    Zhang, Chuhua
    [J]. JOURNAL OF COMPUTATIONAL PHYSICS, 2021, 446
  • [15] Statistical Shape Analysis of Ascending Thoracic Aortic Aneurysm: Correlation between Shape and Biomechanical Descriptors
    Cosentino, Federica
    Raffa, Giuseppe M.
    Gentile, Giovanni
    Agnese, Valentina
    Bellavia, Diego
    Pilato, Michele
    Pasta, Salvatore
    [J]. JOURNAL OF PERSONALIZED MEDICINE, 2020, 10 (02):
  • [16] Flow Diversion for Intracranial Aneurysms A Review
    D'Urso, Pietro I.
    Lanzino, Giuseppe
    Cloft, Harry J.
    Kallmes, David F.
    [J]. STROKE, 2011, 42 (08) : 2363 - 2368
  • [17] Danu M, 2019, INT CONF SYST THEO, P662, DOI [10.1109/ICSTCC.2019.8885576, 10.1109/icstcc.2019.8885576]
  • [18] Our capricious vessels: The influence of stent design and vessel geometry on the mechanics of intracranial aneurysm stent deployment
    De Bock, S.
    Iannaccone, F.
    De Santis, G.
    De Beule, M.
    Mortier, P.
    Verhegghe, B.
    Segers, P.
    [J]. JOURNAL OF BIOMECHANICS, 2012, 45 (08) : 1353 - 1359
  • [19] POD-DL-ROM: Enhancing deep learning-based reduced order models for nonlinear parametrized PDEs by proper orthogonal decomposition
    Fresca, Stefania
    Manzoni, Andrea
    [J]. COMPUTER METHODS IN APPLIED MECHANICS AND ENGINEERING, 2022, 388
  • [20] Interaction between Flow Diverter and Parent Artery of Intracranial Aneurysm: A Computational Study
    Fu, Wenyu