Durability of cantilever inlay-retained fixed dental prosthesis fabricated from multilayered zirconia ceramics with different designs

被引:8
作者
Al-Dwairi, Ziad N. [1 ]
Al-Aghbari, Latifah [1 ]
Husain, Nadin Al-Haj [2 ,3 ]
Ozcan, Mutlu [2 ]
机构
[1] Jordan Univ Sci & Technol JUST, Fac Dent, Dept Prosthodont, Irbid 22110, Jordan
[2] Univ Zurich, Div Dent Biomat, Clin Reconstruct Dent, Zurich, Switzerland
[3] Univ Bern, Sch Dent Med, Dept Reconstruct Dent & Gerodontol, Bern, Switzerland
关键词
All -ceramic restoration; Cantilever; Dental materials; Fracture resistance; Inlay -retained fixed partial dentures; Monolithic zirconia; Preparation design; Prosthodontics; LOAD-BEARING CAPACITY; FRACTURE-RESISTANCE; PARTIAL DENTURES; ANTERIOR; SURVIVAL; TEETH; RESTORATIONS;
D O I
10.1016/j.jmbbm.2022.105547
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: The purpose of this in-vitro study was to investigate the effect of framework design on fracture resistance and failure modes of cantilever inlay-retained fixed partial dentures (IRFDPs) fabricated from two multilayered monolithic zirconia materials.Materials and methods: Seventy-two natural premolar teeth were prepared as abutments for cantilever IRFDPs using three designs: mesial-occlusal (MO) inlay with short buccal and palatal wings (D1), MO inlay with long palatal wing (D2), MO inlay with long palatal wing and occlusal extension (D3). Full-contoured IRFDPs were fabricated from two monolithic zirconia materials; IPS e.max ZirCAD Prime and Zolid Gen-X. Adhesive surfaces were air-abraded and bonded with MDP-containing resin cement. Specimens were subjected to thermocycling (5-55 degrees C, 5000 cycles); then, mechanical loading (1.2 x 106 cycles, 49 N). Surviving specimens were loaded until failure in the universal testing machine. All specimens were examined under stereomicroscope, and two samples from each group were evaluated using Scanning Electron Microscope.Results: Mean failure loads were not significantly different between different framework designs or between two materials. However, IPS e.max ZirCAD Prime showed significantly higher failure rate than Zolid Gen-X during dynamic fatigue (p = 0.009). Samples with D1 design showed higher debonding rate, D2 failed mainly by fracture of the palatal wing and debonding, and D3 failed mainly by fracture of the abutment tooth. Debonded restorations showed mainly mixed failures.Conclusion: Cantilever IRFDPs with framework designs that maximize adhesion to enamel exhibited promising results. IPS e.max ZirCAD Prime was more susceptible to fractures with the long palatal wing design.
引用
收藏
页数:8
相关论文
共 46 条
[1]   Zirconia ceramic inlay-retained fixed dental prostheses - first clinical results with a new design [J].
Abou Tara, Milia ;
Eschbach, Stephanie ;
Wolfart, Stefan ;
Kern, Matthias .
JOURNAL OF DENTISTRY, 2011, 39 (03) :208-211
[2]   Classification and Properties of Dental Zirconia as Implant Fixtures and Superstructures [J].
Ban, Seiji .
MATERIALS, 2021, 14 (17)
[3]   Influence of different preparation forms on the loading-bearing capacity of zirconia cantilever FDPs. A laboratory study [J].
Bishti, Shaza ;
Jaekel, Carsten ;
Kern, Matthias ;
Wolfart, Stefan .
JOURNAL OF PROSTHODONTIC RESEARCH, 2019, 63 (03) :347-353
[4]  
Bömicke W, 2018, J PROSTHET DENT, V119, P89, DOI 10.1016/j.prosdent.2017.03.006
[5]   All-ceramic inlay-retained fixed dental prostheses for replacing posterior missing teeth: A systematic review [J].
Castillo-Oyague, Raquel ;
Sancho-Esper, Rocio ;
Lynch, Christopher D. ;
Suarez-Garcia, Maria-Jesus .
JOURNAL OF PROSTHODONTIC RESEARCH, 2018, 62 (01) :10-23
[6]   Five-year clinical outcome of posterior zirconia ceramic inlay-retained FDPs with a modified design [J].
Chaar, M. Sad ;
Kern, Matthias .
JOURNAL OF DENTISTRY, 2015, 43 (12) :1411-1415
[7]   All-ceramic inlay-retained fixed dental prostheses: An update [J].
Chaar, M. Sad ;
Passia, Nicole ;
Kern, Matthias .
QUINTESSENCE INTERNATIONAL, 2015, 46 (09) :781-788
[8]   A systematic review of the survival and complication rates of inlay-retained fixed dental prostheses [J].
Chen, Junyu ;
Cai, He ;
Suo, Lai ;
Xue, Yiyuan ;
Wang, Jian ;
Wan, Qianbing .
JOURNAL OF DENTISTRY, 2017, 59 :2-10
[9]   Tooth structure removal associated with various preparation designs for anterior teeth [J].
Edelhoff, D ;
Sorensen, JA .
JOURNAL OF PROSTHETIC DENTISTRY, 2002, 87 (05) :503-509
[10]  