In vitro study of the accuracy and efficiency of wireless intraoral scanners at various battery levels

被引:2
作者
Sombun, Santisuk [1 ]
Ongthiemsak, Chakree [1 ]
机构
[1] Prince Songkla Univ, Fac Dent, Dept Prosthet Dent, 15 Kanjanavanich Rd, Hat Yai Dist, Songkhla, Thailand
关键词
Accuracy; Battery level; Image capture; Intraoral scanner; Scan time; Wireless; ARCH IMPRESSIONS; PRECISION; EXECUTION; TRUENESS; TIME;
D O I
10.1016/j.jdent.2023.104686
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To determine the trueness and precision of 2 wireless intraoral scanners (IOSs) under various battery levels, and assess scanning efficiency.Methods: A maxillary cast with 4 metal spheres attached was fabricated. Two wireless IOSs (TRIOS 3 and TRIOS 4) were evaluated under 3 battery levels (1-30%, 31-60%, and 61-100%; n = 30). Six horizontal distances and 1 vertical distance were measured between 4 spherical centers and 1 generated plane. The distance deviations were determined with a coordinate-measuring machine data set. Kruskal-Wallis and Levene tests were used to analyze trueness and precision. Scan time and the number of three-dimensional (3D) images captured were analyzed by using a 2-way analysis of variance.Results: In terms of trueness and precision, no significant differences were found at various battery levels over the majority of the measured distances. TRIOS 4 demonstrated better trueness than TRIOS 3 for cross-arch scan. The 61-100% battery level resulted in the shortest scan time and the least number of 3D images captured (p < 0.001). Scan time and number of 3D images captured were strongly correlated for TRIOS 3 (r = 0.66) and TRIOS 4 (r = 0.89).Conclusions: Changes in battery level had no impact on the trueness and precision of TRIOS 3 and TRIOS 4. High battery level IOSs resulted in faster scans and fewer 3D images captured with less storage space. TRIOS 4 scanned faster, captured fewer images, and demonstrated better trueness than TRIOS 3. Clinical significance: Although all battery levels of wireless IOSs provide comparable trueness and precision, a wireless IOS with a high battery level is more time efficient than one with a low battery level in complete-arch scan.
引用
收藏
页数:7
相关论文
共 50 条
  • [11] Ender A, 2019, INT J COMPUT DENT, V22, P11
  • [12] Ender A, 2013, INT J COMPUT DENT, V16, P11
  • [13] Influence of rescanning mesh holes and stitching procedures on the complete-arch scanning accuracy of an intraoral scanner: An in vitro study
    Gomez-Polo, Miguel
    Piedra-Cascon, Wenceslao
    Methani, Mohammed M.
    Quesada-Olmo, Nieves
    Farjas-Abadia, Mercedes
    Revilla-Leon, Marta
    [J]. JOURNAL OF DENTISTRY, 2021, 110
  • [14] Clinical use of a direct chairside oral scanner: An assessment of accuracy, time, and patient acceptance
    Gruenheid, Thorsten
    McCarthy, Shawn D.
    Larson, Brent E.
    [J]. AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2014, 146 (05) : 673 - 682
  • [15] A new method for the evaluation of the accuracy of full-arch digital impressions in vitro
    Gueth, Jan-Frederik
    Edelhoff, Daniel
    Schweiger, Josef
    Keul, Christine
    [J]. CLINICAL ORAL INVESTIGATIONS, 2016, 20 (07) : 1487 - 1494
  • [16] Effect of Software Version on the Accuracy of an Intraoral Scanning Device
    Haddadi, Yasser
    Bahrami, Golnosh
    Isidor, Flemming
    [J]. INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2018, 31 (04) : 375 - 376
  • [17] International Organization for Standardization, 2022, ISO 20896-1:2019
  • [18] International Organization for Standardization, 2020, ISO 5725-1
  • [19] Effect of different arch widths on the accuracy of three intraoral scanners
    Kaewbuasa, Narin
    Ongthiemsak, Chakree
    [J]. JOURNAL OF ADVANCED PROSTHODONTICS, 2021, 13 (04) : 205 - 215
  • [20] A new method for assessing the accuracy of full arch impressions in patients
    Kuhr, F.
    Schmidt, A.
    Rehmann, P.
    Woestmann, B.
    [J]. JOURNAL OF DENTISTRY, 2016, 55 : 68 - 74