Validity and Reliability of the Kinvent Handheld Dynamometer in the Athletic Shoulder Test

被引:8
|
作者
Olds, Margie [1 ,2 ]
McLaine, Sally [3 ]
Magni, Nico [4 ]
机构
[1] Flawless Mot Ltd, Auckland, New Zealand
[2] Univ Otago, Dept Physiotherapy, Dunedin, New Zealand
[3] Univ Tasmania, Sch Hlth Sci, Physiotas Launceston, Hobart, Tas, Australia
[4] Auckland Univ Technol, Sch Clin Sci, Dept Physiotherapy, Auckland, New Zealand
关键词
strength assessment; torque; force; return to sport; RUGBY UNION; INJURIES;
D O I
10.1123/jsr.2022-0444
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Context: Long-lever shoulder strength tests may aid clinical decision-making regarding return to sport after a shoulder injury. The Athletic Shoulder Test (AST) was developed to measure force production in 3 positions of shoulder abduction (90 degrees, 135 degrees, and 180 degrees) using force plates. However, handheld dynamometers (HHDs) are more portable, affordable, and may provide valid and reliable results which would increase the clinical utility of long-lever tests. HHDs vary in shape, design, and their capacity to report parameters such as rate of force production and require further investigation. The aim of this study was to examine the intrarater reliability of the Kinvent HHD and assess its validity against Kinvent force plates in the AST. Peak force (in kilograms), torque (in Newton meters), and normalized torque (in Newton meters per kilogram) were reported. Design: Validity and reliability study.Methods: Twenty-seven participants with no history of upper limb injury performed the test in a randomized order using the Kinvent HHD and force plates. Each condition was assessed 3 times, and peak force was recorded. Arm length was measured to calculate peak torque. Normalized peak torque was calculated by dividing torque by bodyweight (in kilograms).Results: The Kinvent HHD is reliable when measuring force (intraclass correlation coefficient [ICC] >_ .80), torque (ICC >_ .84), and normalized torque (ICC >_ .64) during the AST. The Kinvent HHD is also valid when compared with the Kinvent force plates for force (ICC >_ .79; r >_ .82), torque (ICC >_ .82; r >_ .76), and normalized torque (ICC >_ .71; r >_ .61). There were no statistically significant differences across the 3 trials on analyses of variance (P > .05).Conclusions: The Kinvent HHD is a reliable tool when used to measure force, torque, and normalized torque in the AST. Furthermore, given the lack of significant difference between trials, clinicians can use one test to accurately report relative peak force/torque/ normalized torque rather than average 3 separate trials. Finally, the Kinvent HHD is valid when compared with Kinvent force plates.
引用
收藏
页码:764 / 772
页数:9
相关论文
共 50 条
  • [1] Concurrent Validity and Reliability of a Handheld Dynamometer in Measuring Isometric Shoulder Rotational Strength
    Chen, Bin
    Liu, Lifen
    Chen, Lincoln Bin
    Cao, Xianxin
    Han, Peng
    Wang, Chenhao
    Qi, Qi
    JOURNAL OF SPORT REHABILITATION, 2021, 30 (06) : 965 - 968
  • [2] Measuring Eccentric Strength of the Shoulder External Rotators Using a Handheld Dynamometer: Reliability and Validity
    Johansson, Fredrik R.
    Skillgate, Eva
    Lapauw, Mattis L.
    Clijmans, Dorien
    Deneulin, Valentijn P.
    Palmans, Tanneke
    Cools, Ann M.
    JOURNAL OF ATHLETIC TRAINING, 2015, 50 (07) : 719 - 725
  • [3] Test-retest reliability of isometric shoulder muscle strength measurement with a handheld dynamometer and belt
    Katoh, Munenori
    JOURNAL OF PHYSICAL THERAPY SCIENCE, 2015, 27 (06) : 1719 - 1722
  • [4] Validity and reliability of a portable handheld dynamometer compared to a fixed isokinetic dynamometer to assess forearm torque strength
    Bonhof-Jansen, E. D. J.
    van Ham, A.
    Kroon, G. J.
    Winter, R. W.
    Brink, S. M.
    HAND SURGERY & REHABILITATION, 2023, 42 (02): : 147 - 153
  • [5] Validity and Reliability of upper limb muscles strength measurements by handheld dynamometer in COPD
    Beaumont, M.
    Peran, L.
    Berriet, A.
    Keller, R.
    Le Ber, C.
    EUROPEAN RESPIRATORY JOURNAL, 2022, 60
  • [6] The Modified-Athletic Shoulder Test: Reliability and validity of a new on-field assessment tool
    Tooth, Camille
    Forthomme, Benedicte
    Croisier, Jean-Louis
    Gofflot, Amandine
    Bornheim, Stephen
    Schwartz, Cedric
    PHYSICAL THERAPY IN SPORT, 2022, 58 : 8 - 15
  • [7] Assessment of isometric strength of the shoulder rotators in swimmers using a handheld dynamometer: a reliability study
    Conceicao, Ana
    Parraca, Jose
    Marinho, Daniel
    Costa, Mario
    Louro, Hugo
    Silva, Antonio
    Batalha, Nuno
    ACTA OF BIOENGINEERING AND BIOMECHANICS, 2018, 20 (04) : 113 - 119
  • [8] Reliability and Validity of Seated Hip Extensor Strength Measurement by Handheld Dynamometer in Older Adults
    Seko, Toshiaki
    Mori, Mitsuru
    Ohnishi, Hirofumi
    Himuro, Nobuaki
    Takahashi, Yui
    Kumamoto, Tsuneo
    Ito, Toshikazu
    JOURNAL OF GERIATRIC PHYSICAL THERAPY, 2019, 42 (04) : E39 - E44
  • [9] Absolute reliability of shoulder joint horizontal adductor muscle strength measurements using a handheld dynamometer
    Hirano, Masahiro
    Katoh, Munenori
    JOURNAL OF PHYSICAL THERAPY SCIENCE, 2015, 27 (07) : 2125 - 2127
  • [10] Measuring seated hip extensor strength using a handheld dynamometer: an examination of the reliability and validity of the protocol
    Seko, Toshiaki
    Kumamoto, Tsuneo
    Miura, Sayo
    Kobayashi, Takumi
    Takahashi, Yui
    Kaneko, Ryousuke
    Morino, Akira
    Ito, Toshikazu
    JOURNAL OF PHYSICAL THERAPY SCIENCE, 2015, 27 (07) : 2179 - 2182