Comparison of mainstream end tidal carbon dioxide on Y-piece side versus patient side of heat and moisture exchanger filters in critically ill adult patients: a prospective observational study

被引:2
作者
Tamashiro, Satoshi [1 ,2 ,3 ]
Nakayama, Izumi [4 ,5 ]
Gibo, Koichiro [6 ]
Izawa, Junichi [4 ,7 ]
机构
[1] Okinawa Prefectural Nanbu Med Ctr, Dept Clin Engn, Okinawa, Japan
[2] Childrens Med Ctr, Okinawa, Japan
[3] Okinawa Prefectural Chubu Hosp, Dept Clin Engn, Okinawa, Japan
[4] Okinawa Prefectural Chubu Hosp, Div Crit Care Med, Dept Internal Med, Okinawa, Japan
[5] Yokohama City Univ, Grad Sch Data Sci, Dept Hlth Data Sci, Kanazawa Ku, Yokohama, Kanagawa, Japan
[6] Okinawa Prefectural Chubu Hosp, Dept Emergency Med, Okinawa, Japan
[7] Kyoto Univ, Sch Publ Hlth, Dept Prevent Serv, Kyoto, Japan
关键词
Capnography; End-tidal carbon dioxide; Humidifier; Intensive care unit; Mechanical ventilation; Monitoring; EQUIVALENCE RANDOMIZED-TRIALS; BREATHING SYSTEM FILTERS; NONINFERIORITY; EXTENSION; INFANTS;
D O I
10.1007/s10877-022-00901-6
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The purpose of the study was to investigate the accuracy of mainstream EtCO2 measurements on the Y-piece (filtered) side of the heat and moisture exchanger filter (HMEF) in adult critically ill patients, compared to that on the patient (unfiltered) side of HMEF. We conducted a prospective observational method comparison study between July 2019 and December 2019. Critically ill adult patients receiving mechanical ventilation with HMEF were included. We performed a noninferiority comparison of the accuracy of EtCO2 measurements on the two sides of HMEF. The accuracy was measured by the absolute difference between PaCO2 and EtCO2. We set the non-inferiority margin at + 1 mmHg in accuracy difference between the two sides of HMEF. We also assessed the agreement between PaCO2 and EtCO2 using Bland-Altman analysis. Among thirty-seven patients, the accuracy difference was - 0.14 mmHg (two-sided 90% CI - 0.58 to 0.29), and the upper limit of the CI did not exceed the predefined margin of + 1 mmHg, establishing non-inferiority of EtCO2 on the Y-piece side of HMEF (P for non-inferiority < 0.001). In the Bland-Altman analyses, 95% limits of agreement between PaCO2 and EtCO2 were similar on both sides of HMEF (Y-piece side, - 8.67 to + 10.65 mmHg; patient side, - 8.93 to + 10.67 mmHg). The accuracy of mainstream EtCO2 measurements on the Y-piece side of HMEF was noninferior to that on the patient side in critically ill adults. Mechanically ventilated adult patients could be accurately monitored with mainstream EtCO2 on the Y-piece side of the HMEF unless their tidal volume was extremely low.
引用
收藏
页码:399 / 407
页数:9
相关论文
共 30 条
  • [1] Ad Hoc Committee for Early Rehabilitation The Japanese Society of Intensive Care Medicine, 2017, J Jpn Soc Intensive Care Med, V24, P255
  • [2] [Anonymous], COVID 19 AN MACH EQ
  • [3] BADGWELL JM, 1987, ANESTH ANALG, V66, P959
  • [4] Effect of Heat and Moisture Exchange Filter on the Dead Space Ventilation in a Pediatric Patient
    Bhagwat, Mrunmayee
    Surve, Rohini M.
    Patwardhan, Prathamesh
    [J]. JOURNAL OF NEUROANAESTHESIOLOGY AND CRITICAL CARE, 2022, 09 (01) : 41 - 43
  • [5] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
  • [6] Bossuyt PM, 2015, BMJ-BRIT MED J, V351, DOI [10.1373/clinchem.2015.246280, 10.1136/bmj.h5527, 10.1148/radiol.2015151516]
  • [7] Mainstream vs. sidestream capnometry for prediction of arterial carbon dioxide tension during supine craniotomy
    Chan, KL
    Chan, MTV
    Gin, T
    [J]. ANAESTHESIA, 2003, 58 (02) : 149 - 155
  • [8] Cheifetz IM, 2007, RESP CARE, V52, P423
  • [9] Prolonged use of heat and moisture exchangers does not affect device efficiency or frequency rate of nosocomial pneumonia
    Davis, K
    Evans, SL
    Campbell, RS
    Johannigman, JA
    Luchette, FA
    Porembka, DT
    Branson, RD
    [J]. CRITICAL CARE MEDICINE, 2000, 28 (05) : 1412 - 1418
  • [10] Comparison of manufacturers' specifications for 44 types of heat and moisture exchanging filters
    Dellamonica, J
    Boisseau, N
    Goubaux, B
    Raucoules-Aimé, M
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2004, 93 (04) : 532 - 539