Safety of using a large femoral head on thin polyethylene for total hip arthroplasty based on different types of polyethylene

被引:1
|
作者
Do, Min Uk [1 ]
Moon, Nam Hoon [2 ]
Suh, Kuen Tak [3 ]
Kim, Jung Shin [1 ]
Lee, Sang-Min [1 ]
Shin, Won Chul [1 ]
机构
[1] Pusan Natl Univ, Sch Med, Res Inst Convergence Biomed Sci & Technol, Dept Orthoped Surg,Yangsan Hosp, 20 Geumo Ro, Yangsan 626770, Gyeongsangnam D, South Korea
[2] Pusan Natl Univ Hosp, Dept Orthoped Surg, Busan, South Korea
[3] Pusan Natl Univ, Sehung Hosp, Dept Orthoped Surg, Sch Med, Busan, South Korea
关键词
CROSS-LINKED POLYETHYLENE; FOLLOW-UP; WEAR; LINERS; DISLOCATION; MINIMUM; RATES; SIZE;
D O I
10.1038/s41598-023-50217-x
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The use of a large femoral head in total hip arthroplasty (THA) to stabilize and reduce the incidence of dislocation is on the increase, but concerns arise when combining them with small acetabular components due to potential mechanical failures in thin polyethylene (PE) liners. A single-institution, retrospective cohort study was conducted on 116 patients with minimum 2-year follow-up who received 36-mm femoral heads and acetabular components <= 52 mm, using either remelted highly cross-linked polyethylene (remelted HXLPE) or vitamin E-infused HXLPE (VEPE). Osteolysis and implant loosening were not observed in either group. Although a fracture of the PE liner was observed in each group (1.7%), the clinical outcomes were excellent, as the mean modified Harris Hip Score (mHHS) at the last follow-up was 93.5. Moreover, the mean linear wear rates measured by digital imaging software in both groups were low, with 0.035 mm/y in remelted HXLPE and 0.030 mm/y in VEPE. In conclusion, The use of a large femoral head on a thin PE liner can be a viable treatment option in patients who need to prioritize stability; however, careful attention should be paid to mechanical fractures of the PE liner.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Wear of polyethylene according to polyethylene type and femoral head material in hip arthroplasty
    Hamadouche, M.
    Zaoui, A.
    Langlois, J.
    El Hage, S.
    Scemama, C.
    Courpied, J. P.
    E-MEMOIRES DE L ACADEMIE NATIONALE DE CHIRURGIE, 2016, 15 (04): : 13 - 16
  • [2] The wear of a polyethylene socket articulating with a zirconia ceramic femoral head in canine total hip arthroplasty
    Hamada, Y
    Horiuchi, T
    Sano, Y
    Usui, I
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1999, 48 (03): : 301 - 308
  • [3] Standing Versus Supine Radiographs to Evaluate Femoral Head Penetration in the Polyethylene Liner After Total Hip Arthroplasty
    Maruyama, Masaaki
    Tensho, Keiji
    Wakabayashi, Shinji
    Hisa, Kenji
    JOURNAL OF ARTHROPLASTY, 2014, 29 (12) : 2415 - 2419
  • [4] Ceramic Head Fracture in Ceramic-on-Polyethylene Total Hip Arthroplasty
    Nho, Jae-Hwi
    Park, Jong-Seok
    Song, Ui-Seoub
    Kim, Woo-Jong
    Suh, You-Sung
    YONSEI MEDICAL JOURNAL, 2013, 54 (06) : 1550 - 1553
  • [5] Modified minimally invasive two-incision total hip arthroplasty using large diameter femoral head
    Park, Kyung Soon
    Yoon, Taek Rim
    Hwang, Sang Yun
    Lee, Keun Bae
    INDIAN JOURNAL OF ORTHOPAEDICS, 2012, 46 (01) : 29 - 35
  • [6] Damage of an Oxinium femoral head and polyethylene liner following 'routine' total hip replacement
    McCalden, R. W.
    Charron, K. D.
    Davidson, R. D.
    Teeter, M. G.
    Holdsworth, D. W.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (03): : 409 - 413
  • [7] Recent update on crosslinked polyethylene in total hip arthroplasty
    Langlois, Jean
    Hamadouche, Moussa
    SICOT-J, 2020, 6
  • [8] Large Femoral Heads in Total Hip Arthroplasty With Vitamin E Highly Cross-Linked Polyethylene: Head Penetration Rates Compared to Highly Cross-Linked Polyethylene
    Nielson, Trent
    Owens, Garrett
    Miller, Brandon
    Meneghini, Ethan
    Deckard, Evan R.
    Meneghini, R. Michael
    JOURNAL OF ARTHROPLASTY, 2022, 37 (07) : S685 - S691
  • [9] Cementless Total Hip Arthroplasty With Alumina-on-Highly Cross-Linked Polyethylene Bearing in Young Patients With Femoral Head Osteonecrosis
    Kim, Young-Hoo
    Choi, Yoowang
    Kim, Jun-Shik
    JOURNAL OF ARTHROPLASTY, 2011, 26 (02) : 218 - 223
  • [10] Polyethylene Liner and Femoral Head Exchange in Total Hip Arthroplasty Factors Associated with Long-Term Success and Failure
    Petis, Stephen M.
    Kubista, Bernd
    Hartzler, Robert U.
    Abdel, Matthew P.
    Berry, Daniel J.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2019, 101 (05) : 421 - 428