Absolute calibration of the spectral sensitivity of an x-ray streak camera over the 0.1-10 keV spectral range equipped with CsI photocathode

被引:1
|
作者
Hubert, S. [1 ]
Boubault, F. [1 ]
机构
[1] CEA, CESTA, F-33116 Le Barp, France
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2023年 / 94卷 / 11期
关键词
CHARGE-COUPLED-DEVICE; DOUBLE-CRYSTAL MONOCHROMATOR; GRAZING-INCIDENCE; SYNCHROTRON; RADIOMETRY; GENERATOR; FACILITY; DETECTOR;
D O I
10.1063/5.0156102
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper, we report the absolute measurement of the spectral sensitivity of a bilamellar tube x-ray streak camera (XRSC) over the 0.1-10 keV range equipped with a CsI photocathode for Laser MegaJoule (LMJ) fusion experiments. This calibration of the XRSC is performed in static mode by using two multi-anode x-ray generators. Two silicon drift detectors (SDDs) previously calibrated at the Physikalisch-Technische-Bundesanstalt radiometric laboratory are used as secondary standards. Both x-ray generators work with a specific monochromator for radiometric measurements. In the sub-keV region, a 1 m-grazing incidence Rowland geometry monochromator specifically developed to handle LMJ's x-ray camera is used, whereas for higher energies (>2 keV), a double-crystal monochromator is employed. The absolute spectral sensitivity of the XRSC is obtained by comparing the CCD counts of the XRSC output with the output counts in x-ray lines recorded by the SDD. The results obtained below 1.2 keV are, to our knowledge, the first measurements of the spectral sensitivity of an XRSC in the soft x-ray range with a CsI photocathode. Comparison with a model describing the spectral dependence of the sensitivity of the XRSC revealed that measurements obtained in the sub-keV region are greater than expected, whereas they agree with the model above 4.5 keV. There may be several contributors to this behavior, including the grain morphology of the CsI layer and exposure to air.
引用
收藏
页数:10
相关论文
共 45 条
  • [1] Absolute x-ray calibration of a gated x-ray framing camera for the Laser MegaJoule facility in the 0.1 keV-1 keV spectral range
    Hubert, S.
    Boubault, F.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2021, 92 (03):
  • [2] Performance of a CsI photocathode in a hard x-ray streak camera
    Hara, T
    Tanaka, Y
    Kitamura, H
    Ishikawa, T
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (10): : 3624 - 3626
  • [3] Theoretical calculation of response sensitivity of CsI photocathode of soft X-ray streak camera
    He, Xiao-An
    Yang, Jia-Min
    Li, Yu-Kun
    Li, Jin
    Xiong, Gang
    ACTA PHYSICA SINICA, 2023, 72 (24)
  • [4] Calculation of CsI photocathode spectral response in 10-100 keV X-ray energy region
    Li Yu-Kun
    Chen Tao
    Li Jin
    Yang Zhi-Wen
    Hu Xin
    Deng Ke-Li
    Cao Zhu-Rong
    ACTA PHYSICA SINICA, 2018, 67 (08)
  • [5] Spectral sensitivity calibration of Au and CsI photocathodes of high speed X-ray scanning camera
    Yuan, Zheng
    Zeng, Peng
    Deng, Bo
    Chen, Tao
    Cao, Zhurong
    Liu, Shenye
    Yuan, Yongteng
    6TH INTERNATIONAL SYMPOSIUM ON ADVANCED OPTICAL MANUFACTURING AND TESTING TECHNOLOGIES: OPTOELECTRONIC MATERIALS AND DEVICES FOR SENSING, IMAGING, AND SOLAR ENERGY, 2012, 8419
  • [6] Fabrication and absolute calibration of X-ray streak camera cathode
    Li, Jin
    Hu, Xin
    Fan, Long
    Chen, Tao
    Yang, Zhiwen
    Deng, Bo
    Li, Yukun
    Cao, Zhurong
    Liu, Shenye
    Qiangjiguang Yu Lizishu/High Power Laser and Particle Beams, 2015, 27 (08):
  • [7] Device for absolute calibration of spectral sensitivity of semiconductor X-ray detectors
    Kropotov, A.Yu.
    Morgun, O.N.
    Chernyi, V.V.
    Instruments and experimental techniques New York, 1988, 31 (3 pt 1): : 572 - 575
  • [8] Absolute calibration of the soft X-ray streak camera static characteristic
    Yi, RQ
    Zheng, ZJ
    Hu, X
    Cui, MQ
    Zhao, YD
    Zhu, PP
    HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2003, 27 (05): : 452 - 454
  • [9] ABSOLUTE SPECTRAL SENSITIVITY OF LINEAR-PHOTODIODE ARRAYS IN 7-20 KEV ENERGY X-RAY RANGE
    DOLBNYA, IP
    KURYLO, SG
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01): : 677 - 679
  • [10] Soft (0.1-10 keV) and hard (>10 keV) X-ray multilayer mirrors for the XEUS astronomical mission
    Pareschi, G
    Cotroneo, V
    OPTICS FOR EUV, X-RAY AND GAMMA-RAY ASTRONOMY, 2004, 5168 : 53 - 64