Validation of the New MINI-CUBE for Clinic Determination of Erythrocyte Sedimentation Rate

被引:3
作者
Tomassetti, Flaminia [1 ,2 ]
Pelagalli, Martina [1 ,2 ]
Nicolai, Eleonora [1 ,2 ]
Sarubbi, Serena [1 ,2 ]
Calabrese, Cinzia [1 ,2 ]
Giovannelli, Alfredo [1 ,2 ]
Codella, Silvia [1 ,2 ]
Viola, Gemma [2 ]
Diamanti, Daniela [3 ]
Massoud, Renato [1 ,2 ]
Bernardini, Sergio [1 ,2 ]
Pieri, Massimo [1 ,2 ]
机构
[1] Univ Roma Tor Vergata, Dept Expt Med, I-00133 Rome, Italy
[2] Tor Vergata Univ Hosp, Dept Lab Med, I-00133 Rome, Italy
[3] DIESSE Diagnost Senese Spa, Res & Dev Dept, I-53035 Siena, SI, Italy
关键词
Erythrocyte sedimentation rate; Blood sedimentation; Method validation; BLOOD;
D O I
10.14740/jh1165
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Erythrocyte sedimentation rate (ESR) indirectly measures blood fibrinogen, and it is altered by all those pathological conditions that modify the aggregation of red blood cells. The international guidelines by the International Council for Standardization in Hematology (ICSH) define the Westergren method as the gold standard for ESR, although it is completely operator-dependent, time-consuming, and requires a patient's blood consumption. Therefore, the validation of new ESR analyzers is needed. The aim of this study is the validation of a new automated ESR analyzer, MINI-CUBE (DIESSE, Diagnostica Senese, Italy). Methods: The samples (n = 270) were collected at the University Hospital of the University of Rome Tor Vergata. A comparison between the automated instrument and the gold standard was performed. Statistical analyses were processed by MedCalc software. Results: The comparison analysis performed on the overall samples reported a good agreement, showing a Spearman rank correlation coefficient of 0.94 (P < 0.001), compared to the Westergren test. The Bland-Altman analysis showed a mean bias of 1.5 (maximum (max.):19.6; minimum (min.):-16.6). Inter-run (level 1 coefficient of variation (CV): 4.9%; level 2 CV: 0.8%), intra-run (level 1 CV: 21.1%; level 2 CV: 3.2%), and inter-instrument (level 1 CV: 27.1%; level 2 CV: 5.6%) precision were also assessed. The hematocrit value did not interfere with the analysis: Spearman rank correlation coefficient of 0.929 (P < 0.001); mean bias of 1.3 (max.:18.3; min.:-15.6). Conclusions: Overall results from MINI-CUBE asserted a good correlation rate with the gold standard, and it could be considered an accurate, and objective alternative for the Westergren test.
引用
收藏
页码:208 / 214
页数:7
相关论文
共 26 条
[1]  
Abramson N, 2006, BLOOD, V107, P4205
[2]  
Cicak H, 2022, Biochem Med (Zagreb), V32, P98
[3]  
Clinical and Laboratory Standards Institute, 2011, Procedures for ESR Test: Approved Standard, V31
[4]  
Clinical and Laboratory Standards Institute, 2014, CLSI EP15 A3 USER VE
[5]  
Clinical and Laboratory Standards Institute, 2010, CLSI EP 28-A3: 2010 EP28-A3c-Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory
[6]  
Approved Guideline, VThird
[7]   Evaluation of the Ves-Matic Cube 200 Erythrocyte Sedimentation Method Comparison With Westergren-Based Methods [J].
Curvers, Joyce ;
Kooren, Jurgen ;
Laan, Maartje ;
van Lierop, Edwin ;
van de Kerkhof, Daan ;
Scharnhorst, Volkher ;
Herruer, Martien .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2010, 134 (04) :653-660
[8]   A short history of the discovery of the erythrocyte sedimentation rate [J].
Grzybowski, A. ;
Sak, J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 (04) :442-444
[9]   A rapid and accurate method for estimating the erythrocyte sedimentation rate using a hematocrit-corrected optical aggregation index [J].
Higuchi, Makoto ;
Watanabe, Nobuo .
PLOS ONE, 2022, 17 (07)
[10]   ICSH recommendations for modified and alternate methods measuring the erythrocyte sedimentation rate [J].
Kratz, A. ;
Plebani, M. ;
Peng, M. ;
Lee, Y. K. ;
McCafferty, R. ;
Machin, S. J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 (05) :448-457