Alternating Look-Locker for quantitative T1, T1? and B1 3D MRI mapping

被引:0
|
作者
Wu, Lin [1 ,2 ]
Carchi, Chris [1 ,2 ]
Michaeli, Shalom [1 ,2 ]
Mangia, Silvia [1 ,2 ]
Idiyatullin, Djaudat [1 ,2 ,3 ]
机构
[1] Univ Minnesota Med Sch, Ctr Magnet Resonance Res, Minneapolis, MN USA
[2] Univ Minnesota Med Sch, Dept Radiol, Minneapolis, MN USA
[3] Univ Minnesota, Ctr MR Res, Radiol Dept, Minneapolis, MN 55455 USA
关键词
B1-map; Look-Locker; rat brain; SWIFT; T-1; T-1 & rho; zero-TE; MAGNETIZATION-TRANSFER; ADIABATIC PULSES; SPIN-LOCKING; RELAXATION; RESONANCE; PHASE; REDUCTION; ARTIFACTS; CONTRAST; BRAIN;
D O I
10.1002/mrm.29839
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop a new MRI method, entitled alternating Look-Locker (aLL), for quantitative T-1, T-1?, and B-1 3D mapping.Methods: A Look-Locker scheme that alternates magnetization from +Z and -Z axes of the laboratory frame is utilized in combination with a 3D Multi-Band Sweep Imaging with Fourier Transformation (MB-SWIFT) readout. The analytical solution describing the spin evolution during aLL, as well as the correction required for segmented acquisition were derived. The simultaneous B-1 and T-1 mapping are demonstrated on an agar/saline phantom and on an in-vivo rat head. T-1? relaxation was achieved by cyclically applying magnetization preparation (MP) modules consisting of two adiabatic pulses. T-1? values in the rat brain in-vivo and in a gadobenate dimeglumine (Gd-DTPA) phantom were compared to those obtained with a previously introduced steady-state (SS) method.Results: The accuracy and precision of the analytical solution was tested by Bloch simulations. With the application of MP modules, the aLL method provides simultaneous T(1 )and T-1? maps. Conversely, without it, the method can be used for simultaneous T-1 and B-1 mapping. T-1? values were similar with both aLL and SS techniques. However, the aLL method resulted in more robust quantita-tive mapping compared to the SS method. Unlike the SS method, the aLL method does not require additional scans for generating T-1 maps.Conclusion: The proposed method offers a new flexible tool for quantitative map-ping of T-1, T-1?, and B-1. The aLL method can also be used with readout schemes different from MB-SWIFT.
引用
收藏
页码:149 / 161
页数:13
相关论文
共 50 条
  • [21] T1 maps in upper abdomen: comparison of Look-Locker and B1-corrected variable-flip-angle (VFA) methods
    Wang, Zhaoqi
    Chu, Funing
    Bai, Bingmei
    Han, Shuai
    Zhang, Hongkai
    Yan, Xu
    Zhang, Shouning
    Dong, Yafeng
    Jiang, Lina
    Zhao, Yan
    Kamel, Ihab R. R.
    Li, Hailiang
    Qu, Jinrong
    CHINESE JOURNAL OF ACADEMIC RADIOLOGY, 2023, 6 (01) : 41 - 46
  • [22] T1 maps in upper abdomen: comparison of Look-Locker and B1-corrected variable-flip-angle (VFA) methods
    Zhaoqi Wang
    Funing Chu
    Bingmei Bai
    Shuai Han
    Hongkai Zhang
    Xu Yan
    Shouning Zhang
    Yafeng Dong
    Lina Jiang
    Yan Zhao
    Ihab R. Kamel
    Hailiang Li
    Jinrong Qu
    Chinese Journal of Academic Radiology, 2023, 6 : 41 - 46
  • [23] Residual learning for 3D motion corrected quantitative MRI: Robust clinical T1, T2 and proton density mapping
    Pirkl, Carolin M.
    Cencini, Matteo
    Kurzawski, Jan W.
    Waldmannstetter, Diana
    Li, Hongwei
    Sekuboyina, Anjany
    Endt, Sebastian
    Peretti, Luca
    Donatelli, Graziella
    Pasquariello, Rosa
    Costagli, Mauro
    Buonincontri, Guido
    Tosetti, Michela
    Menzel, Marion I.
    Menze, Bjoern H.
    MEDICAL IMAGING WITH DEEP LEARNING, VOL 143, 2021, 143 : 618 - 632
  • [24] A fast Look-Locker method for T 1 mapping of the head and neck region
    Chikui, Toru
    Tokumori, Kenji
    Zeze, Ryosuke
    Shiraishi, Tomoko
    Ichihara, Takahiro
    Hatakenaka, Masamitsu
    Yoshiura, Kazunori
    ORAL RADIOLOGY, 2009, 25 (01) : 22 - 29
  • [25] The effect of motion correction interpolation on quantitative T1 mapping with MRI
    Nachmani, Amitay
    Schurr, Roey
    Joskowicz, Leo
    Mezer, Aviv A.
    MEDICAL IMAGE ANALYSIS, 2019, 52 : 119 - 127
  • [26] Look-Locker和B1修正变量翻转角技术在正常乳腺组织T1值测量中的对照研究
    蒋辉
    曲金荣
    邵楠楠
    李祥
    孟帆
    张宏凯
    黎海亮
    放射学实践, 2019, 34 (01) : 26 - 29
  • [27] Silent T1 mapping using the variable flip angle method with B1 correction
    Ljungberg, Emil
    Wood, Tobias
    Solana, Ana Beatriz
    Kolind, Shannon
    Williams, Steven C. R.
    Wiesinger, Florian
    Barker, Gareth J.
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (02) : 813 - 824
  • [28] Simultaneous T1, T2, and T1ρ relaxation mapping of the lower leg muscle with MR fingerprinting
    Sharafi, Azadeh
    Medina, Katherine
    Zibetti, Marcelo W., V
    Rao, Smita
    Cloos, Martijn A.
    Brown, Ryan
    Regatte, Ravinder R.
    MAGNETIC RESONANCE IN MEDICINE, 2021, 86 (01) : 372 - 381
  • [29] 3D magnetic resonance fingerprinting for rapid simultaneous T1, T2, and T1ρ volumetric mapping of human articular cartilage at 3 T
    Sharafi, Azadeh
    Zibetti, Marcelo V. W.
    Chang, Gregory
    Cloos, Martijn
    Regatte, Ravinder R.
    NMR IN BIOMEDICINE, 2022, 35 (12)
  • [30] Flip angle profile correction for T1 and T2 quantification with look-locker inversion recovery 2D steady-state free precession imaging
    Cooper, Mitchell A.
    Nguyen, Thanh D.
    Spincemaille, Pascal
    Prince, Martin R.
    Weinsaft, Jonathan W.
    Wang, Yi