Real-world agreement of same-visit Tono-Pen vs Goldmann applanation intraocular pressure measurements using electronic health records

被引:2
作者
Gui, Haiwen [1 ]
Zhang, Youchen [2 ]
Chang, Robert T. [2 ]
Wang, Sophia Y. [2 ,3 ]
机构
[1] Stanford Univ, Sch Med, Stanford, CA USA
[2] Stanford Univ, Byers Eye Inst, Dept Ophthalmol, Palo Alto, CA USA
[3] 2370 Watson Court, Palo Alto, CA 94303 USA
关键词
Ophthalmology; Intraocular pressure; Electronic health record; Tonometry; Glaucoma; TONOMETRY; GLAUCOMA; BLINDNESS; XL;
D O I
10.1016/j.heliyon.2023.e18703
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: To compare intraocular pressure (IOP) obtained with Tono-Pen (TP) and Goldmann applanation (GAT) using large-scale electronic health records (EHR). Design: Retrospective cohort study. Methods: A single pair of eligible TP/GAT IOP readings was randomly selected from the EHR for each ophthalmology patient at an academic ophthalmology center (2013-2022), yielding 4550 eligible measurements. We used Bland-Altman analysis to describe agreement between TP/GAT IOP differences and mean IOP measurements. We also used multivariable logistic regression to identify factors associated with different IOP readings in the same eye, including demographics, glaucoma diagnosis, and central corneal thickness (CCT). Primary outcome metrics were discrepant measurements between TP and GAT as defined by two methods: Outcome A (normal TP despite elevated GAT measurements), and Outcome B (TP and GAT IOP differences >6 mmHg).Result: The mean TP/GAT IOP difference was 0.15 mmHg ( & PLUSMN; 5.49 mmHg 95% CI). There was high correlation between the measurements (r = 0.790, p < 0.001). We found that TP overestimated pressures at IOP <16.5 mmHg and underestimated at IOP >16.5 mmHg (Fig. 4). Discrepant measurements accounted for 2.6% (N = 116) and 5.2% (N = 238) for outcomes A and B respectively. Patients with thinner CCT had higher odds of discrepant IOP (OR 0.88 per 25 & mu;m increase, CI [0.84-0.92], p < 0.0001; OR 0.88 per 25 & mu;m increase, CI [0.84-0.92], p < 0.0001 for outcomes A and B respectively). Conclusion: In a real-world academic practice setting, TP and GAT IOP measurements demonstrated close agreement, although 2.6% of measurements showed elevated GAT IOP despite normal TP measurements, and 5.2% of measurements were >6 mmHg apart.
引用
收藏
页数:8
相关论文
共 26 条
[1]   EVALUATION OF THE TONO-PEN AND THE PULSAIR TONOMETERS [J].
ARMSTRONG, TA .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1990, 109 (06) :716-720
[2]   Comparison of Goldmann applanation tonometry with the Tonopen for measuring intraocular pressure in a population-based glaucoma survey in rural West Bengal [J].
Bandyopadhyay, M ;
Raychaudhuri, A ;
Lahiri, SK ;
Schwartz, EC ;
Myatt, M ;
Johnson, GJ .
OPHTHALMIC EPIDEMIOLOGY, 2002, 9 (03) :215-224
[3]   Tono-Pen versus Goldmann Applanation Tonometry [J].
Bao, Bo ;
Diaconita, Vlad ;
Schulz, David C. ;
Hutnik, Cindy .
OPHTHALMOLOGY GLAUCOMA, 2019, 2 (06) :435-439
[4]   Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry [J].
Bhartiya S. ;
Bali S.J. ;
Sharma R. ;
Chaturvedi N. ;
Dada T. .
International Ophthalmology, 2011, 31 (4) :297-302
[5]   Real-world comparison between the Tonopen and Goldmann applanation tonometry in a university glaucoma clinic [J].
Blumberg, Max J. ;
Varikuti, Venkata N. V. ;
Weiner, Asher .
INTERNATIONAL OPHTHALMOLOGY, 2021, 41 (05) :1815-1825
[6]   Goldmann tonometry versus the Tono-Pen XL for intraocular pressure measurement: an evaluation of the potential impact on clinical decision making in glaucoma [J].
Carrim, Zia I. ;
Lavy, Tim E. .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2009, 29 (06) :648-651
[7]   Systematic Review of the Agreement of Tonometers with Goldmann Applanation Tonometry [J].
Cook, Jonathan Alistair ;
Botello, Adriana Paola ;
Elders, Andrew ;
Ali, Alia Fathi ;
Azuara-Blanco, Augusto ;
McCormack, Kirsty ;
Burr, Jennifer Margaret .
OPHTHALMOLOGY, 2012, 119 (08) :1552-1557
[8]   Global causes of blindness and distance vision impairment 1990-2020: a systematic review and meta-analysis [J].
Flaxman, Seth R. ;
Bourne, Rupert R. A. ;
Resnikoff, Serge ;
Ackland, Peter ;
Braithwaite, Tasanee ;
Cicinelli, Maria V. ;
Das, Aditi ;
Jonas, Jost B. ;
Keeffe, Jill ;
Kempen, John H. ;
Leasher, Janet ;
Limburg, Hans ;
Naidoo, Kovin ;
Pesudovs, Konrad ;
Silvester, Alex ;
Stevens, Gretchen A. ;
Tahhan, Nina ;
Wong, Tien Y. ;
Taylor, Hugh R. .
LANCET GLOBAL HEALTH, 2017, 5 (12) :E1221-E1234
[9]  
FRENKEL REP, 1988, ARCH OPHTHALMOL-CHIC, V106, P750
[10]   Reduction of intraocular pressure and glaucoma progression - Results from the early manifest glaucoma trial [J].
Heijl, A ;
Leske, MC ;
Bengtsson, B ;
Hyman, L ;
Bengtsson, B ;
Hussein, M .
ARCHIVES OF OPHTHALMOLOGY, 2002, 120 (10) :1268-1279