Generation and comparison of 3D dosimetric reference datasets for COMS eye plaque brachytherapy using model-based dose calculations

被引:2
作者
Fletcher, Elizabeth M. [1 ]
Ballester, Facundo [2 ,3 ]
Beaulieu, Luc [4 ,5 ,6 ]
Morrison, Hali [7 ,8 ,9 ]
Poher, Audran [4 ,5 ,6 ]
Rivard, Mark J. [10 ,12 ]
Sloboda, Ron S. [11 ]
Vijande, Javier [2 ,3 ,13 ]
Thomson, Rowan M. [1 ,14 ]
机构
[1] Carleton Univ, Phys Dept, Carleton Lab Radiotherapy Phys, Ottawa, ON, Canada
[2] Univ Valencia UV, Dept Fis Atom Mol & Nucl, Burjassot, Spain
[3] Univ Valencia UV, Unidad Mixta Invest Radiofis Instrumentac Nucl Med, Inst Invest Sanitaria Fe IIS La Fe, Burjassot, Spain
[4] CHU Quebec, Ctr Rech CHU Quebec, Serv Phys Med & Axe Oncol, Quebec City, PQ, Canada
[5] Univ Laval, Dept Phys Genie Phys & Opt, Quebec City, PQ, Canada
[6] Univ Laval, Ctr Rech Canc, Quebec City, PQ, Canada
[7] Univ Calgary, Dept Oncol, Calgary, AB, Canada
[8] Univ Calgary, Dept Phys & Astron, Calgary, AB, Canada
[9] Tom Baker Canc Clin, Dept Med Phys, Calgary, AB, Canada
[10] Brown Univ, Warren Alpert Med Sch, Dept Radiat Oncol, Providence, RI USA
[11] Cross Canc Inst, Dept Med Phys, Edmonton, AB, Canada
[12] Univ Alberta, Dept Oncol, Edmonton, AB, Canada
[13] IF UV CSIC, Inst Fis Corpuscular, Burjassot, Spain
[14] Carleton Univ, Phys Dept, Carleton Lab Radiotherapy Phys, Ottawa, ON K1S5 B6, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Monte Carlo; ocular brachytherapy; treatment planning; COLLAPSED CONE SUPERPOSITION; MONTE-CARLO-SIMULATION; NUCLEAR-DATA; I-125; AAPM; PD-103; TOPAS; RECOMMENDATIONS; ALGORITHM; ENGINE;
D O I
10.1002/mp.16721
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeA joint Working Group of the American Association of Physicists in Medicine (AAPM), the European Society for Radiotherapy and Oncology (ESTRO), and the Australasian Brachytherapy Group (ABG) was created to aid in the transition from the AAPM TG-43 dose calculation formalism, the current standard, to model-based dose calculations. This work establishes the first test cases for low-energy photon-emitting brachytherapy using model-based dose calculation algorithms (MBDCAs).Acquisition and Validation MethodsFive test cases are developed: (1) a single model 6711 125I brachytherapy seed in water, 13 seeds (2) individually and (3) in combination in water, (4) the full Collaborative Ocular Melanoma Study (COMS) 16-mm eye plaque in water, and (5) the full plaque in a realistic eye phantom. Calculations are done with four Monte Carlo (MC) codes and a research version of a commercial treatment planning system (TPS). For all test cases, local agreement of MC codes was within & SIM;2.5% and global agreement was & SIM;2% (4% for test case 5). MC agreement was within expected uncertainties. Local agreement of TPS with MC was within 5% for test case 1 and & SIM;20% for test cases 4 and 5, and global agreement was within 0.4% for test case 1 and 10% for test cases 4 and 5.Data Format and Usage NotesDose distributions for each set of MC and TPS calculations are available online () along with input files and all other information necessary to repeat the calculations.Potential ApplicationsThese data can be used to support commissioning of MBDCAs for low-energy brachytherapy as recommended by TGs 186 and 221 and AAPM Report 372. This work additionally lays out a sample framework for the development of test cases that can be extended to other applications beyond eye plaque brachytherapy.
引用
收藏
页码:694 / 706
页数:13
相关论文
共 56 条
  • [1] Collapsed cone dose calculations for heterogeneous tissues in brachytherapy using primary and scatter separation source data
    Ahnesjo, Anders
    van Veelen, Bob
    Tedgren, Asa Carlsson
    [J]. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2017, 139 : 17 - 29
  • [2] On the uncertainties of photon mass energy-absorption coefficients and their ratios for radiation dosimetry
    Andreo, Pedro
    Burns, David T.
    Salvat, Francesc
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (08) : 2117 - 2136
  • [3] [Anonymous], 2015, PENELOPE-2014: A code system for Monte Carlo simulation of electron and photon transport
  • [4] A generic high-dose rate 192Ir brachytherapy source for evaluation of model-based dose calculations beyond the TG-43 formalism
    Ballester, Facundo
    Tedgren, Asa Carlsson
    Granero, Domingo
    Haworth, Annette
    Mourtada, Firas
    Fonseca, Gabriel Paiva
    Zourari, Kyveli
    Papagiannis, Panagiotis
    Rivard, Mark J.
    Siebert, Frank-Andre
    Sloboda, Ron S.
    Smith, Ryan L.
    Thomson, Rowan M.
    Verhaegen, Frank
    Vijande, Javier
    Ma, Yunzhi
    Beaulieu, Luc
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3048 - 3062
  • [5] AAPM WGDCAB Report 372: A joint AAPM, ESTRO, ABG, and ABS report on commissioning of model-based dose calculation algorithms in brachytherapy
    Beaulieu, Luc
    Ballester, Facundo
    Granero, Domingo
    Tedgren, asa Carlsson
    Haworth, Annette
    Lowenstein, Jessica R.
    Ma, Yunzhi
    Mourtada, Firas
    Papagiannis, Panagiotis
    Rivard, Mark J.
    Siebert, Frank-Andre
    Sloboda, Ron S.
    Smith, Ryan L.
    Thomson, Rowan M.
    Verhaegen, Frank
    Fonseca, Gabriel
    Vijande, Javier
    [J]. MEDICAL PHYSICS, 2023, 50 (08) : E946 - E960
  • [6] Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: Current status and recommendations for clinical implementation
    Beaulieu, Luc
    Tedgren, Asa Carlsson
    Carrier, Jean-Francois
    Davis, Stephen D.
    Mourtada, Firas
    Rivard, Mark J.
    Thomson, Rowan M.
    Verhaegen, Frank
    Wareing, Todd A.
    Williamson, Jeffrey F.
    [J]. MEDICAL PHYSICS, 2012, 39 (10) : 6208 - 6236
  • [7] Validation of the TOPAS Monte Carlo toolkit for HDR brachytherapy simulations
    Berumen, Francisco
    Ma, Yunzhi
    Ramos-Mendez, Jose
    Perl, Joseph
    Beaulieu, Luc
    [J]. BRACHYTHERAPY, 2021, 20 (04) : 911 - 921
  • [8] The collapsed cone superposition algorithm applied to scatter dose calculations in brachytherapy
    Carlsson, ÅK
    Ahnesjö, A
    [J]. MEDICAL PHYSICS, 2000, 27 (10) : 2320 - 2332
  • [9] Point kernels and superposition methods for scatter dose calculations in brachytherapy
    Carlsson, ÅK
    Ahnesjö, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02) : 357 - 382
  • [10] Experimental verification of Advanced Collapsed-cone Engine for use with a multichannel vaginal cylinder applicator
    Cawston-Grant, Brie
    Morrison, Hali
    Menon, Geetha
    Sloboda, Ron S.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (03): : 16 - 27