Microcatheter movement in the aneurysm due to low-profile visualized intraluminal support deployment: An in vitro study

被引:1
作者
Ikeda, Hiroyuki [1 ]
Kinosada, Masanori [1 ]
Uezato, Minami [1 ]
Kurosaki, Yoshitaka [1 ]
Chin, Masaki [1 ]
Yamagata, Sen [1 ]
机构
[1] Kurashiki Cent Hosp, Dept Neurosurg, 1-1-1 Miwa, Kurashiki 7108602, Japan
关键词
Aneurysm; Device; Stent; LVIS; Microcatheter; Jailing technique; INTRACRANIAL SACCULAR ANEURYSMS; COIL EMBOLIZATION; STENTS;
D O I
10.1016/j.neurad.2022.03.010
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: When a microcatheter is in the aneurysm, it may move due to low-profile visualized intralumi-nal support (LVIS) deployment. This study was designed to determine this mechanism.Methods: Six silicon aneurysm models were created by combining the aneurysm location (side wall or bifur-cation) and the parent vessel configuration (straight, ipsilateral bending, or contralateral bending). After adjusting the microcatheter tip position in the aneurysm by pushing or pulling, an LVIS stent was deployed to cover the aneurysm neck, and the changes in the microcatheter tip position was measured. Pushing and pulling were performed 15 times each for each model, for a total of 180 experiments.Results: In all experiments, the microcatheter tip moved with LVIS deployment. The total movement distance was 3.00</n>1.59 mm, which was significantly different between the push and pull groups (p = 0.049), between the three side-wall aneurysm models (p<0.0001), and between the three bifurcation aneurysm models (p<0.0001). Backward movement in the aneurysm occurred in 21% (37/180). The frequency of backward movement was significantly different between the side-wall and bifurcation aneurysm models (p = 0.0265) and between the push and pull groups (p<0.0001). The forward movement distance was significantly differ-ent between the side-wall (n = 78) and bifurcation (n = 65) aneurysm models (p<0.0001).Conclusions: The aneurysm location, the parent vessel configuration, and adjustment of the microcatheter tip position by pushing or pulling may affect the total movement distance and forward/backward movement of the microcatheter tip due to LVIS deployment.(c) 2022 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:223 / 229
页数:7
相关论文
共 50 条
  • [31] Y-configuration stenting for coiling of wide-neck bifurcation aneurysms using Low-profile Visualized Intraluminal Support Junior
    Park, Keun Young
    Kim, Byung Moon
    Kim, Dong Joon
    Chung, Joonho
    Lee, Jae Whan
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2019, 11 (04) : 400 - 404
  • [32] Low-Profile Visualized Intraluminal Support Jr Braided Stent Versus Atlas Self-Expandable Stent for Treatment of Intracranial Aneurysms: A Single Center Experience
    Monteiro, Andre
    Cortez, Gustavo M.
    Aghaebrahim, Amin
    Sauvageau, Eric
    Hanel, Ricardo A.
    [J]. NEUROSURGERY, 2021, 88 (02) : E170 - E178
  • [33] Low-Profile Visualized Intraluminal Support Junior Flow-Diversion Properties Used for Stabilization of a Superior Cerebellar Artery Pseudoaneurysm
    Winter, K. Austin
    Derrico, Nicholas P.
    Ziemke, Michael K.
    Robbins, Austin C.
    Luzardo, Gustavo
    Washington, Chad W.
    Strickland, Allison
    [J]. WORLD NEUROSURGERY, 2023, 180 : 6 - 9
  • [34] A new low-profile visualized intraluminal support (LVIS) device, LVIS Blue: laboratory comparison between old and new LVIS
    Chung, Joonho
    Matsuda, Yoshikazu
    Nelson, Jessica
    Keigher, Kiffon
    Lopes, Demetrius K.
    [J]. NEUROLOGICAL RESEARCH, 2018, 40 (01) : 78 - 85
  • [35] Enterprise stents versus low-profile visualized intraluminal support stents for stent-assisted coiling of unruptured paraclinoid aneurysms
    Wu, Zhi-Ping
    Yin, Jian-Lin
    Liu, Shan-Kai
    Ji, Song
    Liu, Jian-Yu
    Wang, Hai-Lan
    Zhang, Yi-Sheng
    Zhang, Da-Zhong
    [J]. TECHNOLOGY AND HEALTH CARE, 2023, 31 (05) : 1855 - 1865
  • [36] Endovascular Coil Embolization with Low-Profile Visualized Intraluminal Support Junior Stent for Ruptured Dissecting Aneurysm of Proximal Superior Cerebellar Artery-Case Report and Literature Review
    Onishi, Shumpei
    Sakamoto, Shigeyuki
    Sadatomo, Takashi
    Hara, Takeshi
    Ochiai, Junichiro
    Yuki, Kiyoshi
    Kurisu, Kaoru
    [J]. WORLD NEUROSURGERY, 2019, 122 : 102 - 105
  • [37] Endovascular Treatment of Ruptured Wide-Necked Anterior Communicating Artery Aneurysms Using a Low-Profile Visualized Intraluminal Support (LVIS) Device
    Xue, Gaici
    Liu, Peng
    Xu, Fengfeng
    Fang, Yibin
    Li, Qiang
    Hong, Bo
    Xu, Yi
    Liu, Jianmin
    Huang, Qinghai
    [J]. FRONTIERS IN NEUROLOGY, 2021, 11
  • [38] A comparison between the new Low-profile Visualized Intraluminal Support (LVIS Blue) stent and the Flow Redirection Endoluminal Device (FRED) in bench-top and cadaver studies
    Matsuda, Yoshikazu
    Chung, Joonho
    Keigher, Kiffon
    Lopes, Demetrius
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2018, 10 (03) : 274 - 278
  • [39] Utility of Low-Profile Visualized Intraluminal Support Junior Stent as a Rescue Therapy for Treating Ruptured Intracranial Aneurysms During Complicated Coil Embolization
    Kim, Seung Soo
    Park, Hyun
    Lee, Kwang Ho
    Jung, Seunguk
    Yoon, Chang Hyo
    Kim, Sung Kwon
    Ryu, Kyeong Hwa
    Baek, Hye Jin
    Hwang, Soo Hyun
    Kwon, O-Ki
    [J]. WORLD NEUROSURGERY, 2020, 135 : e710 - e715
  • [40] Intracranial Complex Ruptured Aneurysms Coiled with Overlapping Low-Profile Visualized Intraluminal Support Stents: Another Available Option for Complex Ruptured Intracranial Aneurysms
    Yan, Zhiyuan
    Zheng, Kuang
    Xiong, Ye
    Lan, Fujun
    Wang, Yunhui
    Tan, Xianxi
    Zhong, Ming
    Li, Zequn
    [J]. WORLD NEUROSURGERY, 2019, 125 : E22 - E28