Performance evaluation of the automated body fluid analysis of the new Sysmex XR haematology analyser

被引:2
作者
Van Laer, Christine [1 ,2 ]
Dedeene, Lieselot [3 ]
Gruwier, Lien [3 ]
Vanmechelen, Bram [3 ]
Hwandih, Talent [4 ]
Mintzas, Konstantinos [4 ]
Muenster, Marion [4 ]
Boeckx, Nancy [3 ,5 ]
机构
[1] Univ Hosp Leuven, Dept Lab Med, Clin Dept Lab Med, Herestr 49, B-3000 Leuven, Belgium
[2] Katholieke Univ Leuven, Ctr Mol & Vasc Biol, Dept Cardiovasc Sci, Leuven, Belgium
[3] Univ Hosp Leuven, Dept Lab Med, Clin Dept Lab Med, Leuven, Belgium
[4] Sysmex Europe SE, Norderstedt, Germany
[5] Katholieke Univ Leuven, Dept Oncol, Leuven, Belgium
关键词
automated analysis; body fluid; bronchoalveolar lavage (BAL); performance evaluation; Sysmex XN haematology analyser; Sysmex XR haematology analyser; COUNTING BLOOD-CELLS; BRONCHOALVEOLAR LAVAGE; CEREBROSPINAL-FLUID; CELLULAR ANALYSIS; MICROSCOPY; MODULE; MODE;
D O I
10.1515/labmed-2023-0064
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: The cellular composition of body fluids (BF) provides insight into disease pathology and is an important diagnostic parameter. The well-established Sysmex XN haematology analyser (XN) offers automated BF analysis and is in routine use in many laboratories. In this study the performance of the new Sysmex XR analyser (XR) in testing BF is compared to the XN. Methods: Cerebrospinal, pleural, peritoneal, and synovial fluids, as well as bronchoalveolar lavage (BAL) samples, were processed in BF mode on both analysers. Standard statistical methods were used to evaluate the performance of XR compared to the reference XN. Results: Cell counts were generated from a total of 356 measurements from 307 patient BF samples. After application of exclusion criteria, 86 cerebrospinal, 77 peritoneal, 78 pleural, and 89 synovial fluid sample pairs were compared. An excellent correlation coefficient (r) between the two analysers was observed for all diagnostic parameters; WBC-BF: 0.996, RBC-BF: 0.974, TC-BF: 0.995, MN#: 0.994, MN%: 0.830, PMN#: 0.985 and PMN%: 0.835. The research parameters also performed equally well (r values of 0.900-0.994). The performance of diluted and undiluted samples was equally good. Similar results for diagnostic and research parameters were obtained from 62 BAL samples (r values of 0.899-0.992). Conclusions: This study demonstrates that the automated BF counts on the new Sysmex XR analyser are equivalent to that of the XN for all BF types analysed.
引用
收藏
页码:257 / 265
页数:9
相关论文
共 25 条
[1]   Evaluation of biological fluid analysis using the sysmex XN automatic hematology analyzer [J].
Aguadero, Vicente ;
Cano-Corres, Ruth ;
Berlanga, Eugenio ;
Torra, Montserrat .
CYTOMETRY PART B-CLINICAL CYTOMETRY, 2018, 94 (05) :680-688
[2]  
Akoh Jacob A, 2012, World J Nephrol, V1, P106, DOI 10.5527/wjn.v1.i4.106
[3]   Automated cell count in body fluids: a review [J].
Alcaide Martin, Maria Jose ;
Altimira Queral, Laura ;
Sahuquillo Frias, Laura ;
Valina Amado, Laura ;
Merino, Anna ;
de Guadiana-Romualdo, Luis Garcia .
ADVANCES IN LABORATORY MEDICINE-AVANCES EN MEDICINA DE LABORATORIO, 2021, 2 (02) :149-161
[4]   Comparison between automated and microscopic analysis in body fluids cytology [J].
Bottini, P. V. ;
Pompeo, D. B. ;
Souza, M. I. ;
Garlipp, C. R. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 (02) :e16-e18
[5]   Two-site evaluation of the diagnostic performance of the Sysmex XN Body Fluid (BF) module for cell count and differential in Cerebrospinal Fluid [J].
Buoro, S. ;
Peruzzi, B. ;
Fanelli, A. ;
Seghezzi, M. ;
Manenti, B. ;
Lorubbio, M. ;
Biagioli, T. ;
Nannini, S. ;
Ottomano, C. ;
Lippi, G. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 (01) :26-33
[6]   Cell Population Data and reflex testing rules of cell analysis in pleural and ascitic fluids using body fluid mode on Sysmex XN-9000 [J].
Buoro, Sabrina ;
Mecca, Tommaso ;
Azzara, Giovanna ;
Seghezzi, Michela ;
Dominoni, Paola ;
Crippa, Alberto ;
Ottomano, Cosimo ;
Lippi, Giuseppe .
CLINICA CHIMICA ACTA, 2016, 452 :92-98
[7]   CLINICAL UTILITY OF BODY-FLUID ANALYSES [J].
CHESON, BD .
CLINICS IN LABORATORY MEDICINE, 1985, 5 (02) :195-208
[8]   Body fluid cellular analysis using the Sysmex XN-2000 automatic hematology analyzer: focusing on malignant samples [J].
Cho, Y. -U. ;
Chi, H. -S. ;
Park, S. H. ;
Jang, S. ;
Kim, Y. -J. ;
Park, C. -J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 (03) :346-356
[9]  
Chowdhary Sajeel, 2005, J Natl Compr Canc Netw, V3, P693
[10]  
CLSI, 2006, BOD FLUID AN CELL CO