A computational model to simulate spectral modulation and speech perception experiments of cochlear implant users

被引:3
|
作者
Alvarez, Franklin [1 ,2 ]
Kipping, Daniel [1 ,2 ]
Nogueira, Waldo [1 ,2 ]
机构
[1] Hannover Med Sch, Hannover, Germany
[2] Cluster Excellence Hearing4All, Hannover, Germany
基金
欧洲研究理事会;
关键词
computational model; cochlear implant; neural health; sound coding strategies; speech-in-noise recognition; spectral modulation detection; speech understanding prediction; PULSATILE ELECTRICAL-STIMULATION; AUDITORY-NERVE; SPIRAL GANGLION; INTELLIGIBILITY; DISCRIMINATION; INTERFERENCE; DEGENERATION; RECOGNITION; EXCITATION; RESOLUTION;
D O I
10.3389/fninf.2023.934472
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Speech understanding in cochlear implant (CI) users presents large intersubject variability that may be related to different aspects of the peripheral auditory system, such as the electrode-nerve interface and neural health conditions. This variability makes it more challenging to proof differences in performance between different CI sound coding strategies in regular clinical studies, nevertheless, computational models can be helpful to assess the speech performance of CI users in an environment where all these physiological aspects can be controlled. In this study, differences in performance between three variants of the HiRes Fidelity 120 (F120) sound coding strategy are studied with a computational model. The computational model consists of (i) a processing stage with the sound coding strategy, (ii) a three-dimensional electrode-nerve interface that accounts for auditory nerve fiber (ANF) degeneration, (iii) a population of phenomenological ANF models, and (iv) a feature extractor algorithm to obtain the internal representation (IR) of the neural activity. As the back-end, the simulation framework for auditory discrimination experiments (FADE) was chosen. Two experiments relevant to speech understanding were performed: one related to spectral modulation threshold (SMT), and the other one related to speech reception threshold (SRT). These experiments included three different neural health conditions (healthy ANFs, and moderate and severe ANF degeneration). The F120 was configured to use sequential stimulation (F120-S), and simultaneous stimulation with two (F120-P) and three (F120-T) simultaneously active channels. Simultaneous stimulation causes electric interaction that smears the spectrotemporal information transmitted to the ANFs, and it has been hypothesized to lead to even worse information transmission in poor neural health conditions. In general, worse neural health conditions led to worse predicted performance; nevertheless, the detriment was small compared to clinical data. Results in SRT experiments indicated that performance with simultaneous stimulation, especially F120-T, were more affected by neural degeneration than with sequential stimulation. Results in SMT experiments showed no significant difference in performance. Although the proposed model in its current state is able to perform SMT and SRT experiments, it is not reliable to predict real CI users' performance yet. Nevertheless, improvements related to the ANF model, feature extraction, and predictor algorithm are discussed.
引用
收藏
页数:19
相关论文
共 50 条
  • [31] Comparison of Two Music Training Approaches on Music and Speech Perception in Cochlear Implant Users
    Fuller, Christina D.
    Galvin, John J., III
    Maat, Bert
    Baskent, Deniz
    Free, Rolien H.
    TRENDS IN HEARING, 2018, 22
  • [32] Impact of Reverberation on Speech Perception in Noise in Bimodal/Bilateral Cochlear Implant Users with and without Residual Hearing
    Koenig, Clara
    Baumann, Uwe
    Stoever, Timo
    Weissgerber, Tobias
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (17)
  • [33] Event-related potentials for better speech perception in noise by cochlear implant users
    Soshi, Takahiro
    Hisanaga, Satoko
    Kodama, Narihiro
    Kanekama, Yori
    Samejima, Yasuhiro
    Yumoto, Eiji
    Sekiyama, Kaoru
    HEARING RESEARCH, 2014, 316 : 110 - 121
  • [34] Speech Perception With Music Maskers by Cochlear Implant Users and Normal-Hearing Listeners
    Eskridge, Elizabeth N.
    Galvin, John J., III
    Aronoff, Justin M.
    Li, Tianhao
    Fu, Qian-Jie
    JOURNAL OF SPEECH LANGUAGE AND HEARING RESEARCH, 2012, 55 (03): : 800 - 810
  • [35] Speech Quality Perception in Unilateral Cochlear Implant Users With Single-Sided Deafness
    Kelly, Scott
    Kuhlmey, Megan E.
    Despotidis, Meghan A.
    Alter, Isaac L.
    Hwa, Tiffany P.
    Chern, Alexander
    Lalwani, Anil K.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2025, 172 (01) : 224 - 232
  • [36] Evaluation of the Effect of Musical Perception Activities on Speech Perception in Adult Cochlear Implant Users
    Mutlu, Barak
    Topcu, Merve Torun
    Yuksel, Mustafa
    Kalcioglu, Mahmut Tayyar
    TURKISH ARCHIVES OF OTORHINOLARYNGOLOGY, 2022, 60 (04) : 188 - 198
  • [37] Relationships Among Peripheral and Central Electrophysiological Measures of Spatial and Spectral Selectivity and Speech Perception in Cochlear Implant Users
    Scheperle, Rachel A.
    Abbas, Paul J.
    EAR AND HEARING, 2015, 36 (04) : 441 - 453
  • [38] Loudness Balancing Optimization for Better Speech Intelligibility, Music Perception, and Spectral Temporal Resolution in Cochlear Implant Users
    Deniz, Burcu
    Deniz, Risvan
    Atas, Ahmet
    OTOLOGY & NEUROTOLOGY, 2024, 45 (05) : e385 - e392
  • [39] Spectral sparsification of speech signals and its interaction with top-down mechanisms in adult cochlear implant users
    Lentz, Benjamin
    Volter, Christiane
    Martin, Rainer
    SPEECH COMMUNICATION, 2022, 144 : 67 - 74
  • [40] Advanced Beamformers for Cochlear Implant Users: Acute Measurement of Speech Perception in Challenging Listening Conditions
    Buechner, Andreas
    Dyballa, Karl-Heinz
    Hehrmann, Phillipp
    Fredelake, Stefan
    Lenarz, Thomas
    PLOS ONE, 2014, 9 (04):