Correlation analysis between 2D and 3D patient-specific quality assurance for volumetric modulated arc therapy

被引:0
|
作者
Sharma, S. [1 ]
Sharma, D. [1 ]
Subramani, V.
Gopishankar, N.
Bhaskar, S. [1 ]
Pathy, S. [1 ]
Kumar, P. [1 ,2 ]
Thulkar, S. [1 ,3 ]
Chander, S. [1 ]
机构
[1] All India Inst Med Sci, Dr BRA Inst Rotary Canc Hosp, Dept Radiotherapy, New Delhi, India
[2] AIIMS, Dept Radiotherapy, New Delhi 110029, India
[3] AIIMS, Dept Radiodiag, New Delhi 110029, India
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2023年 / 21卷 / 04期
关键词
Patient specific QA; 2D measurements; COMPASS QA; Correlation analysis; GAMMA PASSING RATES; PER-BEAM; IMRT; METRICS; QA;
D O I
10.52547/ijrr.21.4.12
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Aim is to find correlation between 2D-gamma passing rate and 3D-DVHbased pre-treatment patient-specific quality assurance. Materials and Methods: 21 head and neck and 21 pelvis patients, treated with volumetric modulated arc therapy (VMAT) were selected for this study. All patients were planned with Elekta VersaHD linear accelerator using Monaco (5.11) treatment planning system. 2D-planar dose measurements were performed with IBA-I'matriXX evolution detector-array using MyQA-Patients software. For 2D-Gamma index evaluation, 3%/3mm and 2%/2mm criteria were used. 3D-dose measurements were performed using the IBA-COMPASS system. For 3D measurement, Monaco and COMPASS doses were compared in terms of percentage dose differences to PTV and organs at risk. For PTV D95, D2, and D50 (dose received by 95%, 2%, and 50% volume), similarly for OARs D2 and D50 were noted. 3D Gamma index was also noted. Correlation coefficient and its corresponding two-tailed p-value (<= 0.05, for statistically significant) were calculated for 2D-gamma passing rate and 3D Gamma index & percentage dose differences of 3D-DVH based metrics (Monaco calculated versus COMPASS measured). Strength of correlation will be considered weak or strong based on the r -value. Results: 2D-Gamma index passing rate was 98.6 +/- 1.8%, 92.1 +/- 7.1% and 98.5 +/- 1.3%, 93.5 +/- 4.4% for head-neck and pelvis patients (3%/3mm, 2%/2mm criteria) respectively. Percentage dose-differences for PTV D95, D2, D50 for head-neck and pelvis were: 4.22 +/- 2.09%, 4.25 +/- 2.23%, 3.93 +/- 1.59 & 0.60 +/- 1.96%, 1.53 +/- 1.64%, 1.59 +/- 1.20% respectively. Spine and brainstem D2 were 0.84 +/- 6.10%, 0.77 +/- 2.70%, bladder and rectum D50 were 3.75 +/- 3.31%, -2.19 +/- 3.60%. Conclusion: No strong correlation was observed between the 2D Gamma passing rate and 3D measurements.
引用
收藏
页码:685 / 691
页数:8
相关论文
共 50 条
  • [41] Effects of high-resolution measurements between different multi-row detectors on volumetric modulated arc therapy patient-specific quality assurance
    Kunii, Y.
    Tanabe, Y.
    Higashi, A.
    Nakamoto, A.
    Nishioka, K.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2023, 21 (03): : 413 - 420
  • [42] Investigation of a 2D Ionization Chamber Array for Quality Assurance of Intensity Modulated Arc Therapy
    Koch, N.
    Ashenafi, M.
    Li, J.
    Vanek, K.
    MEDICAL PHYSICS, 2008, 35 (06) : 2945 - +
  • [43] Automated 3D Gamma and Spot Regression Analysis as a Component of Proton Patient-Specific Quality Assurance
    Morales, D. Hernandez
    Liu, W.
    Shan, J.
    Augustine, K.
    Davis, M.
    Bues, M.
    Stoker, J.
    MEDICAL PHYSICS, 2017, 44 (06) : 2994 - 2994
  • [44] Electronic Portal Imaging Device in Pre-Treatment Patient-Specific Quality Assurance of volumetric-modulated arc therapy delivery
    Lau, M. L.
    Abdullah, R.
    Jayamani, J.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2022, 22
  • [45] Correlation between gamma index passing rate and clinical dosimetric difference for pre-treatment 2D and 3D volumetric modulated arc therapy dosimetric verification
    Jin, X.
    Yan, H.
    Han, C.
    Zhou, Y.
    Yi, J.
    Xie, C.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1047):
  • [46] Machine Log File and Calibration Errors-based Patient-specific Quality Assurance (QA) for Volumetric Modulated Arc Therapy (VMAT)
    Ma, Yangguang
    Mou, Xuanqin
    Beeraka, Narasimha M.
    Guo, Yuexin
    Liu, Junqi
    Dai, Jianrong
    Fan, Ruitai
    CURRENT PHARMACEUTICAL DESIGN, 2023, 29 (34) : 2738 - 2751
  • [47] Prediction of patient-specific quality assurance for volumetric modulated arc therapy using radiomics-based machine learning with dose distribution
    Ishizaka, Natsuki
    Kinoshita, Tomotaka
    Sakai, Madoka
    Tanabe, Shunpei
    Nakano, Hisashi
    Tanabe, Satoshi
    Nakamura, Sae
    Mayumi, Kazuki
    Akamatsu, Shinya
    Nishikata, Takayuki
    Takizawa, Takeshi
    Yamada, Takumi
    Sakai, Hironori
    Kaidu, Motoki
    Sasamoto, Ryuta
    Ishikawa, Hiroyuki
    Utsunomiya, Satoru
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (01):
  • [48] Error detection using a convolutional neural network with dose difference maps in patient-specific quality assurance for volumetric modulated arc therapy
    Kimura, Yuto
    Kadoya, Noriyuki
    Tomori, Seiji
    Oku, Yohei
    Jingu, Keiichi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 73 : 57 - 64
  • [49] Deep learning for patient-specific quality assurance of volumetric modulated arc therapy: Prediction accuracy and cost-sensitive classification performance
    Li, Guangjun
    Duan, Lian
    Xie, Lizhang
    Hu, Ting
    Wei, Weige
    Bai, Long
    Xiao, Qing
    Liu, Wenjie
    Zhang, Lei
    Bai, Sen
    Yi, Zhang
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 125
  • [50] Treatment plan complexity quantification for predicting gamma passing rates in patient-specific quality assurance for stereotactic volumetric modulated arc therapy
    Xue, Xudong
    Luan, Shunyao
    Ding, Yi
    Li, Xiangbin
    Li, Dan
    Wang, Jingya
    Ma, Chi
    Jiang, Man
    Wei, Wei
    Wang, Xiao
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024, 25 (09):