摘要
Background: Aim is to find correlation between 2D-gamma passing rate and 3D-DVHbased pre-treatment patient-specific quality assurance. Materials and Methods: 21 head and neck and 21 pelvis patients, treated with volumetric modulated arc therapy (VMAT) were selected for this study. All patients were planned with Elekta VersaHD linear accelerator using Monaco (5.11) treatment planning system. 2D-planar dose measurements were performed with IBA-I'matriXX evolution detector-array using MyQA-Patients software. For 2D-Gamma index evaluation, 3%/3mm and 2%/2mm criteria were used. 3D-dose measurements were performed using the IBA-COMPASS system. For 3D measurement, Monaco and COMPASS doses were compared in terms of percentage dose differences to PTV and organs at risk. For PTV D95, D2, and D50 (dose received by 95%, 2%, and 50% volume), similarly for OARs D2 and D50 were noted. 3D Gamma index was also noted. Correlation coefficient and its corresponding two-tailed p-value (<= 0.05, for statistically significant) were calculated for 2D-gamma passing rate and 3D Gamma index & percentage dose differences of 3D-DVH based metrics (Monaco calculated versus COMPASS measured). Strength of correlation will be considered weak or strong based on the r -value. Results: 2D-Gamma index passing rate was 98.6 +/- 1.8%, 92.1 +/- 7.1% and 98.5 +/- 1.3%, 93.5 +/- 4.4% for head-neck and pelvis patients (3%/3mm, 2%/2mm criteria) respectively. Percentage dose-differences for PTV D95, D2, D50 for head-neck and pelvis were: 4.22 +/- 2.09%, 4.25 +/- 2.23%, 3.93 +/- 1.59 & 0.60 +/- 1.96%, 1.53 +/- 1.64%, 1.59 +/- 1.20% respectively. Spine and brainstem D2 were 0.84 +/- 6.10%, 0.77 +/- 2.70%, bladder and rectum D50 were 3.75 +/- 3.31%, -2.19 +/- 3.60%. Conclusion: No strong correlation was observed between the 2D Gamma passing rate and 3D measurements.
引用
收藏
页码:685 / 691
页数:8
相关论文
共 50 条
  • [1] Correlation between patient-specific quality assurance in volumetric modulated arc therapy and 2D dose image features
    Yixiao Guo
    Jinyan Hu
    Yang Li
    Juntao Ran
    Hongyi Cai
    Scientific Reports, 13
  • [2] Correlation between patient-specific quality assurance in volumetric modulated arc therapy and 2D dose image features
    Guo, Yixiao
    Hu, Jinyan
    Li, Yang
    Ran, Juntao
    Cai, Hongyi
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [3] From 2D to 3D gamma passing rate tolerance and action limits for patient-specific quality assurance in volumetric-modulated arc therapy
    Zarros, Christos
    Patatoukas, George
    Kollaros, Nikos
    Chalkia, Marina
    Kougioumtzopoulou, Andromachi
    Kouloulias, Vasilios
    Platoni, Kalliopi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025,
  • [4] Characterization of a novel 2D array dosimeter for patient-specific quality assurance with volumetric arc therapy
    Stathakis, Sotirios
    Myers, Pamela
    Esquivel, Carlos
    Mavroidis, Panayiotis
    Papanikolaou, Nikos
    MEDICAL PHYSICS, 2013, 40 (07)
  • [5] A study on correlation between 2D and 3D gamma evaluation metrics in patient-specific quality assurance for VMAT
    Rajasekaran, Dhanabalan
    Jeevanandam, Prakash
    Sukumar, Prabakar
    Ranganathan, Arulpandiyan
    Johnjothi, Samdevakumar
    Nagarajan, Vivekanandan
    MEDICAL DOSIMETRY, 2014, 39 (04) : 300 - 308
  • [6] A Systematic Approach to Patient-Specific Quality Assurance for Volumetric Modulated Arc Therapy
    SONG Yu-lin
    Ceferino Obcemea
    Boris Mueller
    Borys Mychalczak
    ChineseJournalofBiomedicalEngineering, 2015, 24 (02) : 58 - 65
  • [7] A retrospective analysis for patient-specific quality assurance of volumetric-modulated arc therapy plans
    Li, Guangjun
    Wu, Kui
    Peng, Guang
    Zhang, Yingjie
    Bai, Sen
    MEDICAL DOSIMETRY, 2014, 39 (04) : 309 - 313
  • [8] Patient Specific Quality Assurance for Volumetric Modulated Arc Therapy (RapidArc) Using COMPASS 3D Dosimetry System
    Arun, G.
    Subramanian, V. Shanmuga
    Swamy, S. Thirumalai
    Kathirvel, M.
    Ramalingam, K.
    Kala, S.
    Kumar, S. Ashok
    Karthikeyan, K.
    Anantharaman, A.
    Kumar, N. Jagdish
    Arivarasan, I.
    Babaiah, M.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [9] Correlation between 2D and 3D patient specific quality assurance in head and neck cancers
    Bhasker, Suman
    STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 (12) : 1123 - 1123
  • [10] A study on the correlation between plan complexity and gamma index analysis in patient specific quality assurance of volumetric modulated arc therapy
    Rajasekaran, Dhanabalan
    Jeevanandam, Prakash
    Sukumar, Prabakar
    Ranganathan, Arulpandiyan
    Johnjothi, Samdevakumar
    Nagarajan, Vivekanandan
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2015, 20 (01) : 57 - 65