Concurrent validity of smartphone-based markerless motion capturing to quantify lower-limb joint kinematics in healthy and pathological gait

被引:21
作者
Horsak, Brian [1 ,2 ]
Eichmann, Anna [3 ]
Lauer, Kerstin [3 ]
Prock, Kerstin [1 ]
Krondorfer, Philipp [1 ]
Siragy, Tarique [1 ]
Dumphart, Bernhard [1 ,2 ]
机构
[1] St Polten Univ Appl Sci, Ctr Digital Hlth & Social Innovat, Campus Pl 1, A-3100 St Polten, Austria
[2] St Polten Univ Appl Sci, Inst Hlth Sci, Campus Pl 1, A-3100 St Polten, Austria
[3] St Polten Univ Appl Sci, Study Program Gait Anal & Rehabil, Campus Pl 1, A-3100 St Polten, Austria
关键词
Markerless motion capture; Gait analysis; Pose estimation; Deep learning; OpenCap; MUSCULOSKELETAL MODEL; RELIABILITY; SIMULATION;
D O I
10.1016/j.jbiomech.2023.111801
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Markerless motion capturing has the potential to provide a low-cost and accessible alternative to traditional marker-based systems for real-world biomechanical assessment. However, before these systems can be put into practice, we need to rigorously evaluate their accuracy in estimating joint kinematics for various gait patterns. This study evaluated the accuracy of a low-cost, open-source, and smartphone-based markerless motion capture system, namely OpenCap, for measuring 3D joint kinematics in healthy and pathological gait compared to a marker-based system. 21 healthy volunteers were instructed to walk with four different gait patterns: physiological, crouch, circumduction, and equinus gait. Three-dimensional kinematic data were simultaneously recorded using the markerless and a marker-based motion capture system. The root mean square error (RMSE) and the peak error were calculated between every joint kinematic variable obtained by both systems. We found an overall RMSE of 5.8 (SD: 1.8 degrees) and a peak error of 11.3 degrees (SD: 3.9). A repeated measures ANOVA with post hoc tests indicated significant differences in RMSE and peak errors between the four gait patterns (p < 0.05). Physiological gait presented the lowest, crouch and circumduction gait the highest errors. Our findings indicate a roughly comparable accuracy to IMU-based approaches and commercial markerless multi-camera solutions. However, errors are still above clinically desirable thresholds of two to five degrees. While our findings highlight the potential of markerless systems for assessing gait kinematics, they also underpin the need to further improve the underlying deep learning algorithms to make markerless pose estimation a valuable tool in clinical settings.
引用
收藏
页数:7
相关论文
共 28 条
  • [1] Dynamic optimization of human walking
    Anderson, FC
    Pandy, MG
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2001, 123 (05): : 381 - 390
  • [2] Baker R., 2013, MEASURING WALKING HD, V56, DOI [10.1111/dmcn.12288, DOI 10.1111/DMCN.12288]
  • [3] OpenPose: Realtime Multi-Person 2D Pose Estimation Using Part Affinity Fields
    Cao, Zhe
    Hidalgo, Gines
    Simon, Tomas
    Wei, Shih-En
    Sheikh, Yaser
    [J]. IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 2021, 43 (01) : 172 - 186
  • [4] A GAIT ANALYSIS DATA-COLLECTION AND REDUCTION TECHNIQUE
    DAVIS, RB
    OUNPUU, S
    TYBURSKI, D
    GAGE, JR
    [J]. HUMAN MOVEMENT SCIENCE, 1991, 10 (05) : 575 - 587
  • [5] AN INTERACTIVE GRAPHICS-BASED MODEL OF THE LOWER-EXTREMITY TO STUDY ORTHOPEDIC SURGICAL-PROCEDURES
    DELP, SL
    LOAN, JP
    HOY, MG
    ZAJAC, FE
    TOPP, EL
    ROSEN, JM
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (08) : 757 - 767
  • [6] Peak Tibiofemoral Contact Forces Estimated Using IMU-Based Approaches Are Not Significantly Different from Motion Capture-Based Estimations in Patients with Knee Osteoarthritis
    Di Raimondo, Giacomo
    Willems, Miel
    Killen, Bryce Adrian
    Havashinezhadian, Sara
    Turcot, Katia
    Vanwanseele, Benedicte
    Jonkers, Ilse
    [J]. SENSORS, 2023, 23 (09)
  • [7] Inertial Sensor-to-Segment Calibration for Accurate 3D Joint Angle Calculation for Use in OpenSim
    Di Raimondo, Giacomo
    Vanwanseele, Benedicte
    van der Have, Arthur
    Emmerzaal, Jill
    Willems, Miel
    Killen, Bryce Adrian
    Jonkers, Ilse
    [J]. SENSORS, 2022, 22 (09)
  • [8] G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences
    Faul, Franz
    Erdfelder, Edgar
    Lang, Albert-Georg
    Buchner, Axel
    [J]. BEHAVIOR RESEARCH METHODS, 2007, 39 (02) : 175 - 191
  • [9] DeeperCut: A Deeper, Stronger, and Faster Multi-person Pose Estimation Model
    Insafutdinov, Eldar
    Pishchulin, Leonid
    Andres, Bjoern
    Andriluka, Mykhaylo
    Schiele, Bernt
    [J]. COMPUTER VISION - ECCV 2016, PT VI, 2016, 9910 : 34 - 50
  • [10] Concurrent assessment of gait kinematics using marker-based and markerless motion capture
    Kanko, M. Robert
    Laende, K. Elise
    Davis, M. Elysia
    Selbie, W. Scott
    Deluzio, J. Kevin
    [J]. JOURNAL OF BIOMECHANICS, 2021, 127