Combined MR and Microwave Imaging - Prototype and Phantom Experiments

被引:0
作者
Meaney, Paul [1 ]
Geimer, Shireen [1 ]
Player, Grace [1 ]
Raynolds, Timothy [2 ]
Yang, Xiaoyu [3 ]
Paulsen, Keith [1 ]
机构
[1] Darmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[2] White River Technol, Lebanon, NH USA
[3] Quality ElectroDynam, Mayfield, OH USA
来源
2023 17TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION, EUCAP | 2023年
关键词
microwave imaging; magnetic resonance imaging; phantom experiments; breast imaging;
D O I
10.23919/EuCAP57121.2023.10133692
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We are developing a microwave breast imaging system that can operate simultaneously in a magnetic resonance imaging system. Being able to directly register the two image sets allows us to exploit the excellent spatial resolution of MR with the more specific nature of the recovered dielectric properties. This fused imaging configuration could ultimately obviate the need for invasive contrast agents such as gadolinium. The integrated system consists of a substantially slimmed down microwave illumination tank integrated with a custom 3-channel MR coil for improved signal-to-noise ratio. The microwave apparatus has been designed to minimize the overall height and be compatible with the MR environment specifically, material choice and cross-system signal coupling. We demonstrate our new imaging tank and present results from initial breast-like phantom experiments. These results are a major step towards integration, implementation of our 3D soft prior imaging algorithm and ultimately patient exams.
引用
收藏
页数:5
相关论文
共 10 条
  • [1] Using a priori data to improve the reconstruction of small objects in microwave tomography
    Fhager, Andreas
    Persson, Mikael
    [J]. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2007, 55 (11) : 2454 - 2462
  • [2] Comparison of no-prior and soft-prior regularization in biomedical microwave imaging
    Golnabi, Amir H.
    Meaney, Paul M.
    Geimer, Shireen D.
    Paulsen, Keith D.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2011, 36 (03) : 159 - 170
  • [3] Gadolinium deposition in the brain: summary of evidence and recommendations
    Gulani, Vikas
    Calamante, Fernando
    Shellock, Frank G.
    Kanal, Emanuel
    Reeder, Scott B.
    [J]. LANCET NEUROLOGY, 2017, 16 (07) : 564 - 570
  • [4] Hosseinzadegan S, 2021, IEEE T MICROW THEORY, V69, P2741, DOI [10.1109/tmtt.2021.3060597, 10.1109/TMTT.2021.3060597]
  • [5] Meaney P.M., 1995, THESIS
  • [6] A Contrast Source Inversion Algorithm Formulated Using the Log-Phase Formulation
    Rubaek, Tonny
    Meaney, Paulm.
    Paulsen, Keith D.
    [J]. INTERNATIONAL JOURNAL OF ANTENNAS AND PROPAGATION, 2011, 2011
  • [7] Application of magnetic resonance imaging to early detection of breast cancer
    Schnall, MD
    [J]. BREAST CANCER RESEARCH, 2001, 3 (01): : 17 - 21
  • [8] Complex permittivities of breast tumor tissues obtained from cancer surgeries
    Sugitani, Takumi
    Kubota, Shin-ichi
    Kuroki, Shin-ichiro
    Sogo, Kenta
    Arihiro, Koji
    Okada, Morihito
    Kadoya, Takayuki
    Hide, Michihiro
    Oda, Miyo
    Kikkawa, Takamaro
    [J]. APPLIED PHYSICS LETTERS, 2014, 104 (25)
  • [9] US. Food & Drug Administration, 2010, FDA DRUG SAF COMM NE
  • [10] Yang X., 2021, P INT SOC MAGN RES M